Nev. Admin. Code § 459.620 - Other radiographic systems: Special purpose systems
For special purpose X-ray systems:
1. A means must be provided to limit the X-ray field in the plane of the image receptor so that the X-ray field is within 2 percent of the source-image receptor distance in each dimension of the X-ray field when the axis of the X-ray beam is perpendicular to the plane of the image receptor.
2. A means must be provided to align the center of the X-ray field with the center of the image receptor to within 2 percent of the source-image receptor distance.
3. Subsections 1 and 2 may be met with a system that meets the requirements for a general purpose X-ray system as specified in NAC 459.618, or, when alignment means are also provided, may be met with either:
(a) An assortment of removable, fixed-aperture, beam-limiting devices sufficient to meet the requirement for each combination of image receptor size and source-image receptor distance for which the unit is designed with each such device having clear and permanent markings to indicate the image receptor size and source-image receptor distance for which it is designed; or
(b) A beam-limiting device having multiple fixed apertures sufficient to meet the requirement for each combination of image receptor size and source-image receptor distance for which the unit is designed. Permanent, clearly legible markings must indicate the image receptor size and source-image receptor distance for which each aperture is designed and indicate which aperture is in position for use.