Tenn. Comp. R. & Regs. 0400-20-07-.68 - DOSIMETRY EQUIPMENT
(1) Except for low
dose-rate remote afterloader sources where the source output or activity is
determined by the manufacturer, a licensee shall have a calibrated dosimetry
system available for use. To satisfy this requirement, one of the following two
conditions must be met:
(a) The system must
have been calibrated using a system or source traceable to the National
Institute of Science and Technology (NIST) and published protocols accepted by
nationally recognized bodies; or by a calibration laboratory accredited by the
American Association of Physicists in Medicine (AAPM). The calibration must
have been performed within the previous 2 years and after any servicing that
may have affected system calibration; or
(b) The system must have been calibrated
within the previous 4 years. 18 to 30 months after that calibration, the system
must have been intercompared with another dosimetry system that was calibrated
within the past 24 months by NIST or by a calibration laboratory accredited by
the AAPM. The results of the intercomparison must have indicated that the
calibration factor of the licensee's system had not changed by more than 2
percent. A licensee may not use the intercomparison result to change the
calibration factor. When intercomparing dosimetry systems to be used for
calibrating sealed sources for therapeutic units, the licensee shall use a
comparable unit with beam attenuators or collimators, as applicable, and
sources of the same radionuclide as the source used at the licensee's facility.
.
(2) A licensee shall
have available for use a dosimetry system available for use for spot-check
output measurements, if applicable. To satisfy this requirement, the system may
be compared with a system that has been calibrated in accordance with paragraph
(1) of this rule. This comparison must have been performed within the previous
year and after each servicing that may have affected system calibration. The
spot-check system may be the same system used to meet the requirement in
paragraph (1) of this rule.
(3) A
licensee shall retain a record of each calibration, intercomparison, and
comparison in accordance with Rule 0400-20-07-.103.