Tenn. Comp. R. & Regs. 0400-20-07-.69 - FULL CALIBRATION MEASUREMENTS ON TELETHERAPY UNITS
(1)
A licensee authorized to use a teletherapy unit for medical use shall perform
full calibration measurements on each teletherapy unit:
(a) Before the first medical use of the unit;
and
(b) Before medical use under
the following conditions:
1. Whenever
spot-check measurements indicate that the output differs by more than 5 percent
from the output obtained at the last full calibration corrected mathematically
for radioactive decay;
2. Following
replacement of the source or following reinstallation of the teletherapy unit
in a new location;
3. Following any
repair of the teletherapy unit that includes removal of the source or major
repair of the components associated with the source exposure assembly;
and
(c) At intervals not
exceeding 1 year.
(2) To
satisfy the requirement of paragraph (1) of this rule, full calibration
measurements must include determination of:
(a) The output within ±3 percent for
the range of field sizes and for the distance or range of distances used for
medical use;
(b) The coincidence of
the radiation field and the field indicated by the light beam localizing
device;
(c) The uniformity of the
radiation field and its dependence on the orientation of the useful
beam;
(d) Timer accuracy and
linearity over the range of use;
(e) On-off error; and
(f) The accuracy of all distance measuring
and localization devices in medical use.
(3) A licensee shall use the dosimetry system
described in paragraph (1) of Rule 0400-20-07-.68 to measure the output for one
set of exposure conditions. The remaining radiation measurements required in
subparagraph (2)(a) of this rule may be made using a dosimetry system that
indicates relative dose rates.
(4)
A licensee shall make full calibration measurements required by paragraph (1)
of this rule in accordance with published protocols accepted by nationally
recognized bodies.
(5) A licensee
shall mathematically correct the outputs determined in subparagraph (2)(a) of
this rule for physical decay for intervals not exceeding 1 month for cobalt-60,
6 months for cesium-137, or at intervals consistent with 1 percent decay for
all other nuclides.
(6) Full
calibration measurements required by paragraph (1) of this rule and physical
decay corrections required by paragraph (5) of this rule must be performed by
the authorized medical physicist.
(7) A licensee shall retain a record of each
calibration in accordance with Rule 0400-20-07-.104.
Notes
Authority: T.C.A. §§ 68-202-101 et seq., 68-202-201 et seq. and 4-5-201 et seq.
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.