Mich. Admin. Code R. 333.5417 - Enclosures

Rule 417.

(1) An enclosure shall be a permanent part of the building or equipment. Portable shields shall not be used for permanent installations.
(2) The degree of protection required for an enclosure shall be determined by the workload, the use and occupancy factors, and the tube potential, tube current, mechanical movement, and distance. The design shall be subject to approval by the department.
(3) For conventional fluoroscopy extraneous light that interferes with the fluoroscopic examination shall be eliminated. Dark adaptation normally is not necessary when using image intensifiers.

Notes

Mich. Admin. Code R. 333.5417
2016 MR 10, Eff. 5/25/2016