25 Pa. Code § 227a.41 - X-ray on status
Open-beam radiation-producing devices must have a conspicuous and active indication of the following, as applicable:
(1) X-ray tube "on-off" status indicator
located near the radiation source housing. The warning lights required under
§
227a.11(b)
(relating to warning devices), meet this requirement if the warning lights are
conspicuous and viewable by anyone near the primary beam .
(2) Shutter "open-closed" status indicator
located at the control panel and near each beam port on the radiation source
housing, if the primary beam is controlled with a shutter . The shutter status
device must be clearly labeled to indicate whether the shutter is open or
closed. The status light at the control panel meets the requirement for the
status light at the beam port if the status light at the control panel is
conspicuous and viewable by anyone near the primary beam .
(3) The X-ray tube "on-off" status indicator
and the shutter "open-closed" status indicators must be of a fail-safe
design .
Notes
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.