105 CMR, § 180.395 - Standard - Microbiology: Quality Control
Chemical and biological solutions, reagents, and antisera shall be tested and inspected as specified below for reactivity and deterioration using positive and negative control organisms. The following requirements apply to the specialties of Bacteriology, Mycology, Parasitology, and Virology:
(A) Materials which shall be
tested and inspected on each day of use include:
(1) Media and discs used for antibiotic
susceptibility testing;
(2)
Reagents and solutions used in the performance of conventional biochemical
testing;
(3) Stains (exception -
Gram Stain).
(B)
Materials which shall be tested and inspected each week of use and when a new
vial or container is opened include:
(1)
Antibiotic discs (exception: antibiotic susceptibility discs);
(2) Chemical strips;
(3) Gram Stain materials.
(C) Materials which shall be
tested and inspected each month of use and when a new vial or container is
opened include:
Antisera.
(D) Materials which shall be tested and
inspected when each new batch or shipment is received include:
(1) Tube and plate media;
(2) Gram negative identification
sets.
(E)
Bacteriology and Mycology. Staining materials shall be
tested for intended reactivity by concurrent application to smears of
micro-organisms with predictable staining characteristics. Each batch of medium
shall be tested before or concurrently with use with selected organisms to
confirm required growth characteristics, selectivity, enrichment, and
biochemical response.
(F)
Parasitology. A reference collection of slides,
photographs, or gross specimens of identified parasites shall be available and
used in the laboratory for appropriate comparison with diagnostic specimens. A
calibrated ocular micrometer shall be used for determining the size of ova and
parasites, if size is a critical factor.
(G)
Virology.
Systems for the isolation of viruses and reagents for the identification of
viruses shall be available to cover the viruses for which services are offered.
Records shall be maintained which reflect the systems used and the reactions
observed. In tests for the identification of viruses, controls shall be
employed which will identify erroneous results. If serodiagnostic tests for
virus diseases are performed, requirements for quality control as specified for
serology shall apply.