10A N.C. Admin. Code 15 .0213 - ADDITIONAL REQUIREMENTS: REGISTERED SERVICES
(a) An
applicant for registration of diagnostic area radiation survey, diagnostic
radiation output measurements or therapeutic calibration services pursuant to
Rule.0205 of this Section shall meet the following additional requirements:
(1) The applicant shall have adequate
radiation survey and radiation measurement equipment appropriate to the
services requested for authorization.
(2) The applicant shall ensure that the
equipment in Subparagraph (a)(1) of this Rule is calibrated at least every 12
months by persons registered to provide such services pursuant to Rule.0205 of
this Section, except as provided in Subparagraph (a)(3) of this Rule. The
agency may approve less frequent calibration of equipment used for therapy
calibration, provided the applicant satisfies the agency that the proposed
frequency and procedures will provide equivalent or better assurance of proper
calibration.
(3) The applicant may
perform the equipment calibrations required in Subparagraph (a)(2) of this Rule
provided that:
(A) such calibrations are
currently traceable to the National Institute of Standards and
Technology;
(B) the calibration
procedures are approved by the agency;
(C) the radiation sources used for such
calibration are licensed or registered as required by the rules in this
Chapter; and
(D) the equipment is
labeled to indicate the date of calibration and records of the calibration are
maintained.
(4) The
applicant shall submit:
(A) a description of
the procedures that will be used in performing area radiation surveys including
a list of all guides and references to the employed;
(B) a copy of all forms, reports and
documents that will be supplied to customers;
(C) samples of three different types of
surveys;
(D) samples of three
reports of diagnostic radiation output measurements; and
(E) samples of three therapeutic calibration
reports.
(b)
An applicant for registration of services pursuant to Rule.0205 of this Section
who proposes to provide diagnostic radiographic, fluoroscopic and therapeutic
facility and shielding design services shall meet the following additional
requirements:
(1) The applicant shall submit
examples of the facility and shielding design which will be provided to
clients.
(2) The applicant shall
submit examples of the calculations which will be performed as part of the
facility and shielding design along with any guides, occupancy factor
rationales, and workload estimation rationales which will be used.
(3) The applicant shall ensure that the
facility and shielding design services provided to licensees and registrants of
the agency satisfy the applicable requirements in this Chapter.
Notes
Eff. June 1, 1989;
Amended Eff. June 1, 1993;
Transferred and Recodified from 15A NCAC 11.0213 Eff. February 1, 2015.
Eff. June 1, 1989;
Amended Eff. June 1, 1993;
Transferred and Recodified from 15A NCAC 11 .0213 Eff. February 1, 2015.