Wash. Admin. Code § 296-842-22010 - Follow these fit-testing procedures for tight-fitting respirators
IMPORTANT:
Exemptions: |
This section does NOT apply to employees who: |
1. Voluntarily use respirators; or |
|
2. Are required to use mouthpiece respirators. |
Note: |
Examples of conditions that may interfere
with the |
1. Moustache, stubble, sideburns, bangs,
hairline, and other types of facial hair in areas where the
|
|
2. Temple bars of corrective eyewear or headgear that extend through the face seal area. |
Table 11
Procedure for Choosing a |
1. Inform the employee: |
a. To choose the most comfortable |
b. That each |
c. That if fitted and used properly, the
|
2. Provide a mirror and show the employee how to: |
a. Put on the |
b. Position the |
c. Set strap tension. |
Note: This instruction does NOT take the place of the employee's formal training since it is only a review. |
3. Review with the employee how to check for a comfortable fit around the nose, cheeks and other areas on the face. |
Tell the employee the |
4. Have the employee hold each facepiece against the face, taking enough time to compare the fit of each. The employee can then either: |
a. Reject any facepiece that clearly does not feel comfortable or fit adequately; or |
b. Choose which facepiece is most acceptable and which are less acceptable, if any. |
Note: |
a. Supply as many |
b. To save time later, during this step note the more acceptable facepieces in case the one chosen fails the fit test or proves unacceptable later. |
5. Have the employee wear the
most acceptable |
a. Ask the employee to observe and comment about the comfort and fit: |
i. Around the nose, cheeks, and other areas on the face |
ii. When talking or wearing eye protection |
b. Have the employee put on the |
c. Evaluate the |
i. Proper chin placement |
ii. Properly tightened straps (do NOT over tighten) |
iii. Acceptable fit across the nose bridge |
iv. |
v. To see if the |
d. Have the employee complete a successful
|
Prior to the |
i. Moving their head from side-to-side; and |
ii. Up and down. |
6. If the employee finds the |
7. Before starting the fit test, you must: |
a. Describe the fit test including screening procedures, employee responsibilities, and test exercises; and |
b. Make sure the employee wears the
|
Table 12
Isoamyl Acetate (Banana Oil) Vapor Test Procedure |
Important: |
1. This is a qualitative fit-test (QLFT) procedure |
2. The success of this test depends on preserving the employee's odor sensitivity to isoamyl acetate (IAA) vapor |
a. Vapor accumulations in ambient air can decrease odor sensitivity. To prevent this: |
i. Prepare ALL solutions in a location separate from screening and test areas |
ii. Conduct screening and tests in separate
well-ventilated rooms. For example, use an exhaust fan or laboratory
|
b. Always use odor-free water, for example, distilled or spring water that is 25°C (77°F). |
3. Isoamyl acetate is also known as isopentyl acetate. |
Important: |
Odor threshold screening determines if the employee can detect weak concentrations of IAA vapor. |
1. Choose an appropriate location to conduct screening. |
Conduct screening and tests in separate well-ventilated rooms. |
2. Prepare a stock solution AT LEAST weekly as follows: |
a. Add one milliliter (ml) of pure IAA to 800 ml of odor-free water in a one-liter glass jar with a metal lid using a measuring dropper or pipette |
b. Seal the jar with the lid and shake it for 30 seconds |
c. Clean the dropper or pipette. |
3. Prepare the odor test solution daily as follows: |
a. Add 0.4 ml from the stock solution to 500 ml of water in a one liter glass jar with a metal lid using a clean pipette or dropper |
b. Seal the jar with the lid and shake it for 30 seconds |
c. Let this solution stand for 2-3 minutes so the IAA concentration above the liquid reaches equilibrium |
d. Label this jar so you know the contents but the employee cannot know its contents, for example, "1." |
Note: |
To maintain the integrity of the test, use labels that peel off easily AND periodically switch the labels. |
4. Prepare a "test blank" solution as follows: |
