Nev. Admin. Code § 459.604 - X-ray and electron therapy installations: Selections of radiation type, energy, and stationary or moving beam
1. In equipment capable of both X-ray therapy and electron therapy:
(a) Irradiation must not be possible until a selection of radiation types, either X rays or electrons, has been made at the treatment control panel; and
(b) The radiation type selected must be displayed at the treatment control panel before and during irradiation.
2. In equipment capable of generating radiation beams of different energies:
(a) Irradiation must not be possible until a selection of energy has been made at the treatment control panel; and
(b) The energy and type of irradiation selected, either X rays or electrons, must be displayed at the treatment control panel before and during irradiation.
3. In equipment capable of both stationary-beam therapy and moving-beam therapy:
(a) Irradiation must not be possible until a selection of stationary-beam therapy or moving-beam therapy has been made at the treatment control panel;
(b) Moving-beam therapy must be controlled so that the required dose monitor units per degree of rotation is obtained; and
(c) The mode of operation must be displayed at the treatment control panel.