N.D. Admin Code 33.1-10-15-07 - Therapeutic radiation machines - Photon therapy systems (five hundred kilovolts and above) and electron therapy systems (five hundred kilo electron volts and above)
1. Possession of
survey instruments. Each facility location authorized to use a therapeutic
radiation machine in accordance with this section shall possess appropriately
calibrated portable monitoring equipment. As a minimum, such equipment shall
include a portable radiation measurement survey instrument capable of measuring
dose rates over the range of ten microsieverts (1 mrem) per hour to ten
millisieverts (1,000 mrem) per hour. The survey instruments shall be operable
and calibrated in accordance with section
33.1-10-15-08.
2. Leakage radiation outside the maximum
useful beam in photon and electron modes.
a.
The absorbed dose due to leakage radiation (excluding neutrons) at any point
outside the maximum-sized useful beam, but within a circular plane of radius
two meters which is perpendicular to and centered on the central axis of the
useful beam at the nominal treatment distance (i.e., patient plane), shall not
exceed a maximum of two-tenths percent and an average of one-tenth percent of
the absorbed dose on the beam axis at the nominal treatment distance.
Measurements shall be averaged over an area not exceeding one hundred square
centimeters at a minimum of sixteen points uniformly distributed in the
plane;
b. Except for the area
defined in subdivision a, the absorbed dose due to leakage radiation (excluding
neutrons) at one meter from the electron path between the electron source and
the target or electron window shall not exceed five-tenths percent of the
absorbed dose on the central axis of the beam at the nominal treatment
distance. Measurements shall be averaged over an area not exceeding one hundred
square centimeters;
c. For
equipment manufactured after January 1, 2011, the neutron absorbed dose outside
the useful beam shall be in compliance with international electrotechnical
commission (IEC) document 601-2-1 (most current revision); and
d. For each therapeutic radiation machine,
the registrant shall determine, or obtain from the manufacturer, the leakage
radiation existing at the positions specified in subdivisions a and c for the
specified operating conditions. Records on leakage radiation measurements shall
be maintained at the installation for inspection by the department.
3. Leakage radiation through
beam-limiting devices.
a. Photon radiation.
All adjustable or interchangeable beam-limiting devices shall attenuate the
useful beam such that at the nominal treatment distance, the maximum absorbed
dose anywhere in the area shielded by the beam-limiting devices shall not
exceed two percent of the maximum absorbed dose on the central axis of the
useful beam measured in a ten centimeter by ten centimeter radiation
field.
b. Electron radiation. All
adjustable or interchangeable electron applicators shall attenuate the
radiation, including photon radiation generated by electrons incident on the
beam-limiting device and electron applicator and other parts of the radiation
head, such that the absorbed dose in a plane perpendicular to the central axis
of the useful beam at the nominal treatment distance shall not exceed:
(1) A maximum of two percent and average of
five-tenths percent of the absorbed dose on the central axis of the useful beam
at the nominal treatment distance. This limit shall apply beyond a line seven
centimeters outside the periphery of the useful beam; and
(2) A maximum of ten percent of the absorbed
dose on the central axis of the useful beam at the nominal treatment distance.
This limit shall apply beyond a line two centimeters outside the periphery of
the useful beam.
c.
Measurement of leakage radiation.
(1) Photon
radiation. Measurements of leakage radiation through the beam-limiting devices
shall be made with the beam-limiting devices closed and any residual aperture
blocked by at least two tenth-value layers of suitable absorbing material. In
the case of overlapping beam-limiting devices, the leakage radiation through
each set shall be measured independently at the depth of maximum dose.
Measurements shall be made using a radiation detector of area not exceeding ten
square centimeters; and
(2)
Electron radiation. Measurements of leakage radiation through the electron
applicators shall be made with the electron beam directed into the air and
using a radiation detector of area up to but not exceeding one square
centimeter suitably protected against radiation which has been scattered from
material beyond the radiation detector. Measurements shall be made using one
centimeter of water equivalent buildup material.
