A. In addition to the definitions in A.R.S.
§
36-2801 and
R9-17-101, the following
definitions apply in this Section unless otherwise stated:
1. "Limit of quantitation" means the lowest
concentration of an analyte that may be detected and the concentration of the
analyte reliably and accurately determined.
2. "Matrix" means the specific
components of a sample, other than the analyte being tested
for.
3.2. "Mid-level standard"
means a standard that is between the highest concentration and lowest
concentration of standards containing the same substances that are used as a
reference when testing for the concentration of an analyte.
4.3.
"Response factor" means the ratio between a signal produced by an analyte
relative to a signal produced by an internal standard at a specific
concentration.
5.4. "Retention time"
means the length of time taken by an analyte to pass through a chromatography
column.
6.5. "Standard" means a
sample of known concentration and containing specific substances that is used
as a reference when testing for the concentration of an analyte.
B. To perform laboratory testing
using chemical analytical methods for any of the analytes in Table 3.1, a
laboratory may use:
1. An established national
or international chemical method; or
2. A laboratory-developed method that was
validated according to:
a. AOAC - Appendix K:
Guidelines for Dietary Supplements and Botanicals, 2013, which is incorporated
by reference, includes no future editions or amendments, and is available at
http://www.eoma.aoac.org/app_k.pdf;
b. USDA - Guidelines for the Validation of
Chemical Methods for the FDA FVM Program, 2nd Edition, April 2015, which is
incorporated by reference, includes no future editions or amendments, and is
available at
https://www.fda.gov/media/81810/download;
or
C. A technical
laboratory director shall ensure that all instruments and equipment used for
testing medical marijuana or a marijuana product by chemical analytical methods
are:
1. Set up, tuned, and calibrated
according to:
a. Manufacturer's acceptance
criteria, or
b. Criteria validated
according to subsection (B), as applicable;
2. Monitored and maintained according to AOAC
- Guidelines for Laboratories Performing Microbiological and Chemical Analyses
of Food, Dietary Supplements, and Pharmaceuticals, Appendix A: Equipment,
August 2018, which is incorporated by reference, includes no future editions or
amendments, and is available at
https://www.aoac.org/aoac-accreditation-guidelines-for-laboratories-alacc;
and
3. Applicable for the analytes
to be tested.
D. A
technical laboratory director shall ensure that for an initial demonstration of
capability:
1. Before implementing a method or
using a new instrument, at least four replicate reference samples including
each analyte that are to be tested using the method or the instrument are:
a. Spiked into a clean matrix that is similar
to the medical marijuana or marijuana product to be tested with a mid-level
standard;
b. Taken through the
entire sample preparation and analysis process;
c. Have a relative standard deviation of no
more than 20%; and
d. Have an
accuracy that meets the acceptance criteria in subsection (K)(2)(d);
2. Whenever a significant change
to instrumentation or to a standard operating procedure occurs, the laboratory
demonstrates, as specified in subsection (D)(1), that acceptable precision and
bias can still be obtained by the changed conditions; and
3. Whenever a new laboratory agent who will
be performing testing on medical marijuana or marijuana products is being
trained, the laboratory agent demonstrates, as specified in subsection (D)(1),
acceptable precision and bias.
E. For potency testing or testing for
pesticides, fungicides, growth regulators, mycotoxins, or residual solvents, a
technical laboratory director shall ensure that the retention time window for
each analyte is established by using the absolute retention time for each
analyte and internal standard from the calibration verification standard,
prepared according to subsection (H) or (J) as applicable, at the beginning of
the analytical sequence.
1. For establishing the retention
time for an analyte, the retention time is determined by three injections, over
the course of a 72hour period, of a standard ± 20% of, as
applicable:
a. The maximum allowable
concentration in Table 3.1 for the analyte; or
b. The mid-level standard for
potency testing; and
2. The width of the retention
time window for each analyte is defined as ±3 times the standard
deviation of the mean absolute retention time that was established during the
72-hour period or 0.1 minutes, whichever is greater.
