Ill. Admin. Code tit. 20, § 1286.310 - Approved Evidentiary Instrument Operation
The following procedures shall be used to obtain a breath sample to determine a subject's BrAC with an approved evidentiary instrument:
a) Prior to obtaining a
breath analysis reading from a subject, the BAO or another agency employee
shall continuously observe the subject for at least 20 minutes.
1) During the 20 minute observation period
the subject shall be deprived of alcohol and foreign substances and shall not
have vomited.
2) If the subject
vomits during the observation (deprivation) period, the process shall be
started over by having the individual rinse the oral cavity with
water.
3) If the individual
continues to vomit, alternate testing shall be considered.
b) After starting the instrument's breath
test sequence, the BAO will obey instrument prompts. When prompted by the
instrument, the BAO shall direct the subject to blow into the instrument. The
subject shall be directed to keep blowing into the instrument until he or she
has submitted an adequate breath sample. Once an adequate breath sample is
collected, the instrument shall complete the test cycle and print and display
the breath analysis reading.
c) A
breath test shall consist of only one breath analysis reading, based on the
instrument's internal operational calculations.
1) A complete and valid breath analysis
reading is denoted by at least one air blank, one subject breath test reading,
and no breakdown message.
2)
Messages such as "refusal", "insufficient sample", "inadequate sample", etc.,
are not breakdowns or malfunctions. These messages indicate the subject's
failure to adequately complete a requested test or tests.