265 CMR, § 6.01 - Minimal Procedures and Equipment
(1) The following minimum procedures shall be
used in the fitting and selling of hearing instruments:
(a) pure tone audiometric testing by air and
bone conduction pathways through a calibrated system in the appropriate
environment to determine the type and degree of hearing deficiency;
(b) effective masking when
indicated;
(c) appropriate testing
using a calibrated system or other acceptable verification technique in the
appropriate environment to determine client's hearing ability, as measured by
the percentage words the client is able to repeat correctly, the client's
ability to discriminate speech, the client's most comfortable and uncomfortable
loudness levels in decibels, and the best fitting arrangement for maximum
hearing aid benefit;
(d) otoscopic
inspection of the outer ear;
(e) a
pertinent case history; and
(f) a
final fitting ensuring physical and operational comfort of the hearing
aid.
(2)
Mandatory Disclosure. Any person with a 15 dB or
greater difference between air conduction and bone conduction hearing, at three
frequencies of 500hz, 1000hz, and 2000hz in either ear, must be advised of the
possibility of medical correction.
(3) At a minimum, the following required
equipment shall be used:
(a) a wide-range
audiometer which meets the specifications of the American National Standards
Institute for diagnostic audiometers;
(b) a speech audiometer or a master hearing
aid in order to determine the most comfortable listening level and speech
discrimination.
Notes
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.