a. Add 500 ml of odor-free water to a one liter glass jar with a metal lid |
b. Seal the jar |
c. Label the jar so you know the contents but the employee cannot know its contents. |
5. Type or neatly print the following instructions on a card and place it on the table in front of the two test jars: |
"The purpose of this test is to find out if you can smell banana oil at a low concentration. While both jars contain water, one ALSO contains a small amount of banana oil. |
Make sure the lid is secure then pick up a jar and shake it for two seconds. Open the jar and sniff at the opening. Repeat this for the second jar. |
Tell the individual conducting the fit test which jar contains banana oil." |
Test Preparations |
6. Choose an appropriate location to conduct fit testing. |
Conduct screening and tests in separate well-ventilated rooms. |
7. Assemble the fit test enclosure in the room. |
a. Invert a clear 55-gallon drum liner over a circular 2-foot diameter frame made of plywood or other lightweight rigid material OR construct a similar enclosure using plastic sheeting |
b. Hang the frame with the plastic covering so the top of the enclosure is about six inches above the employee's head |
c. Attach a small hook inside top center of the enclosure |
d. Tape a copy of the test exercises (see Table 19) to the inside of the test enclosure where the employee can read it. |
8. Have organic vapor cartridges or
equivalent on hand for each employee's chosen |
9. Have ready a 6 x 5-inch piece of paper towel or other porous absorbent single-ply material AND 0.75 ml of pure IAA. Do NOT apply IAA yet. |
Note: |
As an alternative to using the paper towel, you may use an IAA test swab OR ampoule if it has been demonstrated to generate an equivalent test concentration. |
Screening |
10. Have the employee, while
NOT wearing a |
a. If the employee correctly identifies the jar containing IAA, proceed to conduct testing (Step 11) |
b. If the employee is NOT able to correctly identify the jar containing IAA, you must STOP and use a different fit test protocol. |
Testing |
11. BEFORE entering the fit
test room, have the employee attach cartridges, put on, properly
adjust, and |
12.Wet the paper towel with 0.75 ml of pure IAA AND fold it in half. |
13. Pass the paper towel to the employee inside the enclosure AND instruct the employee to hang it on the hook at the top of the enclosure. |
14. Wait two minutes for the IAA vapor to fill the enclosure. |
a. While waiting, explain the fit test, including the purpose of the test exercises, the importance of cooperation, and that you must be informed if a banana-like odor is detected during the test |
b. You may also demonstrate the test exercises. |
15. Have the employee perform the appropriate fit-test exercises in Table 19. |
a. If the employee does NOT detect IAA while performing test exercises, the fit test has been PASSED. Proceed as follows: |
i. BEFORE leaving the
enclosure, have the employee break the |
ii. When exiting the employee must remove the paper towel and give it to the individual conducting the fit test. This prevents IAA vapor from building up in the enclosure during subsequent tests |
iii. The individual conducting the fit test must keep used paper towels in a self-sealing plastic bag to prevent area contamination |
b. If the employee detects IAA during any test exercise, the fit test has FAILED. STOP and have the employee do the following: |
i. Quickly return to the selection room to
remove the |
ii. Select another |
iii. Repeat screening and testing |
At this stage, if the employee fails the screening part of this procedure, the employee can repeat it AFTER waiting at least five minutes for odor sensitivity to return. |
Table 13
Saccharin Aerosol Test Procedure |
Screening Preparations |
Important: |
This is a qualitative fit-test (QLFT) procedure |
Taste threshold screening determines whether the employee being tested can detect the taste of saccharin |
The employee must NOT eat, smoke, chew gum or drink anything but plain water for at least fifteen minutes BEFORE the fit test. Sweet foods or drink consumed before the test may make the employee unable to detect saccharin during screening |
