Fla. Admin. Code Ann. R. 64E-5.504 - Fluoroscopic X-Ray Systems
All fluoroscopic x-ray systems shall meet the following requirements:
(1) Limitation of the
Useful Beam.
(a) The fluoroscopic tube shall
not produce x-rays unless the primary protective barrier is in position to
intercept the entire cross section of the useful beam.
(b) A means shall be provided between the
x-ray source and the patient for stepless adjustment of the size of the x-ray
field.
(c) With the collimating
shutters adjusted to the closed position, the minimum field size at the maximum
SID shall not be greater than 5 x 5 centimeters when measured at the point
where the beam enters the patient.
(d) Limitation to the Imaging Surface.
1. The x-ray field produced by
nonimage-intensified fluoroscopic equipment shall not extend beyond the useable
area of the largest image receptor at any SID.
2. The longitudinal and transverse dimensions
of the x-ray field produced by image-intensified fluoroscopic equipment shall
not extend beyond the corresponding dimensions of the visible area of the image
receptor by more than 3 percent of the SID in either dimension in the plane of
the image receptor and the sum of the excess shall be no greater than 4 percent
of the SID. If the collimation is automatically accomplished, the x-ray field
dimension criteria above shall apply to all film sizes and portions thereof
that the spot film device accommodates and to the dimensions of the input
phosphor, as appropriate. If collimation is not automatic, the x-ray field
dimension criteria shall apply to the useful area of the input
phosphor.
3. Compliance shall be
determined with the beam axis perpendicular to the plane of the image receptor.
For rectangular x-ray fields used with circular image receptors, the error in
alignment shall be determined along the length and width dimensions of the
x-ray field which passes through the center of the visible area of the image
receptor.
4. The center of the
x-ray field in the plane of the image receptor shall be aligned with the center
of the selected portion of the image receptor to within 2 percent of the
SID.
5. Adjustable automatic and
manual collimators shall operate smoothly throughout the entire range of
use.
6. For fluoroscopic systems
with spot film capability, means shall be provided for adjustment of the x-ray
field size in the plane of the film to a size smaller than the selected portion
of the film.
(e) The
requirements of paragraphs (1)(b) and (c), above, are not applicable to mobile
fluoroscopic systems.
(2) Activation of the Fluoroscopic Tube. A
control of the dead-man type shall be incorporated into each fluoroscopic
system such that x-ray production will be terminated at any time pressure is
released from the switch except during the recording of serial fluoroscopic
images with equipment in which means have been provided to permit completion of
any single exposure of the series in progress.
(3) Allowable Entrance Exposure Rate Limits
for Fluoroscopic Equipment.
(a) Fluoroscopic
equipment manufactured after June, 1995, operable at any combination of tube
potential and current that results in an exposure rate greater than 5 roentgens
(1.29 x 10-3 C per kg) per minute at the point where
the center of the useful beam enters the patient shall be equipped with
automatic exposure control. Provision for manual selection of technique factors
can be provided.
(b) Fluoroscopic
equipment shall not be operable at any combination of tube potential and
current that will result in an exposure rate in excess of 10 roentgens (2.58 x
10-3 C per kg) per minute at the point where the
center of the useful beam enters the patient except:
1. During the recording of images from an
x-ray image-intensifier tube using photographic film or a video camera when the
x-ray source is operated in a pulsed mode.
2. When an optional high-level control is
activated. When the high-level control is activated, the equipment shall not be
operable at any combination of tube potential and current that will result in
an exposure rate in excess of 20 roentgens (5.16 x
10-3 C per kg) per minute at the point where the
center of the useful beam enters the patient. Special means to activate
high-level controls shall be required. The high-level control shall only be
operable when continuous manual activation is provided by the
operator.
(c) Special
means to activate high level controls such as additional pressure applied
continuously by the operator shall be required to avoid accidental
use.
(d) A continuous signal
audible to the fluoroscopist shall indicate when the high level control is
being employed.
(e) Compliance with
the dose limits will be determined as follows:
1. Movable grids and compression devices will
be removed from the useful beam during the measurement.
