7.2. Definitions - As used in this Section,
the following definitions apply:
7.2.1.
Accessible Surface - the external surface of the enclosure or housing of the
radiation producing machine as provided by the manufacturer.
7.2.2. Added Filtration - any filtration
which is in addition to the inherent filtration.
7.2.3. Aluminum Equivalent - the thickness of
aluminum (type one thousand one hundred [1100] alloy) affording the same
attenuation, under specified conditions, as the material in question. (The
nominal chemical composition of type one thousand one hundred [1100] aluminum
alloy is ninety-nine [99.00] percent minimum aluminum, twelve one-hundredths
[0.12] percent copper.)
7.2.4.
Assembler - any person engaged in the business of assembling, replacing, or
installing one or more components into an x-ray system or subsystem. The term
includes the owner of an x-ray system or his or her employee or agent who
assembles components into an x-ray system that is subsequently used to provide
professional or commercial services.
7.2.5. Attenuation Block - a block or stack,
having dimensions twenty (20) centimeters by twenty (20) centimeters by three
and eight-tenths (3.8) centimeters, of type 1100 aluminum alloy or other
materials having equivalent attenuation.
7.2.6. Automatic Exposure Control (AEC) - a
device which automatically controls one or more technique factors in order to
obtain at a preselected locations a required quantity of radiation (Includes
devices such as phototimers and ion chambers).
7.2.7. Barrier - the material, excluding
filters, placed in the useful and scattered beam, for protection purposes, to
reduce the radiation exposure.
7.2.8. Beam Axis - a line from the source
through the centers of the x-ray fields.
7.2.9. Beam-limiting device - a device which
provides a means to restrict the dimensions of the x-ray field.
7.2.10. C-arm X-ray System - an x-ray system
in which the image receptor and x-ray tube housing assembly are connected by a
common mechanical support system in order to maintain a desired spatial
relationship. This system is designed to allow a change in the projection of
the beam through the patient without a change in the position of the
patient.
7.2.11. Cephalometric
Device - a device intended for the radiographic visualization and measurement
of the dimensions of the human head.
7.2.12. Certified Components - components of
x-ray systems which are subject to regulations promulgated under
Public Law
90-602, the Radiation Control for Health and Safety Act of 1968, the Food and
Drug Administration.
7.2.13.
Certified System - any x-ray system which has one or more certified
components.
7.2.14. Changeable
Filters - any filter, exclusive of inherent filtration, which can be removed
from the useful beam through any electronic, mechanical, or physical
process.
7.2.15. Coefficient of
Variation - the ratio of the standard deviation to the mean value of a set of
observations. It is estimated using the following equation:
where:
{FUNC s} = Standard deviation of the observed values;
{OVERLINE FUNC ALIGNC x} = Mean value of observations in
sample;
{FUNC x SUB i} = ith observation in
sample;
{FUNC n} = Number of observations in sample.
7.2.16. Collimator - a device or
mechanism by which the x-ray beam is restricted in size.
7.2.17. Computed Tomography (CT) - the
production of a tomogram by the acquisition and computer processing of x-ray
transmission data.
7.2.18. Control
Panel - that part of the x-ray control upon which are mounted the switches,
knobs, pushbuttons, and other hardware necessary for manually setting the
technique factors.
7.2.19. Cooling
Curve - the graphical relationship between heat units stored and cooling
time.
7.2.20. Dead-man Switch - a
switch so constructed that circuit-closing contact can be maintained only be
continuous pressure on the switch by the operator.
7.2.21. Detector - a device which in the
presence of radiation provides a signal or other indication suitable for use in
measuring one or more quantities of incident radiation.
7.2.22. Diagnostic-type Protective Tube
Housing - an x-ray tube housing so constructed that the leakage radiation at a
distance of one (1) meter from the target cannot exceed one hundred (100)
milliroentgens in one (1) hour when the tube is operated at its maximum
continuous rated current for the maximum rated tube potential.
7.2.23. Diagnostic Source Assembly - the tube
housing assembly with a beam-limiting device attached.
