Or. Admin. R. 333-106-0220 - Fluoroscopic Indication of Potential and Current

Current through Register Vol. 61, No. 4, April 1, 2022

During fluoroscopy and cinefluorography X-ray tube potential and current shall be continuously indicated. Deviation of X-ray tube potential and current from the indicated values shall not exceed the maximum deviation stated by the manufacturer.