F. A technical laboratory director shall
ensure that:
1. The laboratory complies with
the following requirements related to calibration and standards:
a. Except as specified in subsection
(F)(1)(c), a minimum of:
i. Five standards are
used for an average response factor or for a linear model,
ii. Six standards are used for a quadratic
model, and
iii. Seven standards are
used for a cubic model;
b. An X-value of zero is not included as a
calibration point;
c. A calibration
curve for heavy metal testing includes a minimum of three standards and a
calibration blank;
d. One standard
is less than or equal to the limit of quantitation;
e. The maximum allowable concentration in
Table 3.1 for an analyte, with or without dilution, is less than the
concentration of the highest calibration standard for the analyte; and
e. Except as specified in
subsection (F)(1)(f) and as applicable, one standard for each analyte is
± 20% of the:
i. Maximum allowable concentration
in Table 3.1 for the analyte, or
ii. Mid-level standard for potency
testing; and
f. For testing for residual
solvents, either:
i. One standard for each analyte
is ± 20% of the maximum allowable concentration in Table 3.1 for the
analyte; or
ii.f. As applicable, a
standard is created containing a concentration of specific analytes that is a
dilution factor from the maximum allowable concentration in Table 3.1 for the
analyte and is used when performing multiple runs on a sample, with or without
dilution, to cover the range of maximum allowable concentrations in Table 3.1;
g. One standard is above the
maximum allowable concentration in Table 3.1 for an
analyte;
2. The acceptance criteria for testing is one
of the following, as applicable:
a. The
maximum relative standard deviation for the average calibration factor, for an
external calibration model, or the response factor, for an internal calibration
model, is no more than 20%; and
b.
For linear and non-linear calibration models, the coefficient of determination
(r2) is greater than or equal to 0.990 with no
rounding;
3. For
chromatographic testing methods using internal standards for calibration:
a. The relative retention time of each
analyte to the internal calibration standard is within 0.06 units;
b. The areas of the peaks for the internal
standards in any sample are between 50 and 200% of the area of the peak of a
mid-level standard used for calibration; and
c. The internal standards:
i. Have retention times similar to the
analytes being tested for,
ii. Do
not interfere with any of the analytes, and
iii. Have similar chemical properties as the
analytes being tested for;
4. For methods testing for heavy metals using
internal standards, the internal standards:
a. Are appropriate for the analyte,
and
b. Do not interfere with any of
the analytes;
5. When
using a selective ion monitoring technique for data gathering, the integration
window includes the entire analyte peak; and
6. All standards included for calibration
that are below the limit of quantitation have a signal-to-noise ratio of at
least 3:1 according to ASTM E685-93, Standard Practice for Testing
Fixed-Wavelength Photometric Detectors Used in Liquid Chromatography (2013),
which is incorporated by reference, includes no future editions or amendments,
and is available at
https://web-store.ansi.org/Standards/ASTM/astme685932013.
G. To obtain an acceptable calibration, a
technical laboratory director, for each calibration event:
1. May use any of the following options:
a. Perform instrument maintenance to optimize
analyte responses, as long as all resulting calibration models meet the
acceptance criteria appropriate for the analyte;
b. If the problem appears to be associated
with a single standard:
i. Reanalyze that one
standard, at the time of calibration and before any samples are analyzed, to
rule out problems due to random error: and
ii. Recalculate and reevaluate the standard
against the acceptance criteria;
c. Narrow the calibration range by replacing
one or more of the calibration standards at the upper or lower ends of the
curve;
d. Narrow the calibration
range by removing data points from either extreme end of the range and
recalculating the calibration function; or
e. Perform a new initial calibration
according to subsection (F); and
2. May not:
a. Remove data points from within a
calibration range while still retaining the extreme ends of the calibration
range,
b. Use non-linear
calibrations to compensate for detector saturation or to avoid proper
instrument maintenance;
c. Use
multiple points at the same calibration level if not also being done for all
quality control samples, such as a sample required in subsection (K), and
samples accepted for testing; or
d.
Include calibration data from another calibration that was run at a different
time.
H. A
technical laboratory director shall ensure that, during each calibration event
for initial calibration verification:
1.
Standards are prepared either from a different source or from a different lot
of standards from the same source than the source from which the initial
calibration standards specified in subsection (F)(1) were obtained and must:
a. Be
i. The maximum allowable
concentrations for an analyte in Table 3.1,
ii. According to subsection
(F)(1)(f)(ii), or
iii. a mid-level
standard; and
b. Contain all
analytes being reported to comply with
R9-17-317(A)(5) ;
and
2. The following
acceptance criteria are used:
a. For potency
testing, 80 to 120% recovery of true value;
b. For testing for pesticides, fungicides,
growth regulators, mycotoxins, or residual solvents other than butanes, 70 to
130% recovery of the true value;
c.
For butanes, 60 to 140% recovery of the true value; and
c.d. For
heavy metal testing, 90 to 110% recovery of the true
value.