Nebulizers must be thoroughly rinsed in water and shaken dry: |
Each morning and afternoon OR |
At least every four hours. |
You may use commercially prepared solutions if they meet the requirements in this procedure. |
1. Obtain a test enclosure ( |
a. Twelve inches in diameter by fourteen inches tall |
b. A clear front portion |
c. Enough space inside to allow free movement
of the head when a |
d. A 3/4 inch (or 1.9 centimeter) hole to accommodate the nebulizer nozzle. The hole must line up in front of the wearer's nose and mouth. |
Note: |
An enclosure similar to the 3M |
This enclosure can also be used for testing. |
2. Obtain and assemble two clean DeVilbiss Model 40 Inhalation Medication Nebulizers OR equivalent. |
3. Prepare the screening solution as follows: |
a. Dissolve 830.0 milligrams of sodium saccharin USP in 100 ml of warm distilled water; or |
b. IF you have already prepared the fit-test solution, you can make the screening solution by adding 1 ml of this solution to 100 ml of distilled water. |
4. Add about 1 ml of the screening solution to one of the nebulizers. |
Mark this nebulizer to distinguish it from the one to be used for fit testing. |
Test Preparations |
5. Prepare the fit-test solution as follows: |
Add 83.0 grams of sodium saccharin to 100 ml of warm water. |
6. Add about 1 ml of the test solution to the second nebulizer. |
Mark this nebulizer to distinguish it from the one to be used for screening. |
7. Have particulate filters ready for the
employee's chosen |
Screening |
8. Have the employee, while NOT wearing a
|
9. Instruct the employee to: |
a. Breathe through a slightly open mouth with tongue extended during screening AND testing |
b. Immediately report when a sweet taste is detected. |
10. Insert the nebulizer into the front hole of the test enclosure AND administer saccharin as follows: |
a. Direct the nozzle away from the employee's nose and mouth |
b. Complete 10 squeezes in rapid succession |
c. Each time firmly squeeze the bulb so it collapses completely, then release and allow it to fully expand. |
11. Ask the employee if a sweet taste is detected. |
a. If YES, screening is completed. Proceed to conduct testing, Step 14, AFTER you: |
i. Ask the employee to remember the taste for reference during the fit test |
ii. Note the employee's taste threshold as "10" regardless of the number of squeezes actually completed |
b. If NO, screening must continue. Proceed to Step 12. |
12. Repeat with 10 more squeezes. Then follow Step 11 again; EXCEPT this time note the employee's taste threshold as "20" IF a sweet taste is reported. |
If a sweet taste is still NOT detected, repeat with 10 more squeezes and follow Step 11 one last time; EXCEPT this time note "30" for the taste threshold IF a sweet taste is reported. |
13. If NO sweet taste is reported after 30 squeezes, you must STOP and choose a different fit-test protocol for the employee. |
Test |
Important: |
Periodically check nebulizers to make sure they do not clog during use. A test is NOT valid if the nebulizer is clogged at the end of the test. |
14. Have the employee attach particulate
filters, put on, properly adjust, and |
15. Instruct the employee to immediately report if a sweet taste is detected. |
16. Insert the nebulizer into the front hole of the test enclosure AND administer the same number of squeezes, either 10, 20, or 30, as noted during screening. |
17. Have the employee perform the appropriate fit-test exercises as described in Table 19. During this step: |
a. Replenish the aerosol in the |
b. The employee must report if a sweet taste is detected: |
If NO saccharin is tasted, the test has been PASSED |
i. If saccharin is tasted the test has
FAILED, have the employee select another |
ii. Repeat screening and testing. |
Table 14
BitrexT Aerosol Test Procedure |
Important: |
This is a qualitative fit-test (QLFT) procedure |
BitrexT (denatonium benzoate) is routinely used as a taste aversion agent in household liquids that children should not drink and is endorsed by the American Medical Association, the National Safety Council, and the American Association of Poison Control Centers |
