§ 820.75 Process validation.
(a) Where the results of a process cannot be fully verified by subsequent inspection and test, the process shall be validated with a high degree of assurance and approved according to established procedures. The validation activities and results, including the date and signature of the individual(s) approving the validation and where appropriate the major equipment validated, shall be documented.
(b) Each manufacturer shall establish and maintain procedures for monitoring and control of process parameters for validated processes to ensure that the specified requirements continue to be met.
(1) Each manufacturer shall ensure that validated processes are performed by qualified individual(s).
(2) For validated processes, the monitoring and control methods and data, the date performed, and, where appropriate, the individual(s) performing the process or the major equipment used shall be documented.
(c) When changes or process deviations occur, the manufacturer shall review and evaluate the process and perform revalidation where appropriate. These activities shall be documented.

Title 21 published on 2013-04-01

The following are only the Rules published in the Federal Register after the published date of Title 21.

For a complete list of all Rules, Proposed Rules, and Notices view the Rulemaking tab.

  • 2013-09-24; vol. 78 # 185 - Tuesday, September 24, 2013
    1. 78 FR 58786 - Unique Device Identification System
      GPO FDSys XML | Text
      DEPARTMENT OF HEALTH AND HUMAN SERVICES, Food and Drug Administration
      Final rule.
      This rule is effective December 23, 2013, except §§ 801.55, 830.10, 830.100, 830.110, 830.120, and 830.130 are effective October 24, 2013. The incorporation by reference of § 830.20 listed in the rule is approved by the Director of the Office of the Federal Register as of December 23, 2013. The incorporation by reference of §§ 830.10 and 830.100 listed in the rule is approved by the Director of the Office of the Federal Register as of October 24, 2013. Certain provisions have later compliance dates as discussed in section VII. B. “Compliance Dates.”
      21 CFR Parts 16, 801, 803, 806, 810, 814, 820, 821, 822, and 830

Title 21 published on 2013-04-01

The following are ALL rules, proposed rules, and notices (chronologically) published in the Federal Register relating to 21 CFR 820 after this date.

  • 2013-09-24; vol. 78 # 185 - Tuesday, September 24, 2013
    1. 78 FR 58786 - Unique Device Identification System
      GPO FDSys XML | Text
      DEPARTMENT OF HEALTH AND HUMAN SERVICES, Food and Drug Administration
      Final rule.
      This rule is effective December 23, 2013, except §§ 801.55, 830.10, 830.100, 830.110, 830.120, and 830.130 are effective October 24, 2013. The incorporation by reference of § 830.20 listed in the rule is approved by the Director of the Office of the Federal Register as of December 23, 2013. The incorporation by reference of §§ 830.10 and 830.100 listed in the rule is approved by the Director of the Office of the Federal Register as of October 24, 2013. Certain provisions have later compliance dates as discussed in section VII. B. “Compliance Dates.”
      21 CFR Parts 16, 801, 803, 806, 810, 814, 820, 821, 822, and 830