Fla. Admin. Code Ann. R. 64E-5.502 - General Requirements
(1)
Administrative Controls.
(a) Registrant. The
registrant shall be responsible for directing the operation of the x-ray
systems which are subject to registration as described in Rule
64E-5.511,
F.A.C. The registrant or the registrant's agent shall assure that the following
requirements are met in the operation of the x-ray system.
1. Any x-ray system which does not meet the
provisions of these regulations shall not be operated for diagnostic or
therapeutic purposes unless the Department determines that such operation will
not endanger the public health, safety and welfare.
2. An x-ray system operator shall receive
instruction on and be competent in the safe use of the x-ray system. A medical
x-ray system operator shall also be certified or authorized in accordance with
Chapter 468, Part IV, F.S.
3. A
chart shall be provided in the vicinity of the diagnostic x-ray system's
control panel, which specifies techniques and procedures to be used for all
examinations which are performed by that system.
4. Except for patients who cannot be moved
out of the room, only the staff and ancillary personnel required for the
medical procedure or training shall be in the room during the radiographic
exposure. Other than the patient being examined:
a. All individuals shall be positioned such
that no part of the body will be struck by the useful beam unless protected by
0.5 millimeter lead equivalent.
b.
Staff and ancillary personnel shall be protected from the direct scatter
radiation by protective aprons or whole body protective barriers of not less
than 0.25 millimeter lead equivalent.
c. Other patients who cannot be removed from
the room shall be protected from the direct scatter radiation by whole body
protective barriers of 0.25 millimeter lead equivalent or shall be so
positioned that the nearest portion of the body is at least 2 meters from both
the tube head and the nearest edge of the useful beam.
d. When a portion of the body of any staff or
ancillary personnel is potentially subjected to stray radiation which could
result in that individual receiving one-fourth of the maximum permissible dose
as defined in Part III, additional protective devices may be required by the
Department.
5. Gonad
shields of not less than 0.25 millimeter lead equivalent shall be used for
patients who have not passed the reproductive age, during radiographic
procedures in which the gonads are in the useful beam, except for cases in
which this would interfere with the diagnostic procedure.
6. Individuals shall not be exposed to the
useful beam except for healing arts purposes and unless such exposure has been
authorized by a licensed practitioner of the healing arts. This provision
specifically prohibits the following:
a.
Exposure of an individual for training, demonstration or other purposes unless
there are also healing arts requirements and a proper prescription has been
provided.
b. Exposure of an
individual for the purpose of a healing arts self-referral program except when
authorized by subparagraph 64E-5.502(1)(a) 1., F.A.C.
c. Advertisement of free x-ray examinations
unless the advertisement states that a determination of need will be made prior
to the x-ray examination.
7. An x-ray system is exempt from the
requirements of subparagraph 64E-5.502(1)(a) 6., F.A.C., only if:
a. The system is a stationary, non-mobile
system installed and used only in a jail or penal institution;
b. The system is used only on legal detainees
in a jail or penal institution, and never on detainees' family members,
children, institution employees, contractors, visitors, the public, or any
other persons;
c. The system is
manufactured, maintained, and operated solely for security screening purposes
in strict compliance with, and fully according to, the most restrictive
standards found in ANSI/HPS N43.17-2009, "Radiation Safety for Personnel
Security Screening Systems Using X-Ray or Gamma Radiation," which is herein
incorporated by reference. This ANSI publication may be examined and inspected
at the Florida Department of Health, Bureau of Radiation Control at Building
4042 Bald Cypress Way, Suite 210, Tallahassee, Florida 32399-1741, and at the
Florida Department of State, Room 701, The Capitol, Tallahassee, Florida
32399-0250, and is available from the American National Standards Institute,
Inc., at http://www.ansi.org. The
agency has determined that posting the publication on the internet for purposes
of public inspection and examination would constitute a violation of federal
copyright law; and,
d. The
institution tracks radiation dose for each detainee, as specified by ANSI/HPS
N43.17-2009, "Radiation Safety for Personnel Security Screening Systems Using
X-Ray or Gamma Radiation," to ensure the detainee does not exceed the
recommended dose limit.
8. When a patient or film must be provided
with auxiliary support during a radiation exposure:
a. Mechanical holding devices shall be used
when the technique permits;
b.
Written safety procedures shall be available to indicate the requirements for
selecting a holder, list the individual projections where holding devices
cannot be used and describe the procedure the holder shall follow;
c. The human holder shall be protected as
required by subparagraph (1)(a)4., above; and,
d. No individual shall be used routinely to
hold film or patients.