4. Filters and wedges.
a. Each wedge filter that is removable from
the system shall be clearly marked with an identification number. For removable
wedge filters, the nominal wedge angle shall appear on the wedge or wedge tray
(if permanently mounted to the tray). If the wedge or wedge tray is
significantly damaged, the wedge transmission factor shall be
redetermined;
b. If the absorbed
dose rate information required by subsection 2 relates exclusively to operation
with a field-flattening filter or beam-scattering foil in place, such foil or
filter shall be removable only by the use of tools; and
c. For equipment which utilizes wedge
filters, interchangeable field-flattening filters, or interchangeable
beam-scattering foils:
(1) Irradiation shall
not be possible until a selection of a filter or a positive selection to use
"no filter" has been made at the treatment control panel, either manually or
automatically;
(2) An interlock
system shall be provided to prevent irradiation if the filter selected is not
in the correct position;
(3) A
display shall be provided at the treatment control panel showing the wedge
filters, interchangeable field-flattening filters, or interchangeable
beam-scattering foils, or both, in use; and
(4) An interlock shall be provided to prevent
irradiation if any filter or beam-scattering foil selection operation, or both,
carried out in the treatment room does not agree with the filter or
beam-scattering foil selection operation, or both, carried out at the treatment
control panel.
5. Stray radiation in the useful beam. For
equipment manufactured after January 1, 2011, the registrant shall determine
during acceptance testing, or obtain from the manufacturer, data sufficient to
ensure that x-ray stray radiation, in the useful electron beam, absorbed dose
at the surface during x-ray irradiation, and stray neutron radiation in the
useful x-ray beam are in compliance with international electrotechnical
commission (IEC) document 601-2-1 (most current revision).
6. Beam monitors. All therapeutic radiation
machines subject to this section shall be provided with redundant
beam-monitoring systems. The sensors for these systems shall be fixed in the
useful beam during treatment to indicate the dose monitor unit rate.
a. Equipment shall be provided with at least
two independently powered integrating dose meters. Alternatively, common
elements may be used if the production of radiation is terminated upon failure
of any common element.
b. The
detector and the system into which that detector is incorporated shall meet the
following requirements:
(1) Each detector
shall be removable only with tools and, if movable, shall be interlocked to
prevent incorrect positioning;
(2)
Each detector shall form part of a beam-monitoring system from whose readings
in dose monitor units the absorbed dose at a reference point can be
calculated;
(3) Each
beam-monitoring system shall be capable of independently monitoring,
interrupting, and terminating irradiation; and
(4) The design of the beam-monitoring systems
shall ensure that the:
(a) Malfunctioning of
one system shall not affect the correct functioning of the other systems;
and
(b) Failure of either system
shall terminate irradiation or prevent the initiation of radiation;
and
(5) Each
beam-monitoring system shall have a legible display at the treatment control
panel. Each display shall:
(a) Maintain a
reading until intentionally reset;
(b) Have only one scale and no electrical or
mechanical scale multiplying factors;
(c) Utilize a design such that increasing
dose is displayed by increasing numbers; and
(d) In the event of power failure, the
beam-monitoring information required in subparagraph c displayed at the control
panel at the time of failure shall be retrievable in at least one system for a
twenty-minute period of time.
7. Beam symmetry.
a. Bent-beam linear accelerators subject to
this section shall be provided with auxiliary devices to monitor beam
symmetry;
b. The devices referenced
in subdivision a shall be able to detect field asymmetry greater than five
percent; and
c. The devices
referenced in subdivision a shall be configured to terminate irradiation if the
specifications in subdivision b cannot be maintained.
8. Selection and display of dose monitor
units.
a. Irradiation shall not be possible
until a new selection of a number of dose monitor units has been made at the
treatment control panel;
b. The
preselected number of dose monitor units shall be displayed at the treatment
control panel until reset manually for the next irradiation;
c. After termination of irradiation, it shall
be necessary to reset the dosimeter display before subsequent treatment can be
initiated; and
d. After termination
of irradiation, it shall be necessary for the operator to reset the preselected
dose monitor units before irradiation can be initiated.