7.2.24. Diagnostic X-ray System - an x-ray
system designed for irradiation of any part of the human or animal body for the
purpose of diagnosis or visualization.
7.2.25. Diagnostic X-ray Imaging System - an
assemblage of components for the generation, emission and reception of x-rays
and the transformation, storage and visual display of the resultant x-ray
image.
7.2.26. Diaphragm - a device
or mechanism by which the x-ray beam is restricted in size.
7.2.27. Direct Scattered Radiation - that
scattered radiation which has been deviated in direction only by materials
irradiated by the useful beam (See "Scattered Radiation").
7.2.28. Entrance Exposure Rate - the exposure
free in air per unit time at the point where the center of the useful beam
enters the patient.
7.2.29. Field
Emission Equipment - equipment which uses an x-ray tube in which electron
emission from the cathode is due solely to the action of an electric
field.
7.2.30. Filter - the
material placed in the useful beam to absorb preferentially selected
radiations.
7.2.31. Fluoroscopic
Imaging Assembly - a subsystem in which x-ray photons produce a visible image.
It includes the image receptor or receptors such as the image intensifier and
spot-film device, electrical interlocks, if any, and structural material
providing linkage between the image receptor and diagnostic source
assembly.
7.2.32. Focal Spot
(Actual) - the area projected on the anode on the x-ray tube bombarded by
electrons accelerated from the cathode and from which the useful beam
originates.
7.2.33. General Purpose
Radiographic X-ray System - any radiographic x-ray system which, by design, is
not limited to radiographic examination of specific anatomical
regions.
7.2.34. Gonadal Shield - a
protective barrier for the testes or ovaries.
7.2.35. Half-value Layer (HVL) - the
thickness of specified material which attenuates the beam of radiation to an
extent such that the exposure rate is reduced to one-half of its original
value. In this definition, the contribution of all scattered radiation, other
than any which might be present initially in the beam concerned, is deemed to
be excluded.
7.2.36. Healing Arts
Screening - the testing of human beings using x-ray machines for the detection
or evaluation of health indications when such tests are not specifically and
individually ordered by a licensed practitioner of the healing arts legally
authorized to prescribe such x-ray tests for the purpose of diagnosis or
treatment.
7.2.37. Heat Unit - a
unit of energy equal to the product of the peak kilovoltage, milliamperes, and
seconds, i.e., kVp x mA x second.
7.2.38. Image Intensifier - a device,
installed in its housing, which instantaneously converts an x-ray pattern into
a corresponding light image of higher intensity.
7.2.39. Image Receptor - any device, such as
a fluorescent screen or radiographic film, which transforms incident x-ray
photons either into a visible image or into another form which can be made into
a visible image by further transformations.
7.2.40. Image Receptor Support - for
mammographic systems, that part of the system designed to support the image
receptor during mammography.
7.2.41. Inherent Filtration - the filtration
of the useful beam provided by the permanently installed components of the tube
housing assembly.
7.2.42. Interlock
- a device for precluding access to an area of radiation hazard either by
preventing entry or by automatically removing the hazard.
7.2.43. Irradiation - the exposure of matter
to ionizing radiation.
7.2.44. kV -
kilovolts.
7.2.45. kVp - the
maximum value of the potential difference across the x-ray tube during an
exposure.
7.2.46. kWs - kilowatt
second.
7.2.47. Lead Equivalent -
the thickness of lead affording the same attenuation, under specified
conditions, as the material in question.
7.2.48. Leakage Technique Factors - the
technique factors associated with the diagnostic source assembly which are used
in measuring leakage radiation. They are defined as follows:
7.2.48.a. For diagnostic source assemblies
intended for capacitor energy storage equipment, the maximum-rated peak tube
potential and the maximum-rated number of exposures in an hour for operation at
the maximum-rated peak tube potential with the quantity of charge per exposure
being ten (10) millicoulombs, i.e., ten (10) milliampere seconds, or the
minimum obtainable from the unit, whichever is larger;
7.2.48.b. For diagnostic source assemblies
intended for field emission equipment rated for pulsed operation, the
maximum-rated peak tube potential and the maximum-rated number of x-ray pulses
in an hour for operation at the maximum-rated peak tube potential;
7.2.48.c. For all other diagnostic source
assemblies, the maximum-rated peak tube potential and the maximum-rated
continuous tube current for the maximum-rated peak tube potential.
