Tenn. Comp. R. & Regs. 0400-20-07-.78 - THERAPY-RELATED COMPUTER SYSTEMS
(1) A licensee
shall perform or shall verify and maintain documentation of acceptance testing
on the treatment planning system in accordance with published protocols
accepted by nationally recognized bodies. At a minimum, the acceptance testing
must include, as applicable, verification of:
(a) The source-specific input parameters
required by the dose calculation algorithm;
(b) The accuracy of dose, dwell time, and
treatment time calculations at representative points;
(c) The accuracy of isodose plots and graphic
displays;
(d) The accuracy of the
software used to determine sealed source positions from radiographic images;
and
(e) The accuracy of electronic
transfer of the treatment delivery parameters to the treatment delivery unit
from the treatment planning system.