(1) Analysis of the contaminants listed in
paragraph (2) of Rule 0400-45-01-.25 for the purpose of determining compliance
with the maximum contaminant level shall be conducted as follows:
(a) Groundwater systems shall take a minimum
of one sample at every entry point to the distribution system which is
representative of each well after treatment (hereafter called a sampling
point). The Department may designate additional sampling points within the
distribution system or at the consumer's tap which more accurately determines
consumer exposure. Each sample must be taken at the same sampling point unless
conditions make another sampling point more representative of each source or
treatment plant or within the distribution system.
(b) Surface water systems shall take a
minimum of one sample at points in the distribution system that are
representative of each source or at each entry point to the distribution system
after treatment (hereafter called a sampling point). The Department may
designate additional sampling points within the distribution system or at the
consumer's tap which more accurately determines consumer exposure. Each sample
must be taken at the same sampling point unless conditions make another
sampling point more representative of each source, treatment plant, or within
the distribution system. Surface water systems include systems with a
combination of surface and ground sources.
(c) If the system draws water from more than
one source and the sources are combined before distribution, the system must
sample at an entry point to the distribution system during periods when water
representative of all sources is being used.
(d) Each community and non-transient
non-community water system shall take four consecutive quarterly samples for
each contaminant listed in paragraph (2) of Rule 0400-45-01-.25 during each
compliance period beginning in the initial compliance period.
(e) If the initial monitoring for
contaminants listed in paragraph (2) of Rule 0400-45-01-.25 has been completed
by December 31, 1992, and none of the contaminants were detected, then each
ground and surface water system shall take one sample annually beginning with
the initial compliance period.
(f)
After a minimum of three years of annual sampling, ground water systems which
have not detected any contaminant listed in paragraph (2) of Rule
0400-45-01-.25 shall take one sample during each compliance period.
(g) If a contaminant listed in paragraph (2)
of Rule 0400-45-01-.25 is detected at a level exceeding 0.0005 mg/l in any
sample, then:
1. The system must monitor
quarterly at each sampling point which resulted in a detection.
2. The Department may decrease the quarterly
monitoring requirement provided it has determined that the system is below the
maximum contaminant level. In no case shall this determination be made unless a
groundwater system takes a minimum of two quarterly samples and a surface water
system takes a minimum of four quarterly samples.
3. If the Department determines that the
system is reliably and consistently below the MCL, the Department may allow the
system to monitor annually. Systems which monitor annually must monitor during
the quarter(s) which previously yielded the highest analytical
result.
(h) Systems
which violate the maximum contaminant level for the contaminants listed in
paragraph (2) of Rule 0400-45-01-.25 must monitor quarterly. After a minimum of
four quarterly samples shows the system is in compliance with the MCL and the
Department determines that the system is below the maximum contaminant level,
the system may monitor annually during the quarter which previously yielded the
highest result.
(i) The Department
may require a confirmation sample for positive or negative results. If a
confirmation sample is required by the Department, the result must be averaged
with the first sampling result and the average is used for the compliance
determination. Results of documented sampling errors may be deleted provided
the Department approves.
(j) The
Department may reduce the total number of analyses a system must conduct by
allowing the use of composite sampling. Composite samples from a maximum of
five sampling points are allowed, provided that the detection limit of the
method used for analysis is less than one-fifth of the MCL. Compositing of
samples must be done in the laboratory and analyzed within 14 days of sample
collection.
1. If the concentration in the
composite sample is greater than or equal to 0.0005 mg/l for any contaminant
listed in Rule 0400-45-01-.25, then a follow-up sample must be taken and
analyzed within 14 days from each sampling point included in the
composite.
2. If duplicates of the
original sample taken from each sampling point used in the original composite
are available, the system may use the duplicate samples instead of resampling.
The duplicate must be analyzed and the results reported to the Department
within 14 days of collection.
3. If
the population served by the system is >3,300 persons, then compositing may
only be permitted at sampling points within the system itself. In systems
serving 3,300 or less persons compositing among different systems provided the
5-sample limit is maintained is permitted.
4. Compositing samples prior to GC analysis:
(i) Add 5 ml or equal larger amounts of each
sample (up to 5 samples are allowed) to a 25 ml glass syringe. Special
precautions must be made to maintain zero headspace in the syringe.
(ii) The samples must be cooled at 4°C
during this step to minimize volatilization losses.
(iii) Mix well and draw out a 5-ml aliquot
for analysis.
(iv) Follow sample
introduction, purging, and desorption steps described in the applicable and
approved method.
(v) If less than
five samples are used for compositing, a proportionately smaller syringe may be
used.
5. Compositing
samples prior to GC/MS analysis:
(i) Inject
5-ml or equal larger amounts of each aqueous sample (up to 5 samples are
allowed) into a 25-ml purging device using the sample introduction technique
described in the method.
(ii) The
total volume of the sample in the purging device must be 25 ml.
(iii) Purge and desorb as described in the
applicable and approved method.
(k) Compliance with paragraph (2) of Rule
0400-45-01-.25 shall be determined based on the analytical results obtained at
each sampling point. If one sampling point is in violation of an MCL, the
system is in violation of the MCL.
1. For
systems monitoring more than once per year, compliance with the MCL is
determined by a running average at each sampling point.
2. Systems monitoring annually or less
frequently whose sample result exceeds the MCL must begin quarterly sampling.