7.2.49. Light Field - that area of
the intersection of the light beam from the beam-limiting device and one (1) of
the set of planes parallel to and including the plane of the image receptor,
whose perimeter is the locus of points at which the illumination is one-fourth
(1/4) of the maximum in the
intersection.
7.2.50. Line-voltage
Regulation - the difference between the no-load and the load line potentials
expressed as a percent of the load line potential. It is calculated using the
following equation:
Percent line-voltage regulation = 100
(Vn-Vl)/Vl
where:
Vn = No-load line potential;
and
Vl = Load line potential.
7.2.51. mA -
milliampere.
7.2.52. mAs -
milliampere second.
7.2.53. Maximum
Line Current - the root-mean-square current in the supply line of an x-ray
machine operating at its maximum rating.
7.2.54. Mobile X-ray Equipment - x-ray
equipment mounted on a permanent base with wheels or casters for moving while
completely assembled.
7.2.55.
Patient - an individual or animal subjected to healing arts examination,
diagnosis, or treatment.
7.2.56.
PBL - See Positive Beam Limitation.
7.2.57. Peak Tube Potential - the maximum
value of the potential difference across the x-ray tube during an
exposure.
7.2.58. Phantom - a
volume of material behaving in a manner similar to tissue with respect to the
attenuation and scattering of radiation. This requires that both the atomic
number (Z) and the density of the material be similar to that of
tissue.
7.2.59. PID - a device on
dental x-ray equipment used to indicate the beam position and to establish a
definite source-surface (skin) distance. It may or may not incorporate or serve
as a beam-limiting device.
7.2.60.
Portable X-ray Equipment - x-ray equipment designed to be
hand-carried.
7.2.61. Position
Indicating Device - a device on dental x-ray equipment used to indicate the
beam position and to establish a definite source-surface (skin) distance. It
may or may not incorporate or serve as a beam-limiting device.
7.2.62. Positive Beam Limitation - the
automatic or semi-automatic adjustment of an x-ray beam to the size of the
selected image receptor, whereby exposures cannot be made without such
adjustment.
7.2.63. Primary
Protective Barrier - the material, excluding filters, placed in the useful
beam, for protection purposes, to reduce the radiation exposure.
7.2.64. Protective Apron - an apron made of
radiation absorbing materials used to reduce radiation exposure.
7.2.65. Protective Glove - a glove made of
radiation absorbing materials used to reduce radiation exposure.
7.2.66. Qualified Expert - an individual who
has demonstrated to the satisfaction of the agency that he possesses the
knowledge and training to measure ionizing radiation, to evaluate safety
techniques, and to advise regarding radiation protective needs.
7.2.67. Radiation Detector - a device which
in the presence of radiation provides a signal or other indication suitable for
use in measuring one or more quantities of incident radiation.
7.2.68. Radiation Therapy Simulation System -
a radiographic or fluoroscopic x-ray system intended for localizing the volume
to be exposed during radiation therapy and confirming the position and size of
the therapeutic irradiation field.
7.2.69. Radiograph - an image receptor on
which the image is created directly or indirectly by an x-ray pattern and
results in a permanent record.
7.2.70. Radiographic Imaging System - any
system whereby a permanent or semi-permanent image is recorded on an image
receptor by the action of ionizing radiation.
7.2.71. Rating - the operating limits as
specified by the component manufacturer.
7.2.72. Recording - producing a permanent
form of an image resulting from x-ray photons.
7.2.73. Registrant as used in this Section -
any person who owns or possesses and administratively controls and x-ray system
which is used to deliberately expose humans or animals to the useful beam of
the system and is required by the provisions of Sections 4. and 5. of this rule
to register with this agency.