9. Air kerma rate or absorbed dose rate. A
system shall be provided from whose readings the air kerma rate or absorbed
dose rate at a reference point can be calculated. The radiation detectors
specified in subsection 6 may form part of this system. In addition:
a. The dose monitor unit rate shall be
displayed at the treatment control panel;
b. If the equipment can deliver under any
conditions an air kerma rate or absorbed dose rate at the nominal treatment
distance more than twice the maximum value specified by the manufacturer, a
device shall be provided which terminates irradiation when the air kerma rate
or absorbed dose rate exceeds a value twice the specified maximum. The dose
rate at which the irradiation will be terminated shall be a record maintained
by the registrant;
c. If the
equipment can deliver under any fault conditions an air kerma rate or absorbed
dose rate at the nominal treatment distance more than ten times the maximum
value specified by the manufacturer, a device shall be provided to prevent the
air kerma rate or absorbed dose rate anywhere in the radiation field from
exceeding twice the specified maximum value and to terminate irradiation if the
excess absorbed dose at the nominal treatment distance exceeds four gray (400
rad); and
d. For each therapeutic
radiation machine, the registrant shall determine, or obtain from the
manufacturer, the maximum values specified in subdivisions b and c for the
specified operating conditions. Records of these maximum values shall be
maintained at the installation for inspection by the department.
10. Termination of irradiation by
the beam-monitoring system or systems during stationary beam radiation therapy.
a. Each primary system shall terminate
irradiation when the preselected number of dose monitor units has been detected
by the system;
b. If the original
design of the equipment included a secondary dose monitoring system, that
system shall be capable of terminating irradiation when not more than fifteen
percent or forty dose monitor units above the preselected number of dose
monitor units set at the control panel has been detected by the secondary dose
monitoring system; and
c. An
indicator on the control panel shall show which monitoring system has
terminated irradiation.
d. For new
equipment, a secondary dose monitoring system must be present. That system must
be capable of terminating irradiation when not more than ten percent or
twenty-five dose monitoring units above the preselected number of dose monitor
units set at the control panel has been detected by the secondary dose
monitoring system.
11.
Termination of irradiation. It shall be possible to terminate irradiation and
equipment movement or go from an interruption condition to termination
condition at any time from the operator's position at the treatment control
panel.
12. Interruption of
irradiation. It shall be possible to interrupt irradiation and equipment
movements at any time from the treatment control panel. Following an
interruption it shall be possible to restart irradiation by operator action
without any reselection of operating conditions. If any change is made of a
preselected value during an interruption, irradiation and equipment movements
shall be automatically terminated.
13. Timer. A suitable irradiation control
device shall be provided to terminate the irradiation after a preset time
interval.
a. A timer shall be provided which
has a display at the treatment control panel. The timer shall have a preset
time selector and an elapsed time indicator;
b. The timer shall be a cumulative timer that
activates with an indication of "BEAM-ON" and retains its reading after
irradiation is interrupted or terminated. After irradiation is terminated and
before irradiation can be reinitiated, it shall be necessary to reset the
elapsed time indicator;
c. After
termination of irradiation and before irradiation can be reinitiated, it shall
be necessary to reset the preset time selector; and
d. The timer shall terminate irradiation when
a preselected time has elapsed, if the dose monitoring systems have not
previously terminated irradiation.
14. Selection of radiation type. Equipment
capable of both x-ray therapy and electron therapy shall meet the following
additional requirements:
a. Irradiation shall
not be possible until a selection of radiation type (x-rays or electrons) has
been made at the treatment control panel;
b. The radiation type selected shall be
displayed at the treatment control panel before and during
irradiation;
c. An interlock system
shall be provided to ensure that the equipment can principally emit only the
radiation type that has been selected;
d. An interlock system shall be provided to
prevent irradiation with x-rays, except to obtain an image, when electron
applicators are fitted;
e. An
interlock system shall be provided to prevent irradiation with electrons when
accessories specific for x-ray therapy are fitted; and
f. An interlock system shall be provided to
prevent irradiation if any selected operations carried out in the treatment
room do not agree with the selected operations carried out at the treatment
control panel.
15.
Selection of energy. Equipment capable of generating radiation beams of
different energies shall meet the following requirements:
a. Irradiation shall not be possible until a
selection of energy has been made at the treatment control panel;
b. An interlock system shall be provided to
prevent irradiation if any selected operations carried out in the treatment
room do not agree with the selected operations carried out at the treatment
control panel;
c. The nominal
energy value selected shall be displayed at the treatment control panel until
reset manually for the next irradiation. After termination of irradiation, it
shall be necessary to reset the nominal energy value selected before subsequent
treatment can be initiated;
d.