9. Exposure Procedures Designed to Minimize
Patient and Personal Exposure.
a. The speed
of film or screen and film combinations shall be the fastest speed consistent
with the diagnostic objectives of the examination.
b. The radiation exposure to the patient
shall be the minimum exposure required to produce images of good diagnostic
quality.
c. Portable or mobile
equipment shall be used only for examinations where it is impractical to
transfer the patient to a stationary radiographic installation.
d. X-ray systems subject to Rule
64E-5.505,
F.A.C., shall not be utilized in procedures where the source to patient
distance is less than 30 centimeters.
e. A person shall not perform fluoroscopic
imaging or otherwise expose a human to x-rays from a fluoroscopic system unless
the person is a:
(I) Licensed practitioner as
that term is defined in Section
468.301, F.S.;
or
(II) Certified radiologist
assistant practicing in accordance with the requirements of Chapter 468, Part
IV, F.S.; or
(III) Certified
general radiographer practicing in accordance with the requirements of Chapter
468, Part IV, F.S.; and
(A) The general
radiographer has been trained and authorized in writing by the licensed
practitioner in charge to perform the specified imaging; and
(B) The specified imaging does not rely upon
the general radiographer to provide any diagnostic interpretation, or to
determine suspicious areas for additional imaging, or to otherwise modify the
scope of authorization for the imaging; and
(C) The specified imaging is designed to
prevent or reduce exposure to patients by facilitating proper location and
positioning for the authorized radiographic imaging.
10. Personnel
Monitoring. All individuals who are associated with the operation of an x-ray
system are subject to the occupational exposure limits and the requirements for
the determination of the doses stated in Rules
64E-5.304
and
64E-5.308,
F.A.C. In addition, when protective clothing or devices are worn on portions of
the body and a personnel monitoring device is required, at least one such
device shall be utilized as follows:
a. When
a protective apron is worn, the monitoring device shall be worn at the collar
outside of the apron.
b. The dose
to the whole body shall be recorded in the records required by Rule
64E-5.339,
F.A.C. If more than one device is used and a record is made of the data, each
dose shall be identified with the area where the device was worn on the
body.
11. Healing arts
self-referral. Only healing arts self-referral programs for mammography
screening will be authorized by the Department.
(b) Information and Maintenance Records and
Associated Information. The registrant shall maintain at least the following
information for each x-ray system:
1. Tube
rating charts and cooling curves.
2. Record of surveys, calibrations,
maintenance, modifications from the original schematics and drawings performed
on the x-ray machine along with the names of persons who performed the
service.
3. A copy of all
correspondence with the Department regarding each x-ray system.
4. An x-ray log containing the patient's
name, the type of examination and the dates the examinations were performed.
When the patient or film must be provided with human auxiliary support, the
name of the human holder shall be recorded.
(c) Maintenance of X-ray Equipment. X-ray
systems and accessory devices shall be maintained in good working condition,
both mechanically and electrically, so that the clinical objectives may be
fulfilled without risk of unproductive exposure due to equipment failure or
malfunction.
(2)
Shielding.
(a) Each x-ray facility shall have
primary and secondary protective barriers as needed to assure that an
individual will not receive a radiation dose in excess of the limits specified
in Part III of these regulations.
(b) Structural shielding in walls and other
vertical barriers required for personnel protection shall extend without breach
from the floor to a height of at least 7 feet (2.1 m).
(c) Doors, door frames, windows and window
frames shall have the same lead equivalent shielding as that required in the
wall or other barrier in which they are installed.
(d) In computation of protective barrier
requirements, the maximum anticipated workload, use factors, occupancy factors
and the potential for radiation exposure from other sources shall be taken into
consideration.
(e) Prior to
construction, the floor plans and equipment arrangement of all new
installations, or modifications of existing installations, utilizing x-ray
energies of 200 keV and above for diagnostic or therapeutic purposes shall be
submitted to the Department for review and approval.
1. The plans shall show, as a minimum, the
following:
a. The normal location of the
x-ray system's radiation port; the port's travel and traverse limits; general
direction of the useful beam; locations of any windows and doors; the location
of the operator's booth; and the location of the x-ray control panel.
b. The structural composition and thickness
or lead equivalent of all walls, doors, partitions, floor and ceiling of the
room concerned.
c. The dimensions
of the room concerned.
d. The type
of occupancy of all adjacent areas inclusive of space above and below the room
concerned. If there is an exterior wall, the distance to the closest area where
it is likely that individuals may be present.
e. The make and model of the x-ray equipment
and the maximum technique factors.
f. The type of examinations or treatments
which will be performed with the equipment.