7.2.74. Scattered Radiation - radiation that,
during passage through matter, has been deviated in direction.
7.2.75. Secondary Protective Barrier - a
barrier sufficient to attenuate the stray radiation to the required
degree.
7.2.76. Shutter - a device
attached to the tube housing assembly which can intercept the entire cross
Sectional area of the useful beam and which has a lead equivalency not less
than that of the tube housing assembly.
7.2.77. Source-image Distance (SID) - the
distance from the source to the center of the input surface of the image
receptor.
7.2.78. Source - the
focal spot of the x-ray tube.
7.2.79. Spot Film - a radiograph which is
made during a fluoroscopic examination to permanently record conditions which
exist during that fluoroscopic procedure.
7.2.80. Spot-film Device - a device intended
to transport or position a radiographic image receptor between the x-ray source
and fluoroscopic image receptor. It includes a device intended to hold a
cassette over the input end of an image intensifier for the purpose of making a
radiograph.
7.2.81. SSD - the
distance between the source and the skin entrance plane of the
patient.
7.2.82. Stationary X-ray
Equipment - x-ray equipment which is installed in a fixed location.
7.2.83. Stray Radiation - the sum of leakage
and scattered radiation.
7.2.84.
Technique Factors - the following conditions of operation:
7.2.84.a. For capacitor energy storage
equipment, peak tube potential in kV and quantity of charge in mAs;
7.2.84.b. For field emission equipment rated
for pulsed operation, peak tube potential in kV, and number of x-ray
pulses;
7.2.84.c. For CT x-ray
systems designed for pulsed operation, peak tube potential in kV, scan time in
seconds, and either tube current in mA, x-ray pulse width in seconds, and the
number of x-ray pulses per scan, or the product of tube current, x-ray pulse
width, and the number of x-ray pulses in mAs;
7.2.84.d. For CT x-ray systems not designed
for pulsed operation, peak tube potential in kV, and either tube current in mA
and scan time in seconds, or the product of tube current and exposure time in
mAs and the scan time when the scan time and exposure time are equivalent;
and
7.2.84.e. For all other
equipment, peak tube potential in kV, and either tube current in mA and
exposure time in seconds, or the product of tube current and exposure time in
mAs.
7.2.85. Termination
of Irradiation - the stopping of irradiation in a fashion which will not permit
continuance of irradiation without the resetting of operating conditions at the
control panel.
7.2.86.
Therapeutic-type Tube Housing:
7.2.86.a. For
x-ray therapy equipment not capable of operating at five hundred (500) kVp or
above, the following definition applies: An x-ray tube housing so constructed
that the leakage radiation averaged over any one hundred (100)
cm2 area at a distance of one meter from the source
does not exceed one (1) Roentgen in an hour when the tube is operated at its
maximum rated continuous current for the maximum rated tube
potential.
7.2.86.b. For x-ray
therapy equipment capable of operating at five hundred (500) kVp or above, the
following definition applies: An x-ray tube housing so constructed that the
leakage radiation averaged over any one hundred (100)
cm2 area at a distance of one meter from the source
does not exceed one-tenth (0.1) percent of useful beam dose rate at one meter
from the source for any of its operating conditions.
7.2.87. Tomogram - the depiction of the x-ray
attenuation properties of a Section through the body.
7.2.88. Tube Housing Assembly - the tube
housing with tube installed. It includes high-voltage or filament transformers
and other appropriate elements when such are contained within the tube
housing.
7.2.89. Tube Rating Chart
- the set of curves which specify the rated limits of operation of the tube in
terms of the technique factors.
7.2.90. Useful Beam - the radiation which
passes through the tube housing port and the aperture of the beam-limiting
device when the exposure switch or timer is activated.
7.2.91. Variable-aperture Beam-limiting
Device - a beam-limiting device which has capacity for stepless adjustment of
the x-ray field size at a given SID.