Irradiation shall not be possible until the appropriate flattening filter or
scattering foil for the selected energy is in its proper location;
e. An interlock system shall be provided to
terminate irradiation if the energy of the electrons striking the x-ray target
or electron window deviates by more than twenty percent or three megaelectron
volts, whichever is smaller, from the selected nominal energy; and
f. For equipment manufactured after January
1, 2011, the selection of energy shall be in compliance with international
electrotechnical commission (IEC) document 601-2-1 (most current
revision).
16. Selection
of stationary beam radiation therapy or moving beam radiation therapy.
Therapeutic radiation machines capable of both stationary beam radiation
therapy and moving beam radiation therapy shall meet the following
requirements:
a. Irradiation shall not be
possible until a selection of stationary beam radiation therapy or moving beam
radiation therapy has been made at the treatment control panel;
b. The mode of operation shall be displayed
at the treatment control panel;
c.
An interlock system shall be provided to ensure that the equipment can operate
only in the mode that has been selected;
d. An interlock system shall be provided to
prevent irradiation if any selected parameter in the treatment room does not
agree with the selected parameter at the treatment control panel;
e. Moving beam radiation therapy shall be
controlled to obtain the selected relationships between incremental dose
monitor units and incremental movement.
(1)
An interlock system shall be provided to terminate irradiation if the number of
dose monitor units delivered in any ten degrees of rotation or one centimeter
of linear motion differs by more than twenty percent from the selected
value;
(2) Where the angle
terminates the irradiation in moving beam radiation therapy, the dose monitor
units delivered shall differ by less than five percent from the dose monitor
unit value selected;
(3) An
interlock shall be provided to prevent motion of more than five degrees or one
centimeter beyond the selected limits during moving beam radiation
therapy;
(4) An interlock shall be
provided to require that a selection of direction be made at the treatment
control panel in all units which are capable of both clockwise and
counterclockwise moving beam radiation therapy; and
(5) Moving beam radiation therapy shall be
controlled with both primary position sensors and secondary position sensors to
obtain the selected relationships between incremental dose monitor units and
incremental movement;
f.
Where the beam monitor system terminates the irradiation in moving beam
radiation therapy, the termination of irradiation shall be as required by
subsection 10; and
g. An interlock
system shall be provided to terminate irradiation if movement:
(1) Occurs during stationary beam radiation
therapy; or
(2) Does not start or
stops during moving beam radiation therapy unless such stoppage is a preplanned
function.
17.
Facility design requirements for therapeutic radiation machines operating above
five hundred kilovolts. In addition to shielding adequate to meet requirements
of section
33.1-10-15-09, the following
design requirements are made:
a. Protective
barriers. All protective barriers shall be fixed, except for access doors to
the treatment room or movable beam interceptors;
b. Control panel. In addition to other
requirements specified in this chapter, the control panel shall also:
(1) Be located outside the treatment
room;
(2) Provide an indication of
whether electrical power is available at the control panel and if activation of
the radiation is possible;
(3)
Provide an indication of whether radiation is being produced; and
(4) Include an access control (locking)
device that will prevent unauthorized use of the therapeutic radiation
machine;
c. Viewing
systems. Windows, mirrors, closed-circuit television, or an equivalent viewing
system shall be provided to permit continuous observation of the patient
following positioning and during irradiation and shall be so located that the
operator may observe the patient from the treatment control panel. The
therapeutic radiation machine shall not be used for patient irradiation unless
at least one viewing system is operational;
d. Aural communications. Provision shall be
made for continuous two-way aural communication between the patient and the
operator at the control panel. The therapeutic radiation machine shall not be
used for irradiation of patients unless continuous two-way aural communication
is possible;
e. Room entrances.