2. The fluoroscope's radiation output will be
maximized.
a. Systems with automatic exposure
controls such as automatic brightness control will have sufficient lead or lead
equivalent materials placed in the useful beam to produce the maximum
output.
b. Systems without
automatic exposure controls or systems with a manual mode in addition to
automatic exposure control modes will have the current and potential set to
produce the maximum output. Attenuating material will be placed in the useful
beam to protect the imaging system. If the registrant has a written radiation
protection program restricting the range of current and potential the tests
will be performed within the range of allowed values.
c. Patient support device height and SID,
where adjustable, will be varied to produce the maximum output. If the
registrant has a written radiation protection program restricting the range of
patient support device heights or SIDs the tests will be performed within the
range of allowed values.
3. The exposure rate will be measured at the
following points on the centerline of the beam unless the specified geometry is
prohibited by a written radiation protection program.
a. At least one centimeter above the patient
support device and corrected for distance to show the actual entrance exposure
rate at the top surface of the patient support device for:
(I) Fluoroscopes where the x-ray tube is
fixed under the patient support device.
(II) C-arm systems or stationary c-arm
fluoroscopes where the x-ray tube can be rotated under the patient support
device. The x-ray tube will be positioned as close to the patient support
device as possible.
b.
At 30 centimeters above the patient support device with the end of the
beam-limiting device or spacer assembly positioned as close as possible to the
point of measurement for:
(I) Fluoroscopes
where the x-ray tube is fixed above the patient support device.
(II) C-arm systems or stationary c-arm
fluoroscopes where the x-ray tube can be rotated above the patient support
device.
c. At a point 15
centimeters laterally from the centerline of the patient support device or from
the centerline of the patient if the registrant has a written radiation
protection program specifying placement of the patient not on the centerline of
the patient support device in the direction of the x-ray tube with the input
surface of the fluoroscopic imaging assembly positioned as close to the edge of
the patient support device as possible but no closer than 15 cm for:
(I) Fluoroscopes where the x-ray tube is
fixed laterally to the patient support device.
(II) C-arm systems or stationary c-arm
fluoroscopes where the x-ray tube can be rotated lateral to the patient support
device.
d. At 30
centimeters from the input surface of the fluoroscopic imaging assembly,
provided that the end of the beam-limiting device or spacer is no closer than
30 centimeters from the input surface of the fluoroscopic imaging assembly, for
mobile c-arm fluoroscopes. Spacers or other attachments normally used can not
be removed to allow measuring from a point closer to the actual input surface.
(f) Periodic
Measurement of Entrance Exposure Rates. The entrance exposure rate shall be
measured before use on humans after the completion of any initial or subsequent
installation and after any maintenance of the system that might affect the
exposure rate.
(g) For
cinefluoroscopy, the maximum exposure at the face of the input phosphor with
the grid removed and with an attenuation block in the beam shall not exceed 40
microroentgens (0.010 µC per kg) per frame. The maximum exposure shall be
measured before use on humans after the completion of any initial or subsequent
installation and after any maintenance of the system which might affect the
maximum exposure.
(4)
Barrier Transmitted Radiation Rate Limits.
(a) The exposure rate due to transmission
through the primary protective barrier and frame assembly with the attenuation
block in the useful beam combined with radiation from the image intensifier if
provided shall not exceed 2 milliroentgens (0.516 µC per kg) per hour at
10 centimeters from any accessible surface of the fluoroscopic imaging assembly
beyond the plane of the image receptor for each roentgen per minute of entrance
exposure rate.
(b) Measuring
Compliance with Barrier Transmission Limits.
1. The exposure rate due to transmission
through the primary protective barrier combined with radiation from the image
intensifier shall be determined by measurements averaged over an area no
greater than 100 square centimeters with no linear dimension greater than 20
centimeters.
2. If the source is
below the tabletop, the measurement shall be made with the input surface of the
fluoroscopic imaging assembly positioned 30 centimeters above the
tabletop.
3. If the source is above
the tabletop and the SID is variable, the measurement shall be made with the
end of the beam limiting device or spacer assembly as close to the tabletop as
it can be placed but not closer than 30 centimeters.
4. Movable grids and compression devices
shall be removed from the useful beam during the measurement.
5. The attenuation block shall be positioned
in the useful beam 10 centimeters toward the input surface of the imaging
assembly from the point at which the entrance exposure rate was
measured.
6. The maximum beam size
shall be used during measurements.
(5) Indication of Potential and Current.
During fluoroscopy and cinefluorography, x-ray tube potential and current shall
be continuously indicated.