I. A
technical laboratory director shall ensure that for the limit of quantitation:
1. The limit of quantitation is initially
verified by the analysis of at least seven replicate samples, spiked with all
analytes at the limit of quantitation, and processed through all preparation
and analysis steps for each method;
2. The signal-to-noise ratio of the replicate
samples in subsection (I)(1) is at least 5:1 according to ASTM E685-93,
Standard Practice for Testing Fixed-Wavelength Photometric Detectors Used in
Liquid Chromatography (2013), which is incorporated by reference, includes no
future editions or amendments, and is available at
https://webstore.ansi.org/Standards/ASTM/astme685932013;
3. The mean recovery of the replicate samples
in subsection (I)(1) is:
a. For potency
testing, ± 20% of the true value;
b. For testing for pesticides, fungicides,
growth regulators, mycotoxins, or residual solvents, ± 50% of the true
value; and
c. For heavy metal
testing, ± 35% of the true value;
4. The relative standard deviation of the
replicate samples in subsection (I)(1) is less than 20%;
5. The limit of quantitation is, as
applicable, no greater than:
a. Half the
maximum allowable concentrations for an analyte in Table 3.1;
b. For chlorfenapyr, cyfluthrin, or
cypermethrin, the maximum allowable concentrations for the analyte in Table
3.1; or
c. 1.0 mg/g for each
analyte for potency testing;
6. Any changes to specific sample amounts,
dilutions, or volumes employed are reflected in the limit of quantitation
stated on a sample report;
7. The
signal-to-noise ratio in subsection (I)(2) is reverified each time the
instrument used for testing is calibrated; and
8. Documentation of the current limit of
quantitation is maintained for each analyte, matrix, and instrument.
J. Except as provided in
subsection (P), a technical laboratory director shall ensure that for batch
analysis:
1. Continuing calibration
verification standards:
a. Are prepared and
spiked with a mid-level concentration of all analytes from the same calibration
standard source used to prepare the standards specified in subsection (F)(1);
and
i. Initially, with a concentration
± 20% of, as applicable, the maximum allowable concentration for an
analyte in Table 3.1, according to subsection (F)(1)(f)(ii), or the mid-level
standard for potency testing for all analytes being reported to comply with
R9-17-317(A)(5) ; and
ii. Subsequently, with a
concentration at or between the highest concentration and lowest concentration
of standards for the analytes in the batch;
b. Have the following acceptance criteria:
i. For potency testing, 80 - 120% recovery of
true value;
ii. For testing for
pesticides, fungicides, growth regulators, or mycotoxins, or residual solvents
other than butanes, 70 - 130% recovery of the true value;
iii. For butanes, 60 - 140% recovery of the
true value; and
iii.iv. For heavy metal
testing, 90 - 110% recovery of the true value;
2. If internal standards are used in
continuing calibration verification, the acceptability criteria of the internal
standards is determined as follows:
a. For
testing for pesticides, fungicides, growth regulators, mycotoxins, or residual
solvents by mass spectrometry, if the area of the peak for an internal standard
is different by a factor of two from the area of the respective standard in
subsection (F)(1)(e), for the most recent initial calibration sequence,
according to subsection (F):
i. The mass
spectrometer is inspected for malfunctions and corrected, and
ii. Reanalysis of the continuing calibration
verification meets acceptance criteria in subsection (J)(1)(b)(ii) before any
samples are tested; and
b. For heavy metal testing:
i. The intensity of an internal standard is
monitored for each analysis to ensure that the intensity does not vary by more
than ± 30%, with respect to the intensity during the initial calibration
in subsection (F); and
ii. If the
intensity of an internal standard is outside the range also observed in the
calibration blank required in subsection (F)(1)(c):
(1) Testing is stopped until the problem is
corrected, the instrument is recalibrated, and the new calibration is
verified;
(2) Reanalysis of the
continuing calibration verification meets acceptance criteria in subsection
(J)(1)(b)(iii) before any samples are tested; and
(3) The affected samples are retested;
and
3. The frequency of continuing calibration
verification is as follows:
a. For testing by
a method other than mass spectrometry:
i. At
the beginning of the test;
ii.
After every 20 samples, not counting a quality control sample, such as a sample
required in subsection (K); and
iii. At the end of the test; and
b. For testing by mass
spectrometry:
i. At the beginning of the
testing,
ii. After every 12 hours
of running, and
iii. At the end of
the run.