The employee must NOT eat, smoke, chew gum or drink anything but plain water for at least fifteen minutes BEFORE the fit test. |
Screening Preparations |
Important: |
Taste threshold screening determines whether the employee being tested can detect the taste of BitrexT |
Nebulizers must be thoroughly rinsed in water and shaken dry: |
Each morning and afternoon |
OR |
At least every four hours. |
You may use commercially prepared solutions if they meet the requirements in this procedure. |
1. Obtain a test enclosure that meets the following specifications: |
a. Twelve inches in diameter by fourteen inches tall |
b. A clear front portion |
c. Enough space inside the front to allow
free movement of the head when a |
d. 3/4 inch (or 1.9 centimeter) hole to accommodate the nebulizer nozzle. The hole must line up in front of the wearer's nose and mouth. |
Note: |
An enclosure similar to the 3M |
This enclosure can also be used for testing. |
2. Obtain and assemble two clean DeVilbiss Model 40 Inhalation Medication Nebulizers OR equivalent: |
3. Prepare the screening solution as follows: |
a. Make up a 5% salt solution by dissolving 5.0 grams of salt (sodium chloride) into 100 ml of distilled water |
b. Dissolve 13.5 milligrams of BitrexT in the salt solution. |
4. Add about 1 ml of the screening solution to one of the nebulizers. |
Mark this nebulizer to distinguish it from the one to be used for fit testing. |
Test Preparations |
5. Prepare the fit test solution. |
a. Dissolve 10.0 grams of salt (sodium chloride) into 200 ml of distilled water |
b. Add 337.5 milligrams of BitrexT to the warmed salt solution. |
6. Add about 1 ml of the test solution to the second nebulizer. |
Mark this nebulizer to distinguish it from the one used for screening. |
7. Have particulate filters ready for the
employee's chosen |
Screening |
Important: The employee must NOT eat, smoke, chew gum or drink anything but plain water for at least fifteen minutes BEFORE the screening and test |
8. Have the employee, while NOT
wearing a |
9. Instruct the employee to: |
a. Breathe through a slightly opened mouth with tongue extended during screening AND testing |
b. Immediately report when a bitter taste is detected. |
10. Insert the nebulizer into the front hole of the test enclosure AND administer BitrexT as follows: |
a. Direct the nozzle away from the employee's nose and mouth |
b. Complete 10 squeezes in rapid succession |
c. Each time firmly squeeze the bulb so it collapses completely, then release and allow it to fully expand. |
11. Ask the employee whether a bitter taste is detected. |
a. If YES, screening is completed. Proceed to conduct testing, Step 14, AFTER you: |
i. Ask the employee to remember the taste for reference during the fit test |
ii. Note the employee's taste threshold as "10," regardless of the number of squeezes actually completed |
b. If NO, screening must continue. Proceed to Step 12. |
12. Repeat with 10 more squeezes. Then follow Step 11 again; EXCEPT this time note the employee's taste threshold as "20" IF a bitter taste is reported. |
If a bitter taste is still NOT detected repeat with 10 more squeezes and follow Step 11 one last time; EXCEPT this time note "30" for the taste threshold IF a bitter taste is reported. |
13. If NO bitter taste is reported after 30 squeezes, you must STOP and choose a different fit-test protocol for the employee. |
Test |
14. Have the employee attach particulate
filters, put on, properly adjust, and |
15. Instruct the employee to: |
a. Breathe through a slightly opened mouth with tongue extended during screening AND testing |
b. Immediately report when a bitter taste is detected. |
16. Insert the nebulizer into the front hole of the test enclosure AND administer the same number of squeezes, either 10, 20, or 30, as noted during screening. |
17. Have the employee perform the appropriate fit-test exercises as described in Table 19. During this step: |
a. Replenish the aerosol in the |
b. The employee must report if a bitter taste is detected: |
i. If NO BitrexT is tasted, the test has been PASSED |
ii. If BitrexT is tasted the test has FAIL ED. Have the employee: |
