Wash. Admin. Code § 296-150V-1400 - What are the gas piping hanger and support requirements for a conversion vendor unit or medical unit?
(1) All
gas piping must be adequately supported by galvanized or equivalently protected
metal straps or hangers at intervals of not more than 4 feet, except where
adequate support and protection is provided by structural members.
(2) Gas pipe supply connections must be
rigidly anchored to a structural member within 6 inches of the supply
connections.
Notes
Statutory Authority: Chapter 43.22 RCW. 99-18-069, § 296-150V-1400, filed 8/31/99, effective 10/1/99.