K. Except as provided in subsection (P), a
technical laboratory director shall ensure that for batch analysis, which may
contain no more than 20 samples accepted for testing:
1. A method blank, with a matrix similar to
each type of sample matrix to be tested within the batch:
a. Contains the same internal standards as
the samples in the batch,
b. Is
prepared and tested with each batch, and
c. Produces results below the limit of
quantitation;
2. Except
as provided in subsection (R), a laboratory control sample and duplicate, with
a matrix similar to each type of sample matrix to be tested within the batch:
a. Are prepared
i. The maximum allowable
concentrations for an analyte in Table 3.1,
ii. According to subsection
(F)(1)(f)(ii), or
iii. with a mid-level
standard;
b. Are spiked
before extraction;
b.c. Are carried through
all stages of sample preparation and included with each analytical batch;
and
c.d. Have either the
following acceptance criteria:
i. For potency
testing, 80 - 120% recovery of true value;
ii. Except as specified in
subsection (K)(2)(c)(iii), for testing for pesticides, fungicides, or growth
regulators, 70 - 130% recovery of the true value;
iii. For Acequinocyl, Bifenthrin,
Fludioxomil, Hexythiazox, Imazalil, Naled, Imidacloprid, and Spiroxamine, 70 -
130% recovery of the true value or according to control limits derived
according to R9-17-404.05(B)(10) ;
iv. For residual solvents except
propane and butane, 70 - 130% recovery of the true value;
v. For propane or butane, 60 -
140% recovery of the true value;
vi.ii. For pesticides,
fungicides, growth regulators, mycotoxins, or residual solvents other than
butanes, 70 - 130% recovery of the true value or according to control limits
derived according to
R9-17-404.05(B)(10)
;
iii. For butanes, 60 - 140%
recovery of the true value or acceptance criteria within statistically derived
limits developed by the laboratory; and
vii.iv. For heavy metal
testing, 80 - 120% recovery of the true value or acceptance criteria within
statistically derived limits developed by the laboratory;
3. The relative percent difference
for the laboratory control sample and duplicate for each analyte, calculated on
the basis of concentration or amount, is no more than 20%; and
4. For all new matrix types to be tested, a
matrix spike derived from a dispensary submitted sample:
a. Is prepared for each analyte in Table 3.1
with a mid-level standard;
b. Is
carried through all stages of sample preparation and included with each
analytical batch of up to 20 samples for each matrix type; and
c. Has either the following acceptance
criteria or acceptance criteria within statistically derived limits developed
by the laboratory:
i. For potency testing, 80
- 120% recovery of true value or according to control limits derived according
to
R9-17-404.05(B)(10)
;
ii. For testing for pesticides,
fungicides, growth regulators, mycotoxins, or residual solvents, 70 - 130%
recovery of the true value or according to control limits derived according to
R9-17-404.05(B)(10)
; and
iii. For heavy metal testing,
75 - 125% recovery of the true value.
L. A technical laboratory director shall
ensure that:
1. Except as provided in
subsection (P), for potency testing or testing for pesticides, fungicides,
growth regulators, mycotoxins, or residual solvents by mass spectrometry, the
relative intensities of the characteristic ions agrees within 30% of the
relative intensities of these ions in the reference spectrum; and
2. For heavy metal testing, the intensity of
each internal standard is monitored for each analysis to ensure that the
intensity does not vary more than ±30%, with respect to the intensity of
the internal standard during the initial calibration specified in subsection
(F).
M. A technical
laboratory director shall ensure that:
1. In
testing, by a method other than mass spectrometry, the resolution of
chromatographic peaks is maintained so that the height of the valley between
two chromatographic peaks is less than 50% of the lower peak height;
and
2. For testing by mass
spectrometry methods, the resolution of chromatographic peaks is maintained so
that the height of the valley between two chromatographic peaks is less than
50% of the average of the two peak heights.
N. A technical laboratory director shall
ensure that confirmation for testing for pesticides, fungicides, growth
regulators, or residual solvents by a method other than mass spectrometry:
1. Is performed using:
a. A second column:
i. That has a stationary phase dissimilar to
the stationary phase in the primary column, and
ii. From which the analyte is eluted in a
different order than from the primary column;
b. A different instrument type, such as gas
chromatography followed by mass spectrometry;
c. Gas chromatography with two different
types of detectors; or
d. Other
recognized confirmation techniques;
2. Meets the applicable criteria in
subsections (D) through (M); and
3.
Includes as part of the confirmation of the analyte:
a. An evaluation of the agreement of the
quantitative values of the results from both methods of testing; and
b. Determination of the relative percent
difference between the values.