Treatment room entrances shall be provided with warning lights in a readily
observable position near the outside of all access doors, which will indicate
when the useful beam is "ON" and when it is "OFF";
f. Entrance interlocks. Interlocks shall be
provided such that all access controls are activated before treatment can be
initiated or continued. If the radiation beam is interrupted by any access
control, it shall not be possible to restore the machine to operation without
resetting the access control and reinitiating irradiation by manual action at
the control panel;
g. Beam
interceptor interlocks. If the shielding material in any protective barrier
requires the presence of a beam interceptor to ensure compliance with
33.1-10-04.2 [ 10 CFR 20.1301a and 10 CFR 20.1301b] of these regulations,
interlocks shall be provided to prevent the production of radiation, unless the
beam interceptor is in place, whenever the useful beam is directed at the
designated barriers;
h. Emergency
cutoff switches. At least one emergency power cutoff switch shall be located in
the radiation therapy room and shall terminate all equipment electrical power,
including radiation and mechanical motion. This switch is in addition to the
termination switch required by subsection 11. All emergency power cutoff
switches shall include a manual reset so that the therapeutic radiation machine
cannot be restarted from the unit's control console without resetting the
emergency cutoff switch;
i. Safety
interlocks. All safety interlocks shall be designed so that any defect or
component failure in the safety interlock system prevents or terminates
operation of the therapeutic radiation machine; and
j. Surveys for residual radiation. Surveys
for residual activity shall be conducted on all therapeutic radiation machines
capable of generating photon and electron energies above ten million volts
prior to machining, removing, or working on therapeutic radiation machine
components which may have become activated due to photo-neutron
production.
18. Radiation
therapy physicist support.
a. The services of
a radiation therapy physicist shall be required in facilities having
therapeutic radiation machines with energies of five hundred kilovolts and
above. The radiation therapy physicist shall be responsible for:
(1) Full calibrations required by subsection
20 and protection surveys required by subsection 1 of section
33.1-10-15-04;
(2) Supervision and review of
dosimetry;
(3) Beam data
acquisition and transfer for computerized dosimetry, and supervision of its
use;
(4) Quality assurance,
including quality assurance check review required by subdivision e of
subsection 21;
(5) Consultation
with the authorized user in treatment planning, as needed; and
(6) Perform calculations and assessments
regarding misadministrations.
b. If the radiation therapy physicist is not
a full-time employee of the registrant, the operating procedures required by
subsection 17 shall also specifically address how the radiation therapy
physicist is to be contacted for problems or emergencies, as well as the
specific actions, if any, to be taken until the radiation therapy physicist can
be contacted.
19.
Operating procedures.
a. No individual, other
than the patient, shall be in the treatment room during treatment or during any
irradiation for testing or calibration purposes;
b. Therapeutic radiation machines shall not
be made available for medical use unless the requirements of subsection 1 of
section 33.1-10-15-04 and subsections 20
and 21 of this section have been met;
c. Therapeutic radiation machines, when not
in operation, shall be secured to prevent unauthorized use;
d. When adjustable beam-limiting devices are
used, the position and shape of the radiation field shall be indicated by a
light field;
e. If a patient must
be held in position during treatment, mechanical supporting or restraining
devices shall be used; and
f. A
copy of the current operating and emergency procedures shall be maintained at
the therapeutic radiation machine control console.
20. Acceptance testing, commissioning, and
full calibration measurements.
a. Acceptance
testing, commissioning, and full calibration of a therapeutic radiation machine
subject to this section shall be performed by, or under the direct supervision
of, a radiation therapy physicist.
b. Acceptance testing and commissioning shall
be performed in accordance with "American association of physicists in medicine
code of practice for radiotherapy accelerators: report of American association
of physicists in medicine radiation therapy task group 45" and shall be
conducted before the first medical use following installation or reinstallation
of the therapeutic radiation machine.
c. Full calibration shall include measurement
of all parameters required by table II of "comprehensive QA for radiation
oncology: report of American association of physicists in medicine radiation
therapy committee task group 40" and shall be performed in accordance with
"American association of physicists in medicine code of practice for
radiotherapy accelerators: report of American association of physicists in
medicine radiation therapy task group 45". Although it shall not be necessary
to complete all elements of a full calibration at the same time, all parameters
(for all energies) shall be completed at intervals not exceeding twelve
calendar months, unless a more frequent interval is required in table
II.