A. Select another |
B. Repeat all screening and testing steps. |
Table 15
Irritant Smoke (Stannic Chloride) Test Procedure |
Important: |
DO NOT USE A TEST ENCLOSURE OR |
This is a qualitative fit-test (QLFT) procedure |
During this test an employee is |
Employee sensitivity varies, and certain
employees may respond more intensely than others |
Conduct fit testing in an area with adequate
ventilation to prevent |
Screening AND Test Preparations |
Important: |
Sensitivity screening is necessary to determine whether the employee can detect a weak concentration of irritant smoke AND whether any gross facepiece leakage is detected. |
1. Obtain only stannic chloride (ventilation) smoke tubes, AND an aspirator squeeze bulb OR use a low-flow air pump set to deliver 200 milliliters of air flow per minute. |
2. Equip the employee's chosen |
Screening |
Important: |
When performing sensitivity screening checks use only the MINIMUM amount of smoke necessary to elicit a response from the employee. |
3. Advise the employee that the smoke can be
irritating to eyes, lungs, and nasal passages AND
instruct the employee to keep eyes closed while |
4. Break both ends of the ventilation smoke
tube AND fit a short piece of plastic tubing, for
example, two-to-six inches of tygon tubing, over one end to prevent
|
5. While the employee is NOT
wearing a |
Carefully direct a small amount of irritant smoke toward the employee. |
Test |
Test 6. Have the employee attach |
7. Remind the employee to keep eyes closed during testing. |
8. Direct a stream of irritant smoke toward
the |
a. Begin at least 12 inches from the facepiece AND move the smoke around the whole perimeter of the mask |
b. Gradually make two more passes around the
perimeter of the facepiece, moving to within 6 inches of the
|
c. STOP at any time the employee
detects smoke in the facepiece. If this occurs a different |
9. Have the employee perform appropriate fit-test exercises in Table 19 IF the employee has NOT had an involuntary response such as evidence of coughing, flinching, or other response, OR detected smoke in the facepiece. |
Continue to direct smoke from a distance of 6 inches around the facepiece perimeter |
If smoke is detected at any time the test has
FAILED. A different |
If NO smoke is detected proceed to Step 10. |
10. Have the employee remove the |
a. Continue to use the smoke tube used for fit testing |
b. Carefully direct a SMALL amount of irritant smoke toward the employee |
i. The test has been PASSED IF the employee responds to the smoke |
ii. The fit test is VOIDED IF the employee does NOT respond to the smoke. |
Table 16
Ambient Aerosol Condensation Nuclei Counter (PortacountT) Test Procedure |
Important: |
This is a quantitative (QNFT) fit-test procedure |
This method uses a particle counting
instrument that measures and compares the particle concentration both
inside and outside the |
Particles in the ambient air are used as the test aerosol. |
Test Preparations |
1. Obtain a test instrument such as a PortacountT. |
2. Have probed respirators available for
each |
Note: A probed |
3. Follow the test instrument manufacturer's instructions for test preparation, including particle, zero, and system checks. Make sure the instrument's pass OR fail criterion is programmed to the following MINIMUM performance levels: |
a. For half-facepiece respirators, an
overall minimum |
b. For full-facepiece respirators, an
overall minimum |
4. Have high-efficiency particulate air
(HEPA) filters, OR other |
If you will use a sampling adapter instead of probed respirators be sure to have the correct type for the respirators chosen. |
Test |
5. Properly attach the sampling line to the facepiece probe or sampling adapter. |
6. Have the employee attach |
a. Proper chin placement |
b. Properly tightened straps (do NOT over tighten) |
c. Acceptable fit across the nose bridge |
d. |
e. To see if the |
Note: Wearing the |
7. Have the employee perform a |
a. Determine the cause; and |
b. If leakage is due to a poorly fitting facepiece, have the employee: |
i. Choose another |
ii. Start again at Step 6. |
8. Start the fit test cycle. |
