Tenn. Comp. R. & Regs. 0400-20-07-.65 - INSTALLATION, MAINTENANCE, ADJUSTMENT, AND REPAIR
(1) Only a
person specifically licensed by the Division, the U.S. Nuclear Regulatory
Commission, or an Agreement State shall install, maintain, adjust, or repair a
remote afterloader unit, teletherapy unit, or gamma stereotactic radiosurgery
unit that involves work on the source(s) shielding, the source(s) driving unit,
or other electronic or mechanical component that could expose the source(s),
reduce the shielding around the source(s), or compromise the radiation safety
of the unit or the source(s).
(2)
Except for low dose-rate remote afterloader units, only a person specifically
licensed by the Division, the U.S. Nuclear Regulatory Commission, or an
Agreement State shall install, replace, relocate, or remove a sealed source or
source contained in other remote afterloader units, teletherapy units, or gamma
stereotactic radiosurgery units.
(3) For a low dose-rate remote afterloader
unit, only a person specifically licensed by the Division, the U.S. Nuclear
Regulatory Commission, or an Agreement State or an authorized medical physicist
shall install, replace, relocate, or remove a sealed source(s) contained in the
unit.
(4) A licensee shall retain a
record of the installation, maintenance, adjustment, and repair of remote
afterloader units, teletherapy units, and gamma stereotactic radiosurgery units
in accordance with Rule 0400-20-07-.101.
Notes
Authority: T.C.A. ยงยง 68-202-101 et seq., 68-202-201 et seq. and 4-5-201 et seq.