Or. Admin. R. 333-123-0070 - Control Panel Functions

A control panel must be designed to provide:

(1) An indication of whether electrical power is available at the control panel and if activation of the electronic brachytherapy source is possible;
(2) An indication of whether X-rays are being produced;
(3) A means for displaying electronic brachytherapy source accelerating voltage(kV), beam current(ยตA), pre-set value of radiation dose(cGy) and real time display of dose delivered;
(4) The means for terminating an exposure at any time; and
(5) An access locking control device that can prevent unauthorized use of the electronic brachytherapy device.

Notes

Or. Admin. R. 333-123-0070
PH 4-2013, f. & cert. ef. 1-29-13

Stat. Auth.: ORS 453.605 - 453.807

Stats. Implemented: ORS 453.605 - 453.807