d. The radiation therapy
physicist shall perform all elements of a full calibration necessary to
determine that all parameters are within acceptable limits:
(1) Whenever quality assurance check
measurements indicate that the radiation output differs by more than five
percent from the value obtained at the last full calibration and the difference
cannot be reconciled. Therapeutic radiation machines with multi-energy or
multimode capabilities, or both, shall only require measurements for those
modes or energies, or both, which are not within their acceptable range;
and
(2) Following any component
replacement, major repair, or modification of components that could
significantly affect the characteristics of the radiation beam. If the repair,
replacement, or modification does not affect all modes or energies,
measurements shall be performed on the affected mode or energy that is in most
frequent clinical use at the facility. The remaining energies or modes may be
validated with quality assurance check procedures against the criteria in
paragraph 1.
e. The
registrant shall maintain a record of each calibration in an auditable form for
the duration of the registration. The record shall include the date of the
calibration; the manufacturer's name, model number, and serial number for the
therapeutic radiation machine; the model numbers and serial numbers of the
instruments used to calibrate the therapeutic radiation machine; and the
signature of the radiation therapy physicist responsible for performing the
calibration.
21. Periodic
quality assurance checks.
a. Periodic quality
assurance checks shall be performed on all therapeutic radiation machines
subject to this section at intervals not to exceed those specified in
"comprehensive QA for radiation oncology: report of American association of
physicists in medicine radiation therapy committee task group 40";
b. To satisfy the requirement of subdivision
a, quality assurance checks shall include determination of central axis
radiation output and a representative sampling of periodic quality assurance
checks contained in "comprehensive QA for radiation oncology: report of
American association of physicists in medicine radiation therapy committee task
group 40". Representative sampling shall include all referenced periodic
quality assurance checks in an interval not to exceed twelve consecutive
calendar months;
c. The registrant
shall use a dosimetry system that has been intercompared within the previous
twelve months with the dosimetry system described in subdivision a of
subsection 3 of section
33.1-10-15-04 to make the
periodic quality assurance checks required in subdivision b;
d. The registrant shall perform periodic
quality assurance checks required by subdivision a in accordance with
procedures established by the radiation therapy physicist;
e. The registrant shall review the results of
each periodic radiation output check according to the following procedures:
(1) The authorized user and radiation therapy
physicist shall be immediately notified if any parameter is not within its
acceptable tolerance. The therapeutic radiation machine shall not be made
available for subsequent medical use until the radiation therapy physicist has
determined that all parameters are within their acceptable
tolerances;
(2) If all quality
assurance check parameters appear to be within their acceptable range, the
quality assurance check shall be reviewed and signed by either the authorized
user or radiation therapy physicist within three treatment days; and
(3) The radiation therapy physicist shall
review and sign the results of each radiation output quality assurance check at
intervals not to exceed one month;
f. Therapeutic radiation machines subject to
this section shall have safety quality assurance checks listed in
"comprehensive QA for radiation oncology: report of American association of
physicists in medicine radiation therapy committee task group 40" performed at
intervals not to exceed one week;
g. To satisfy the requirement of subdivision
e, safety quality assurance checks shall ensure proper operation of:
(1) Electrical interlocks at each external
beam radiation therapy room entrance;
(2) Proper operation of the "BEAM-ON",
interrupt, and termination switches;
(3) Beam condition indicator lights on the
access doors, on the control console, and in the radiation therapy
room;
(4) Viewing
systems;
(5) Electrically operated
treatment room doors from inside and outside the treatment room; and
(6) At least one emergency power cutoff
switch. If more than one emergency power cutoff switch is installed and not all
switches are tested at once, each switch shall be tested on a rotating basis.
Safety quality assurance checks of the emergency power cutoff switches may be
conducted at the end of the treatment day in order to minimize possible
stability problems with the therapeutic radiation machine;
h. The registrant shall promptly repair any
system identified in subdivision g that is not operating properly;
and
i. The registrant shall
maintain a record of each quality assurance check required by subdivisions a
and b for three years. The record shall include the date of the quality
assurance check; the manufacturer's name, model number, and serial number of
the therapeutic radiation machine; the manufacturer's name, model number, and
serial number for the instruments used to measure the radiation output of the
therapeutic radiation machine; and the signature of the individual who
performed the periodic quality assurance check.
Notes
General Authority: NDCC 23.1-03-02; S.L. 2017, ch. 199, § 1
Law Implemented: NDCC 23.1-03-03; S.L. 2017, ch. 199, § 18
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.