Nev. Admin. Code § 459.712 - Equipment control: Radiation survey instruments
1. The
registrant shall maintain sufficient calibrated and operable radiation survey
instruments at each location where radiation producing equipment is used.
Instrumentation required by this section must have a range such that 2
millirems (0.02 millisievert) per hour through 1 rem (0.01 sievert) per hour
can be measured.
2. Each radiation
survey instrument must be calibrated:
(a)
Against appropriate energy at intervals not exceeding 6 months and, except for
battery changes, after each servicing of the instrument;
(b) So that accuracy within plus or minus 20
percent can be demonstrated at each point checked; and
(c) At two or more widely separated points,
other than zero, on each scale, as follows:
(1) For linear scale instruments, at two
points located approximately one-third and two-thirds of full scale on each
scale;
(2) For logarithmic scale
instruments, at the mid-range of each decade and at two points of at least one
decade; and
(3) For digital
instruments, at 3 points between 2 millirems (0.02 millisievert) per hour and 1
rem (0.01 sievert) per hour.
3. Records of these calibrations must be
maintained for at least 3 years after the calibration date for inspection by
the Division.
Notes
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.