a. Follow the manufacturer's instructions for operating the test instrument |
b. Have the employee perform the appropriate fittest exercises in Table 19 |
The test instrument will automatically stop
and calculate the overall |
The test has been PASSED if the
overall |
The test has FAILED if the
overall |
Note: If the test has failed, have the employee
select another |
Table 17
Controlled Negative Pressure (CNP) Test Procedure |
Important: |
This is a quantitative fit-test (QNFT) procedure |
This method determines |
Instruments used must have a nonadjustable test pressure of 15.0 mm water pressure |
Measurements occur while employees remain still AND hold their breath for 10 seconds |
No test aerosols are used. |
Test Preparations |
1. Make sure the individual conducting the fit test is thoroughly trained to perform this test. |
2. Obtain a CNP test instrument such as a Fit-Tester 3000T. Make sure: |
a. Defaults are set at: |
i. -15mm (-0.58 inches) of water test pressure; and |
ii. A modeled inspiratory flow rate of 53.8 liters per minute |
b. It has an effective audio warning device or visual screen tracing that signals when employees fail to hold their breath. |
Note: You are not required to obtain test recording and printing equipment such as computers OR printers. Hand recording results is acceptable |
To see default settings, check the instrument's "REDON protocol." |
3. Obtain facepiece adapters appropriate for
each test |
Note: Adapters are either a
one-piece (for SCBA facepieces), OR two-piece (for dual cartridge
facepieces) device providing a manifold and breathing valve system.
For positive pressure respirators, you will need to obtain an
additional fitting, available from the |
To obtain adapters, contact the CNP
instrument's distributor, Occupational Health Dynamics,
OR the |
Test |
Important: |
The |
4. Explain the test procedure to the employee. |
5. Train the employee on how to hold a breath for at least 10 seconds. |
6. Prepare the |
a. Remove or prop open the inhalation valves. If a breathing tube is present, disconnect it |
b. Replace cartridges, if present, with the manifold and breathing valve adapter |
For positive pressure facepieces, mount the manufacturer's additional fitting followed by the manifold-breathing valve adapter |
c. Connect the |
7. Have the employee put on, adjust, and
|
8. Turn on the instrument AND have the employee stand and perform the fit-test exercises in Table 19. Once exercises begin, any adjustments will void the test and you must begin again. |
9. Once test exercises are completed, ask
the employee about facepiece comfort. If the employee states the
|
10. Determine the overall |
Overall |
Where: |
n = The number of exercises; |
ffE1 = The |
ffE2 = The |
ffE3 = The |
ffEn = The |
The test is PASSED IF the
overall |
The test has FAILED IF the |
If the test has FAILED you must
have the employee select another |
Table 18
Generated Aerosol Test Procedure |
Important: |
* This is a quantitative (QNFT) fit-test procedure |
* In this method, a test aerosol is used to challenge the facepiece seal while aerosol concentrations inside and outside the facepiece are measured during test exercises |
* Special equipment is needed to generate, disperse, detect, and measure test aerosols. |
Test Preparations |
1. Test aerosol. |
Use a particulate, for example, corn oil, polyethylene glycol 400, di-2-ethyl hexyl sebacate, or sodium chloride. |
2. Instrumentation. |
Do ALL the following: |
a. Obtain and use aerosol generation, dilution, and measurement systems appropriate for particulates |
b. Use an aerosol-generating instrument that will maintain test concentrations within a 10% variation |
c. Select a sampling instrument that allows for a computer record or strip chart record to be created |
The record must show the rise and fall of test agent concentration during each inhalation and exhalation at fit factors of at least 2000. |
Note: Integrators, or computers
that integrate the amount of test agent penetration leakage into the
|
d. Minimize the time interval between the activity and the recording of the activity so you can clearly connect what you see to what is being recorded. For example, use a small diameter and length of sampling line. |