(6)
Source-to-Skin Distance. Positive means shall be provided to assure the
source-to-skin distance shall not be less than:
(a) Thirty-eight centimeters on stationary
fluoroscopes installed after January 1, 1977;
(b) Thirty-five and one-half centimeters on
stationary fluoroscopes installed prior to January 1, 1977;
(c) Thirty centimeters on all mobile
fluoroscopes;
(d) Twenty
centimeters for image intensified fluoroscopes used for specific surgical
applications. Written safety procedures must be provided to the operator of the
fluoroscope and precautionary measures followed during the use of this
device;
(e) Nineteen centimeters
for extremity-use-only fluoroscopes;
(f) Ten centimeters for extremity-use-only
fluoroscopes used for specific surgical applications. Written safety procedures
must be provided to the operator of the fluoroscope and precautionary measures
followed during the use of this device.
(7) Fluoroscopic Timer. A cumulative timing
device activated by the fluoroscopic exposure switch shall be provided, the
maximum cumulative time of which shall not exceed 5 minutes without resetting.
The timer shall indicate the passage of the predetermined period of exposure by
an audible signal or termination of the exposure. If such a signal is utilized,
it shall continue while x-rays are produced until the timing device is
reset.
(8) Mobile Fluoroscopes. In
addition to the other requirements of this section, mobile fluoroscopes shall
provide intensified imaging.
(9)
Control of Scatter Radiation.
(a) Fluoroscopic
table designs shall be such that scattered radiation which originates beneath
the tabletop is attenuated by not less than 0.25 millimeters lead equivalent,
and that no unprotected part of any staff or ancillary person's body shall be
exposed to unattenuated scattered radiation.
(b) Fluoroscopic equipment configuration
shall be such that no portion of any staff or ancillary person's body, except
the extremities, shall be exposed to the unattenuated scattered radiation
emanating from above the tabletop unless:
1.
Such person is at least 120 centimeters from the center of the useful beam,
or
2. The radiation has passed
through not less than 0.25 millimeter lead equivalent material.
(c) Exceptions to paragraph
(10)(b), above, may be made in some special procedures where a sterile field
will not permit the use of the normal protective barriers. Where the use of
prefitted sterilized covers for the barriers is practical, the Department shall
not permit such exception.
(10) Photofluorographic Medical X-Ray
Systems.
(a) In addition to other applicable
sections of these regulations, photofluorographic x-ray systems shall conform
with the following requirements:
1. Usage
shall be limited to diagnostic radiography of the lungs and other soft tissues
of the thoracic region.
2.
Personnel monitoring shall be provided for all individuals who operate
photofluorographic apparatus.
3.
The average exposure, including backscatter, for chests measuring 25
centimeters in thickness shall not exceed 100 millirems (1.0 mSv) at the point
where the x-ray beam enters the patient.
(b) Photofluorographic x-ray systems shall
not be installed unless specifically approved by the Department.
(11) Radiation Therapy Simulation
Systems. Radiation therapy simulation systems shall be exempt from all the
requirements of subsections (1), (3), (4), (5) and (8), above, provided that:
(a) Such systems are designed and used in
such a manner that no person other than the patient is in an unprotected area
during periods of time when the system is producing x-rays; and,
(b) Systems that do not meet the requirements
of subsection (8), above, are provided with a means of indicating the
cumulative time that an individual patient has been exposed to x-rays. In such
cases, the timer shall be reset between examinations.
(c) The exposure rate measured at the point
where the center of the useful beam enters the patient shall not exceed 20
roentgens (5.16 mC per kg) per minute, except during the recording of
fluoroscopic images.
(12) For remotely operated fluoroscopic
systems:
(a) The remote control panel shall
be installed so as to require the operator to stand behind a permanent
protective barrier meeting the requirements of paragraph
64E-5.502(2)(a)
-(c), F.A.C. The barrier must be wide enough to prevent the secondary scatter
radiation from striking the operator directly when the machine is operated from
the remote control panel.
(b) The
operator must be able to see and hear the patient when behind the
barrier.
(c) The barrier shall be
constructed of material of sufficient density to meet or exceed the barrier
requirements of sub-subparagraph
64E-5.502(1)(a)
4.b., F.A.C.
Notes
Rulemaking Authority 404.051, 404.22 FS. Law Implemented 404.051, 404.22 FS.
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.