Wash. Admin. Code § 173-408-120 - Test methods and procedures
(1)
Hydrocarbon detector specifications: Any instrument used for the measurement of
methane must be a hydrocarbon detector or other equivalent instrument approved
by the department or local authority that meets the following calibration,
specifications, and performance criteria, as applicable:
(a) EPA Reference Method 21, Determination of
Volatile Organic Compound Leaks, 40 C.F.R. Part 60, Appendix A (in effect on
the date in WAC
173-400-025), which is
incorporated by reference herein, except as follows:
(i) "Methane" replaces all references to
volatile organic compounds (VOC).
(ii) The calibration gas shall be
methane.
(b) EPA Other
Test Method 51 (OTM-51) as specified in WAC
173-408-990 (Appendix II) of this
chapter.
(c) Other approved EPA
test methods with concurrent department or local authority approval.
(2) Determination of landfill gas
heat input capacity: The landfill gas HIC must be determined in accordance with
this subsection:
(a) MSW landfills without
carbon adsorption or passive venting systems: The HIC must be calculated using
the procedure as specified in WAC
173-408-980 (Appendix I).
Additional information may be requested by the department or local authority as
necessary to verify the HIC from the MSW landfill. Site-specific data may be
substituted when available.
(b) MSW
landfills with carbon adsorption systems: The landfill gas HIC at a MSW
landfill with a carbon adsorption system must be determined by measuring:
(i) The actual total landfill gas flow rate,
in standard cubic feet per minute (scfm), using a flow meter or other flow
measuring device such as a standard pitot tube; and
(ii) The methane concentration (percent by
volume) using a hydrocarbon detector meeting the requirements of subsection (1)
of this section. The total landfill gas flow rate must be multiplied by the
methane concentration and then multiplied by the gross heating value (GHV) of
methane of 1,012 Btu/scf to determine the landfill gas HIC.
(c) MSW landfills with passive
venting systems: The landfill gas HIC at a MSW landfill with a passive venting
system must be determined using both of the following, and is the higher of
those determined values:
(i) The calculation
described in (a) of this subsection; and
(ii) The owner or operator must measure:
(A) The actual landfill gas flow rates (in
units of scfm), using a flow meter or other flow measuring device such as a
standard pitot tube; and
(B) The
methane concentration (percent by volume), using a hydrocarbon detector meeting
the requirements of subsection (1) of this section, from each venting pipe that
is within the waste mass. Each gas flow rate must then be multiplied by its
corresponding methane concentration to obtain the individual methane flow rate.
The individual methane flow rates must be added together and then multiplied by
the GHV of methane of 1,012 Btu/scf to determine the landfill gas
HIC.
(3) Surface emissions monitoring procedures:
The owner or operator of a MSW landfill must measure the landfill surface
concentration of methane using a hydrocarbon detector meeting the requirements
of subsection (1) of this section. The landfill surface must be inspected using
the following procedures:
(a) Monitoring area:
The entire landfill surface must be divided into individually identified 50,000
square foot grids. The grids must be used for both instantaneous and integrated
surface emissions monitoring.
(i) Testing must
be performed by holding the hydrocarbon detector's probe within three inches of
the landfill surface while traversing the grid, except where alternatives to
EPA Reference Method 21 are used.
(ii) The walking pattern must be no more than
25-foot spacing intervals and must traverse each monitoring grid.
(A) If the owner or operator measures no
exceedances of the limits in WAC
173-408-100(2),
after any four consecutive quarterly monitoring periods, the walking pattern
spacing may be increased to 100-foot intervals. The owner or operator must
return to a 25-foot spacing interval upon detection of any exceedances of the
limits in WAC
173-408-100(2)
that cannot be remediated within 10 calendar days or upon any exceedances
detected during a compliance inspection.
(B) If an owner or operator of a MSW landfill
can demonstrate that in the three years before the effective date of this
chapter that there were no measured exceedances of the limits in WAC
173-408-100(2),
by annual or quarterly instantaneous surface emissions monitoring, the owner or
operator may increase the walking pattern spacing to 100-foot intervals. The
owner or operator must return to a 25-foot spacing interval upon detection of
any exceedances of the limits in WAC
173-408-100(2)
that cannot be remediated within 10 calendar days, or upon any exceedances
detected during a compliance inspection. The demonstration must prove to the
satisfaction of the department or local authority that any instrument used for
methane detection meets the requirements of subsection (1) of this
section.
