Mich. Admin. Code R. 333.5719 - Initial and annual medical physicist performance evaluations
Rule 719.
(1) A
medical physicist shall complete an initial performance evaluation of the CT
scanner before use on human patients and annually thereafter.
(2) A calibrated dosimetry system shall be
used to measure the radiation output of a CT scanner. Calibration of the
dosimetry system shall be within the preceding 24 months and shall be traceable
to a national standard as specified in R 333.5012(1).
(3) A performance evaluation should include
the following:
(a) Alignment light
accuracy.
(b) Alignment of table to
gantry.
(c) Table and gantry
tilt.
(d) Slice localization from
scanned projection radiograph.
(e)
Table increment accuracy.
(f) Slice
thickness.
(g) Image quality,
including the following:
(i) High-contrast
resolution.
(ii) Low-contrast
resolution.
(iii) Image
uniformity.
(iv) Noise.
(v) Artifact evaluation.
(h) CT number accuracy and
linearity.
(i) Dosimetry, including
the following:
(i) Dose indicator such as
computed tomography dose index (CTDI).
(ii) Patient radiation dose for
representative examinations.
(j) Safety evaluation, including the
following:
(i) Visual inspection.
(ii) Audible and visual signals.
(iii) Posting requirements.
(iv) Scattered radiation
measurements.
(k) Review
of the ongoing quality control program, including test results and corrective
action.
(4) The medical
physicist shall prepare a report that includes all of the following:
(a) A summary of the performance evaluation
required under subrule (1) of this rule.
(b) Recommendations for necessary
improvements.
(c) Type of dosimetry
system used, including the date of the last calibration.
(5) The report required under subrule (4) of
this rule shall be provided to the CT facility within 30 days after completion
of the evaluation.
Notes
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.