N.J. Admin. Code § 7:28-16.4 - Multiple dental radiographic tube installations

Current through Register Vol. 54, No. 7, April 4, 2022

(a) No person shall use x-ray equipment in a multiple dental radiographic tube installation set-up or cause it to be used unless the following requirements are met:
1. It shall be possible to activate only one dental radiographic tube at any one time.
2. Where two or more radiographic tubes are controlled by one exposure switch, the dental radiographic tube which has been selected shall be clearly indicated prior to initiation of the exposure. For certified units only, there shall be an indicator on both the x-ray control and at or near the dental radiographic tube housing assembly which has been selected.
3. It shall be possible to energize a dental radiographic tube from an exposure switch located at a specific dental radiographic tube's remote station only when that specific dental radiographic tube is selected.
4. It shall be possible to energize a dental radiographic tube from the main control panel exposure switch only when that specific dental radiographic tube is selected.