3. Test enclosure. |
Do ALL the following: |
a. Make sure the enclosure is equipped and constructed to effectively: |
i. Maintain a uniform concentration of the test agent inside the enclosure. For example, the enclosure must be large enough to allow ALL employees freedom of movement during testing WITHOUT disturbing the test concentration or measurement instrument |
ii. Keep the test agent from contaminating
the air outside the enclosure. For example, use a HEPA |
iii. Allow the individual conducting the fit test to view the employee during the test |
b. Make sure the tubing used to collect
samples from the enclosure AND |
c. If sodium chloride is used, relative humidity inside the enclosure must be kept below 50%. |
4. Prepare test respirators. |
Do ALL the following: |
a. Inspect test respirators regularly for missing parts AND damage |
b. Keep test respirators in proper working order |
c. Make sure in-mask sampling probes are: |
i. Designed and installed so the air sample will be drawn from the employee's breathing zone; midway between the nose and mouth; and |
ii. The probe extends inside the facepiece at least 1/4 inch |
d. Make sure sampling ports such as probes, or adapters on respirators are constructed and installed so they do NOT: |
i. Block air flow into the sampling line |
ii. Leak |
iii. Interfere with the |
Have high efficiency particulate air (HEPA)
filters OR P100 series |
Replace filters when increased breathing
resistance is detected OR when the test agent has
altered the |
Test |
Important: |
Throughout the test, maintain the employee's
|
If a single peak penetration exceeds 5% for half facepieces OR 1% for full facepieces: |
STOP the test; and |
Have the employee select another |
5. Have the employee attach filters, put on,
adjust, and |
a. Be sure to crimp the sampling line to
avoid pressure leaks during the |
b. Have the employee adjust the |
6. OPTIONAL Step. To save time conduct a screening test to quickly identify poorly fitting respirators. |
Note: You may use a qualitative screening test OR an ambient aerosol condensation nuclei counter instrument in the count mode. |
7. Make sure test aerosol concentration is reasonably stable. |
If a canopy or shower curtain enclosure is used, determine stability of the test aerosol concentration AFTER the employee enters the enclosure. |
8. Have the employee enter the test
enclosure and connect the |
9. Immediately after entering the enclosure
measure test aerosol concentration inside the |
Make sure the peak penetration does NOT exceed 5% for half facepieces, OR 1% for full facepieces. |
10. Have employee perform the appropriate fit-test exercises in Table 19. |
Do NOT adjust the |
11. Calculate the overall |
a. PASSED IF the minimum |
b. IF a passing |
Calculations |
Important: |
Do NOT count the grimace exercise measurements during these calculations |
Take into account the limitations of instrument detection when determining fit factors. |
12. Calculate individual fit factors for EACH exercise by applying the following: |
Exercise |
a. To determine the average test enclosure concentration use one of the following methods: |
i. Arithmetic average of the concentration before and after each test (an average of two values per entire test) |
ii. Arithmetic average of concentration before and after each exercise (an average of two values per exercise) |
iii. True average measured continuously
during the |
b. Determine the test aerosol concentration
inside the |
Average peak penetration values. Determine aerosol penetration for each exercise by: |
i. Using integrators or computers that calculate the actual test agent penetration; or |
ii. Average the peak heights shown on the strip chart recording, graph, or by computer integration |
c. Maximum peak penetration. Use strip chart recordings to determine the highest peak penetration for each exercise and use this value |
d. Area under the peaks. Use computerized integration or other appropriate calculations to integrate the area under individual peaks for each exercise. |
13. Using individual exercise fit factors