7.2.92. Visible Area - that portion of the
input surface of the image receptor over which incident x-ray photons are
producing a visible image.
7.2.93.
X-ray Exposure Control - a device, switch, button or other similar means by
which an operator initiates or terminates the radiation exposure. The x-ray
exposure control may include such associated equipment as timers and back-up
timers.
7.2.94. X-ray Equipment -
an x-ray system, subsystem, or component thereof. Types of x-ray equipment are
as follows:
7.2.94.a. Mobile X-ray Equipment -
x-ray equipment mounted on a permanent base with wheels or casters for moving
while completely assembled.
7.2.94.b. Portable X-ray Equipment - x-ray
equipment designed to be hand-carried.
7.2.94.c. Stationary X-ray Equipment - x-ray
equipment which is installed in a fixed location.
7.2.95. X-ray Field - that area of the
intersection of the useful beam and any one of the set of planes parallel to
and including the plane of the image receptor, whose perimeter is the locus of
points at which the exposure rate is one-fourth of the maximum in the
intersection.
7.2.96. X-ray
High-voltage Generator - a device which transforms electrical energy from the
potential supplied by the x-ray control to the tube operating potential. The
device may also include means for transforming alternating current to direct
current, filament transformers for the x-ray tubes, high-voltage switches,
electrical protective devices, and other appropriate elements.
7.2.97. X-ray System - an assemblage of
components for the controlled production of x-rays. It includes minimally an
x-ray high-voltage generator, an x-ray control, a tube housing assembly, a
beam-limiting device, and the necessary supporting structures. Additional
components which function with the system are considered integral parts of the
system.
7.2.98. X-ray Table - a
patient support device with its patient support structure (tabletop) interposed
between the patient and the image receptor during radiography or fluoroscopy.
This includes, but is not limited to, any stretcher equipped with a radiolucent
panel and any Table equipped with a cassette tray (or bucky), cassette tunnel,
image intensifier, or spot-film device beneath the Tabletop.
7.2.99. X-ray Tube - any electron tube which
is designed for the conversion of electrical energy into x-ray
energy.
7.12. Therapeutic
Radiation Machines
7.12.a. Purpose and Scope.
7.12.a.1. This Section establishes
requirements, for which the registrant is responsible, for use of therapeutic
radiation machines. The provisions of this Section are in addition to, and not
in substitution for, other applicable provisions of this rule.
7.12.a.2. The use of therapeutic radiation
machines shall be by, or under the supervision of, a licensed practitioner of
the healing arts who meets the training and experience criteria established by
Subdivision 12.3.c.
7.12.b. Definitions. As used in this Section,
the following definitions apply:
7.12.b.1.
Absorbed Dose (D) - the mean energy imparted by ionizing radiation to matter.
Absorbed dose is determined as the quotient of dE by dM, where dE is the mean
energy imparted by ionizing radiation to matter of mass dM.. The SI unit of
absorbed dose is joule per kilogram and the special name of the unit of
absorbed dose is the Gray (Gy). The previously used special unit of absorbed
dose (rad) is being replaced by the gray.
7.12.b.2. Absorbed Dose Rate - absorbed dose
per unit time, for machines with timers, or dose monitor unit per unit time for
linear accelerators.
7.12.b.3.
Accessible Surface - surface of equipment or of an equipment part that can be
easily or accidentally touched by persons without the use of a tool.
7.12.b.4. Added Filtration - any filtration
which is in addition to the inherent filtration.
7.12.b.5. Air Kerma (K) - the kinetic energy
released in air by ionizing radiation. Kerma is determined as the quotient of
dE by dM, where dE is the sum of the initial kinetic energies of all the
charged ionizing particles liberated by uncharged ionizing particles in air of
mass dM.. The SI unit of air kerma is joule per kilogram and the special name
for the unit of kerma is the gray (Gy).
7.12.b.6. Barrier - a barrier of radiation
absorbing materials used to reduce radiation exposure.
7.12.b.7. Beam Axis - the axis of rotation of
the beam limiting device.
7.12.b.8.