The system will not be considered in violation of the MCL until it has
completed one year of quarterly sampling.
3. If any sample result will cause the
running annual average to exceed the MCL, at any sampling point, the system is
out of compliance with the MCL immediately.
4. If a system fails to collect the required
number of samples, compliance will be based on the total number of samples
collected.
5. If a sample result is
less than the detection limit, zero will be used to calculate the annual
average.
(l) Analysis
for the contaminants listed in paragraph (2) of Rule 0400-45-01-.25 shall be
conducted using methods described in paragraphs (2) and (5) of this
rule.
(m) The Department may allow
the use of monitoring data collected after January 1, 1988, for the purposes of
monitoring compliance. If the data are generally consistent with the other
requirements in this rule, the Department may approve the use of such data,
(i.e., a single sample rather than four quarterly samples) to satisfy the
initial monitoring requirements of subparagraph (1)(d) of this rule.
(n) The Department may increase required
monitoring where necessary to detect variations in the quality of the water
produced by the system.
(o) Each
approved laboratory must determine the method detection limit (MDL) as defined
in Appendix B, 40 Code of Federal Regulations Part 136 at which it is capable
of detecting VOCs. The acceptable MDL is 0.0005 mg/l. This concentration is the
detection concentration for purposes of this rule.
(p) The Department may specify times within a
compliance period for a public water system to complete its monitoring activity
to insure that existing laboratory capacity is not exceeded.
(2) Reserved
(3) Analysis for vinyl chloride is required
only of ground water systems that have detected one or more of the following
two-carbon Organic Chemicals:
Trichloroethylene
Tetrachloroethylene
1, 2-Dichloroethane
1, 1,
1-Trichloroethane
Cis-1, 2-dichloroethylene
Trans-1, 2-dichloroethylene
1,1-Dichloroethylene
(4) The analysis for vinyl
chloride is required at each distribution or entry point at which one or more
of the two-carbon organic compounds listed in paragraph (3) of this rule were
found. If the first analysis does not detect vinyl chloride, the system may,
upon written notification to the Department, reduce the frequency of vinyl
chloride monitoring to a minimum of once every three years for that sample
location or other sample locations which are more representative of the same
source.
(5) Reserved
(6) Each community and non-transient water
system which does not detect a contaminant listed in paragraph (2) of Rule
0400-45-01-.25 may apply to the Department for a waiver from the requirement of
subparagraphs (1)(d) and (e) of this rule after completing the initial
monitoring. (For the purpose of this paragraph, detection is defined as >
0.0005 mg/l.) A waiver shall be effective for no more than six years (two
compliance periods).
A system must provide the following data and
analysis to the Department with a request for a waiver:
(a) any previous use (including transport,
storage, and/or disposal) of a contaminant within the watershed or zone of
influence;
(b) previous analytical
results;
(c) the proximity of the
system to a potential point or non-point source of contamination. Point sources
include spills and leaks of chemicals at or near a water treatment facility or
at manufacturing, distribution, or storage facilities, or from hazardous and
municipal waste landfills and other waste handling or treatment
facilities;
(d) the environmental
persistence and transport of the contaminants;
(e) the number of persons served by the
public water system and the proximity of a smaller system to a larger
system;
(f) how well the water
source is protected against contamination such as whether it is a surface or
ground water system. Ground water systems must consider factors such as depth
of the well, the type of soil, and wellhead protection. Surface water systems
must consider watershed protection; and
(g) any other information required by the
Department.
(7) The
Department shall determine a reduced monitoring frequency which the system must
implement, if the waiver is granted. The Department shall consider the
following factors to determine the reduced monitoring frequency:
(a) the watershed recharge area;
and
(b) the type of risks of
contamination to the water source.
The Department shall either grant or
deny a request for the waiver. The Department will notify the system of the
decision, in writing, with the basis for the decision and any reduced
monitoring frequency, if the waiver is granted. The term of a waiver shall not
exceed six (6) years (two compliance periods).
A system which has
been granted a waiver pursuant to this rule, must take a minimum of one (1)
sample during the term of the waiver regardless of the reduced monitoring
frequency.
A waiver for monitoring under this rule may be
initiated by the Department or upon written request of the system as set forth
in paragraph (6) of this rule. The Department shall review and, where
appropriate, revise its determination of the appropriate monitoring frequency
when the system submits new monitoring data or when other relevant data or
information becomes available. A system must reapply for a waiver for each
compliance period.
(8) As a condition of the waiver, a system
must take one sample at each sampling point during the time the waiver is
effective (i.e., one sample during two compliance periods or six years), and
update its vulnerability assessment considering the factors listed in paragraph
(7) of this rule. Based on this vulnerability assessment, the Department must
confirm that the system is non-vulnerable. If the Department does not make this
reconfirmation within three years of the initial determination, then the waiver
is invalidated and the system is required to sample annually as specified in
subparagraph (1)(c) of this rule. Surface systems shall monitor at the
frequency specified by the Department (if any).
(9) All new systems or systems that use a new
source of water must demonstrate compliance with the MCL within a period of
time specified by the Department. The system must also comply with the initial
sampling frequencies specified by the Department to ensure a system can
demonstrate compliance with the MCL. Routine and increased monitoring
frequencies shall be conducted in accordance with the requirements of this
rule.