(ffE) calculate the overall |
a. Convert each exercise |
b. Determine the average penetration value; |
c. Convert the average penetration value
back to a |
Use this equation to calculate the
overall |
Overall |
Where: |
n = The number of exercises; |
ffE1 = The |
ffE2 = The |
ffE3 = The |
ffEn = The |
Table 19
Fit-Test Exercises |
|||
Important: |
|||
This list applies when you use any fit test |
|||
Employees tested must perform ALL exercises marked with an "X" as described for the fit-test procedure used |
|||
Once exercises begin, any adjustments made void the test AND you must begin again |
|||
After test exercises are completed, you must
ask the employee about the comfort of the |
|||
When the controlled negative pressure
procedure is used, STOP and repeat the test if the
employee adjusts the |
|||
Controlled negative pressure tests conducted according to the method published in 29 C.F.R. 1910.134, Appendix A are an acceptable alternative to the method outlined below. |
|||
Description of Required Fit-Test Exercises |
Fit-Test Procedures |
||
Qualitative Procedures |
Quantitative Procedures; EXCEPT the CNPP |
Controlled Negative Pressure Procedure (CNPP) |
|
Normal breathing |
|||
Breathe normally, while standing for one minute |
X |
X |
|
Deep breathing |
|||
Breathe slowly and deeply while standing for one minute |
X |
X |
|
Take caution to avoid hyperventilating |
|||
Head side to side |
|||
Slowly turn head from side to side while standing for one minute, pausing at each extreme position to inhale |
X |
X |
|
Be careful to NOT bump the
|
|||
Head up and down |
|||
Slowly move head up and down while standing for one minute, inhaling in the up position |
X |
X |
|
Be careful to NOT bump the
|
|||
Talking |
|||
Talk slowly and loud enough to be heard clearly by the individual conducting fit testing for one minute. Choose ONE of the following: |
|||
Read from a prepared text such as the Rainbow Passage1 |
X |
X |
|
Count backward from 100 |
|||
Recite a memorized poem or song. |
|||
Grimace |
|||
Smile or frown for fifteen seconds. |
X |
||
Bending over |
|||
Bend over to touch toes while standing. Repeat at a comfortable pace for one minute |
|||
OR |
X |
X |
|
Jog in place for one minute if the test
enclosure, such as a |
|||
Normal breathing |
|||
Breathe normally while standing for one minute |
X |
X |
|
Face forward |
|||
Premeasurement activity: Stand and breath normally, without talking, for 30 seconds |
X |
||
Measurement position: Face forward while holding breath for 10 seconds |
|||
Bending over |
|||
Premeasurement activity: While standing, bend at the waist, as if to touch toes |
X |
||
Measurement position: Hold the bending position with face parallel to the floor while holding breath for 10 seconds |
|||
Head shaking |
|||
Premeasurement activity: Vigorously shake head from side to side for about 3 seconds while shouting |
X |
||
Measurement position: Face forward, while holding breath for 10 seconds |
|||
Redon-1 |
|||
Premeasurement activity: Loosen
all facepiece straps and remove the |
X |
||
Measurement position: Face forward while holding breath for 10 seconds |
|||
Redon-2 |
|||
Repeat the premeasurement activity and measurement position described in Redon-1 |
X |
1The Rainbow Passage:
"When the sunlight strikes raindrops in the air, they act like a prism and form a rainbow. The rainbow is a division of white light into many beautiful colors. These take the shape of a long round arch, with its path high above, and its two ends apparently beyond the horizon. There is, according to legend, a boiling pot of gold at one end. People look, but no one ever finds it. When a man looks for something beyond reach, his friends say he is looking for the pot of gold at the end of the rainbow."
Notes
Statutory Authority: RCW 49.17.050 and 29 C.F.R. Subpart Z. 09-19-119, § 296-842-22010, filed 9/22/09, effective 12/1/09. Statutory Authority: RCW 49.17.010, 49.17.040, 49.17.050, 49.17.060. 07-05-072, § 296-842-22010, filed 2/20/07, effective 4/1/07; 03-20-114, § 296-842-22010, filed 10/1/03, effective 1/1/04.
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.