N.Y. Comp. Codes R. & Regs. Tit. 12 § 38.22 - Surveys, checks and tests
(a)
Surveys--general. Each person who possesses any radiation source shall make, or
cause to be made, the applicable surveys required under this section and such
additional surveys as may be necessary for such person to comply with other
provisions of this Part (rule).
(b)
Instrumentation. Each person required to perform a survey by this Part (rule)
shall be provided with or have available appropriate calibrated and operable
instruments capable of detecting and measuring radiation.
(c) Radiation equipment--surveys. Every
installation and mobile source wherein radiation equipment is to be used shall
be surveyed during the initial operation of such equipment and whenever any
change is made in such installation or mobile source or in its use that might
increase the radiation level to which an individual could be exposed. When
vibrations or other physical conditions exist in such installations or mobile
sources which may cause changes in the protective features, surveys shall be
made at least every six months.
(d)
Radioactive materials--surveys. Any installation wherein radioactive materials
are handled or installed which has any readily accessible area in which there
is reasonable expectation that a radiation level will exist in excess of 2
millirems per hour, or any mobile source which could create such a radiation
level, shall be surveyed during the initial operation and whenever any change
is made in such installation or mobile source or in its use that might increase
the radiation level to which an individual could be exposed.
(e) Unsealed sources--surveys. Radiation
installations wherein radioactive material not contained in a sealed source is
handled or installed shall be surveyed at least once a month for radioactive
contamination. With regard to radioactive contamination of surfaces, compliance
with section
38.17(a) of this
Part (rule) may be assumed if the limits specified in Table 5 of section
38.41 of this Part (rule) are
met.
(f) Sealed sources--leak test.
(1) Each sealed source containing radioactive
material other than hydrogen 3, with a half life greater than 30 days and in
any form other than gas, shall be tested for leakage prior to initial use and
at successive intervals thereafter not exceeding six months, except that each
source designed for the purpose of emitting alpha particles shall be tested at
intervals not exceeding three months. Notwithstanding the periodic leak test
required by this paragraph, any licensed sealed source is exempt from such leak
tests when the source contains 3.7 MBq (100 microcuries) or less of beta and/or
gamma emitting material or 370 kBq (10 microcuries) or less of alpha emitting
material. Except for alpha sources, the periodic leak test required by this
paragraph shall not apply to sealed sources that are stored and not being used.
The sources excepted from this requirement shall be tested for leakage prior to
any use or transfer to another person unless they have been leak tested within
six months prior to the date of use or transfer. In the absence of the delivery
of a certificate by the transferor to the transferee indicating that a test
pursuant to the applicable provisions of this Part (rule) was made within six
months prior to the transfer, the source shall not be used until tested for
leakage. If there is reason to suspect that a sealed source might have been
damaged, or might be leaking, such source shall be tested for leakage before
further use.
(2) The test sample
shall be taken from selected accessible surfaces of the sealed source or from
surfaces of the device in which the source is permanently mounted or stored.
The selected surfaces should be those on which contamination would be expected
to accumulate if the source was leaking. Where applicable, the test sample
shall be taken with the source in the " off"position.
(3) The leak test technique shall be capable
of detecting 185 Bq (0.005 microcurie) of removable radioactivity, or from
sealed radium sources, the escape of radon at the rate of 37 Bq (0.001
microcuries) per 24 hours.
(4)
Detection of a leak in any sealed source in excess of the sensitivity levels
set forth in this subdivision shall result in immediate suspension of the use
of such source until it is decontaminated and repaired or disposed of in
accordance with section
38.20 of this Part
(rule).
(5) Records of leak test
results shall be kept in units of microcuries per test sample and maintained
for inspection by the commissioner. A report shall be submitted to the
commissioner for each source found to be leaking in excess of the above
sensitivity levels within five days of detection of the leak and shall describe
the equipment involved, the test results, and the corrective actions
taken.
(g) Protective
devices. All protective devices such as interlocks, safety switches, fume
hoods, filters and trapping devices for radioactive gases, shall be maintained
in good repair and proper operating condition.