2. Information on the anticipated maximum
workload of the x-ray system.
3. If
the services of a qualified person have been utilized to determine the
shielding requirements, a copy of the report, including all basic assumptions
used, shall be submitted with the plans.
(3) X-ray Film Processing Facilities and
Practices.
(a) Processing Facilities. Each
installation using a radiographic x-ray system shall provide suitable equipment
for handling and processing radiographic film in accordance with the following
provisions:
1. The area in which undeveloped
films are handled for processing shall be devoid of light with the exception of
light in the wave lengths having no significant effect on the radiographic
film.
2. Film pass boxes, if
provided, shall be so constructed as to exclude light when film is placed in or
removed from the boxes, and shall incorporate adequate shielding to prevent
exposure of undeveloped film to stray radiation.
3. Darkrooms used by more than one individual
shall be provided a positive method to prevent accidental entry while
undeveloped films are being handled or processed.
4. Where film is developed manually,
a. At least one tri-sectional tank made of
mechanically rigid, corrosion resistant material shall be utilized;
and
b. The temperature of each
solution shall be maintained within the range of 60º F to 80º F
(16º C to 27º C). Film shall be developed in accordance with the
time-temperature relationships specified by the film manufacturer, or, in the
absence of such recommendations by the film manufacturer, with the following
time-temperature chart:
Time-Temperature Chart | ||
Thermometer Reading | Minimum Developing | |
(Degrees) | Time (Minutes) | |
C | F | |
26.7 | 80 | 2 |
26.1 | 79 | 2 |
25.6 | 78 | 2 1/2 |
25.0 | 77 | 2 1/2 |
24.4 | 76 | 3 |
23.9 | 75 | 3 |
23.3 | 74 | 3 1/2 |
22.8 | 73 | 3 1/2 |
22.2 | 72 | 4 |
21.7 | 71 | 4 |
21.1 | 70 | 4 1/2 |
20.6 | 69 | 4 1/2 |
20.0 | 68 | 5 |
19.4 | 67 | 5 1/2 |
18.9 | 66 | 5 1/2 |
18.3 | 65 | 6 |
17.8 | 64 | 6 1/2 |
17.2 | 63 | 7 |
16.7 | 62 | 8 |
16.1 | 61 | 8 1/2 |
15.6 | 60 | 9 1/2 |
c.
Devices shall be utilized which will:
(I)
Indicate the actual temperature of the developer; and
(II) Signal the passage of a preset time as
short as 2 minutes.
(b) Precautionary Practices.
1. Film shall be stored in a cool, dry place
and shall be protected from exposure to stray radiation. Film in open packages
shall be stored in a light tight container.
2. Film cassettes and intensifying screens
shall be inspected periodically and shall be cleaned and replaced as necessary
to best assure radiographs of good diagnostic quality.
3. Outdated x-ray film shall not be used for
human diagnostic radiographs, unless the film has been stored in accordance
with the manufacturer's recommendations and a sample of the film passes a
sensitometric test for normal ranges of base fog and speed.
4. Film developing solutions shall be
prepared in accordance with the directions given by the manufacturer, and shall
be maintained in strength by replenishment or renewal so that full development
is accomplished within the time specified by the manufacturer.
5. Safe light and darkroom fog shall be such
that, when a radiographic film is exposed to radiation to achieve a density of
1.0 and is exposed for 1 minute on any darkroom working surface, the film shall
not have a density change greater than 0.1.
(c) Automatic Processors and Other Closed
Processing Systems. Preventive maintenance shall be performed on the unit,
except for extended periods of nonuse, on a frequency basis which is not less
than that schedule recommended by the manufacturer. In the event that no
schedule is available from the manufacturer, a maintenance schedule shall be
established which will preserve good diagnostic film quality.
(d) Radiographic Film Quality.
1. Developed radiographs of patients or
phantoms shall have an optical density of 0.5 to 2.0 in the area of clinical
interest to allow for diagnostic interpretation of the image, unless justified
due to special circumstances. Radiographs which provide the necessary
diagnostic information shall not be repeated for the sole purpose of meeting
the stated density range.
2.
Radiographic film used for diagnostic purposes shall be free from light fog and
artifacts.
Notes
Rulemaking Authority 404.051, 404.081, 404.141, 404.22 FS. Law Implemented 404.051, 404.081, 404.141, 404.22, 468.302 FS.
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.