(iii) Surface
testing must be terminated when the average wind speed exceeds five miles per
hour, or the instantaneous wind speed exceeds 10 miles per hour. Surface
testing can continue when the average wind speed is five miles per hour or
less. The department or local authority may approve alternatives to this wind
speed surface testing termination for MSW landfills consistently having
measured winds in excess of these specified limits. Average wind speed must be
determined on a 15-minute average using an on-site anemometer with a continuous
recorder for the entire duration of the monitoring event.
(iv) Surface emissions testing must be
conducted only when there has been no measurable precipitation in the preceding
72 hours. The department or local authority may approve alternatives to this
procedure for MSW landfills that cannot meet the requirements of this
subsection.
(v) Monitoring should
be conducted during average barometric pressure conditions to the extent
possible.
(b)
Instantaneous surface emissions monitoring procedures:
(i) The owner or operator must record any
instantaneous surface readings of methane 200 ppmv or greater, other than those
measured by "nonrepeatable, momentary readings," as defined in WAC
173-408-020;
(ii) Surface areas of the MSW landfill that
exceed a methane concentration limit of 500 ppmv must be marked and remediated
in accordance with WAC
173-408-110 (1)(b) and
(c);
(iii) The entirety of landfill surface areas
with cover penetrations, distressed vegetation, cracks, or seeps must also be
inspected visually and with a hydrocarbon detector that meets the requirements
of subsection (1) of this section. Exceedances of a methane concentration limit
of 500 ppmv must be marked and remediated in accordance with WAC
173-408-110 (1)(b) and
(c);
(iv) The location of each monitored
exceedance must be marked, and the location and concentration recorded. The
location must be recorded using an instrument with an accuracy of at least 14
feet. The coordinated must be in decimal degrees with at least five decimal
places; and
(v) The wind speed and
barometric pressure must be recorded during the sampling period.
(c) Integrated surface emissions
monitoring procedures:
(i) Integrated surface
readings must be recorded and then averaged for each grid;
(ii) Individual monitoring grids that exceed
an average methane concentration of 25 ppmv must be identified and remediated
in accordance with WAC
173-408-110 (1)(b) and
(d); and
(iii) The wind speed and barometric pressure
must be recorded during the sampling period.
(4) Gas collection and control system leak
procedures: The owner or operator of a MSW landfill, or third-party owner or
operator of a landfill gas control system, must measure leaks using a
hydrocarbon detector meeting the requirements of subsection (1) of this
section.
(5) Determination of
expected gas generation flow rate: The expected gas generation flow rate must
be determined as prescribed by the 2006 Intergovernmental Panel on Climate
Change (IPCC) Guidelines for National Greenhouse Gas Inventories, Chapter 3,
which is incorporated by reference herein, using a recovery rate of 75
percent.
(6) Control device
destruction efficiency determination: The control device destruction efficiency
must be determined according to the following methods:
(a) Enclosed combustors: One of the following
test methods in 40 C.F.R., Part 60, Appendix A (in effect on the date in WAC
173-400-025), must be used to
determine the efficiency of the control device:
(i)
U. S. EPA Reference Method 18, Measurement of Gaseous Organic
Compound Emissions by Gas Chromatography (in effect on the date in WAC
173-400-025);
(ii) U. S. EPA Reference Method 25,
Determination of Total Gaseous Nonmethane Organic Emissions as Carbon (in
effect on the date in WAC
173-400-025);
(iii) U.S. EPA Reference Method 25A,
Determination of Total Gaseous Organic Concentration Using a Flame Ionization
Analyzer (in effect on the date in WAC
173-400-025); or
(iv) U. S. EPA Reference Method 25C,
Determination of Nonmethane Organic Compounds in Landfill Gases (in effect on
the date in WAC
173-400-025).
(v) The following equation must be used to
calculate destruction efficiency:
Click here to view image
(b) Open flares: Open flares must meet the
requirements of 40 C.F.R.
60.18 (in effect on the date in WAC
173-400-025).
(7) Determination of gauge
pressure: Gauge pressure must be determined using a hand-held manometer,
magnehelic gauge, or other pressure measuring device approved by the department
or local authority. The device must be calibrated and operated in accordance
with the manufacturer's specifications.
(8) Alternative test methods: Alternative
test methods may be used if they are approved in writing by the department or
local authority.
Notes
Reviser's note: The brackets and enclosed material in the text of the above section occurred in the copy filed by the agency and appear in the Register pursuant to the requirements of RCW 34.08.040.
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.