O. If the relative percent difference between
the values obtained according to subsection (N) is more than 40%, a technical
laboratory director shall ensure that:
1. The
chromatograms are checked to see if an obviously overlapping peak is causing an
erroneously high result, and the chromato-graphic conditions are reviewed;
and
2. Either:
a. If a problem is found with one of the
tests, the result from the other test is reported; and
b. If there is no evidence of a
chromatographic problem, the higher result is reported.
P. A technical laboratory director
may release testing results that are scientifically valid and defensible,
according to
R9-17-404.06(B)(3),
with the following data qualifier notations if:
1. The target analyte detected in the
calibration blank required in subsection (F)(1)(c) or the method blank
specified in subsection (K)(1) is at or above the limit of quantitation, but
the sample result:
a. For potency testing, is
below the limit of quantitation - B1; or
b. When testing for pesticides, fungicides,
growth regulators, mycotoxins, heavy metals, or residual solvents, is below the
maximum allowable concentration in Table 3.1 for the analyte - B2;
2. The limit of quantitation and
the sample results were adjusted to reflect sample dilution - D1;
3. The relative intensity of a characteristic
ion in a sample analyte exceeded the acceptance criteria in subsection (L)(1)
with respect to the reference spectra, indicating interference - I1;
4. When testing for pesticides, fungicides,
growth regulators, mycotoxins, heavy metals, or residual solvents, the percent
recovery of a laboratory control sample is greater than the acceptance limits
in subsection (K)(2)(c), but the sample's target analytes were not detected
above the maximum allowable concentrations in Table 3.1 for the analytes in the
sample - L1;
5. The recovery from
the matrix spike in subsection (K)(4) was:
a.
High, but the recovery from the laboratory control sample in subsection (K)(2)
was within acceptance criteria - M1,
b. Low, but the recovery from the laboratory
control sample in subsection (K)(2) was within acceptance criteria - M2,
or
c. Unusable because the analyte
concentration was disproportionate to the spike level, but the recovery from
the laboratory control sample in subsection (K)(2) was within acceptance
criteria - M3;
6. The
analysis of a spiked sample required a dilution such that the spike recovery
calculation does not provide useful information, but the recovery from the
associated laboratory control sample in subsection (K)(2) was within acceptance
criteria - M4;
7. The analyte
concentration was determined by the method of standard addition, in which the
standard is added directly to the ali-quots of the analyzed sample -
M5;
8. A description of the
variance is described in the final report of testing according to
R9-17-404.06(B)(3)(d)(ii)
- N1;
9. The relative percent
difference for the laboratory control sample and duplicate exceeded the limit
in subsection (K)(3), but the recovery in subsection (K)(2) was within
acceptance criteria - R1;
10. The
relative percent difference for a sample and duplicate exceeded the limit in
subsection (O) - R2; or
11. The
recovery from initial or continuing calibration verification standards is
greater than the acceptance limits in subsection (H)(2) or (J)(1)(b) as
applicable, but the sample's target analytes were not detected above the
maximum allowable concentrations in Table 3.1 for the analytes in the sample -
V1.
Q. A technical
laboratory director shall include in the final report of testing, according to
R9-17-404.06(B)(3)(d)(iii),
the following data qualifier notations if:
1.
Sample integrity was not maintained - Q1;
2. The sample is heterogeneous, and sample
homogeneity could not be readily achieved using routine laboratory practices -
Q2; or
3. Testing result is for
informational purposes only and cannot be used to satisfy dispensary testing
requirements in
R9-17-317.01(A) or
labeling requirements in
R9-17-317 - Q3.
R. For batch analysis of samples to determine
potency, a technical laboratory director may check precision by using either a
duplicate laboratory control sample or a duplicate sample prepared from the
medical marijuana or marijuana product being tested, according to requirements
in subsections (K)(2) and (3).
S. A
technical laboratory director shall ensure that the reporting units for:
1. Pesticides, fungicides, growth regulators,
heavy metals, or residual solvents are in parts per million (ppm);
2.3. Potency are:
a. In either:
i. Percent (w/w) relative to the bulk plant
material or marijuana product, as applicable; or
ii. Number of milligrams per designated unit;
and
b. For:
i. Total tetrahydrocannabinol, the sum of
tetrahydrocannabinolic acid (THC-A), multiplied by 0.877, and
delta-9-tetra-hydrocannabinol ([DELTA]9-THC); and
ii. Total cannabidiol, the sum of
cannabidiolic acid (CBD-A), multiplied by 0.877, and cannabidiol
(CBD).