Beam-limiting Device - a field defining collimator, integral to the therapeutic
radiation machine, which provides a means to restrict the dimensions of the
useful beam.
7.12.b.9. Beam
Monitoring System - a system designed and installed in the radiation head to
detect and measure the radiation present in the useful beam.
7.12.b.10. Beam Scattering Foil - a thin
piece of material (usually metallic) placed in the beam to scatter a beam of
electrons in order to provide a more uniform electron distribution in the
useful beam.
7.12.b.11. Bent Beam
Linear Accelerator - a linear accelerator geometry in which the accelerated
electron beam must change direction by passing through a bending
magnet.
7.12.b.12. Changeable
Filters - any filter, exclusive of inherent filtration, which can be removed
from the useful beam through any electronic, mechanical, or physical
process.
7.12.b.13. Contact Therapy
System - a therapeutic radiation machine with a short target to skin distance
(TSD), usually less than five (5) centimeters.
7.12.b.14. Detector - a device which, in the
presence of radiation provides, by either direct or indirect means, a signal or
other indication suitable for use in measuring one or more quantities of
incident radiation.
7.12.b.15. Dose
Monitor Unit (DMU) - a unit response from the beam monitoring system from which
the absorbed dose can be calculated.
7.12.b.16. External Beam Radiation Therapy -
therapeutic irradiation in which the source of radiation is at a distance from
the body.
7.12.b.17.
Field-flattening Filter - a filter used to homogenize the absorbed dose rate
over the radiation field.
7.12.b.18. Filter - material placed in the
useful beam to change beam quality in therapeutic radiation machines subject to
Subdivision 7.12.f.
7.12.b.19.
Gantry - that part of a radiation therapy system supporting and allowing
movements of the radiation head about a center of rotation.
7.12.b.20. Gray (Gy) - the SI unit of
absorbed dose, kerma, and specific energy imparted equal to 1 joule per
kilogram. The previous unit of absorbed dose (rad) is being replaced by the
Gray. (one [1] Gy=one hundred [100] rad).
7.12.b.21. Half-value Layer (HVL) - the
thickness of a specified material which attenuates x-radiation or gamma
radiation to an extent such that the air kerma rate, exposure rate or absorbed
dose rate is reduced to one-half (½) of the value
measured without the material at the same point.
7.12.b.22. Interlock - a device preventing
the start or continued operation of equipment unless certain predetermined
conditions prevail.
7.12.b.23.
Interruption of Irradiation - the stopping of irradiation with the possibility
of continuing irradiation without resetting of operating conditions at the
control panel.
7.12.b.24.
Irradiation - the exposure of a living being or matter to ionizing
radiation.
7.12.b.25. Isocenter -
the center of the sphere through which the useful beam axis passes while the
gantry moves through its full range of motions.
7.12.b.26. Kilovolt (kV) [kilo electron volt
(keV)] - the energy equal to that acquired by a particle with one electron
charge in passing through a potential difference of one thousand volts in a
vacuum. [Note: current convention is to use kV for photons and keV for
electrons.]
7.12.b.27. Lead
Equivalent - the thickness of the material in question affording the same
attenuation, under specified conditions, as lead.
7.12.b.28. Leakage Radiation - radiation
emanating from the radiation therapy system except for the useful
beam.
7.12.b.29. Light Field - the
area illuminated by light, simulating the radiation field.
7.12.b.30. mA - milliampere.
7.12.b.31. Megavolt (MV) [mega electron volt
(MeV)] - the energy equal to that acquired by a particle with one electron
charge in passing through a potential difference of one million volts in a
vacuum. [Note: current convention is to use MV for photons and MeV for
electrons.]
7.12.b.32. Monitor Unit
(MU) -a unit response from the beam monitoring system from which the absorbed
dose can be calculated.
7.12.b.31.
Moving Beam Radiation Therapy - radiation therapy with any planned displacement
of radiation field or patient relative to each other, or with any planned
change of absorbed dose distribution. It includes arc, skip, conformal,
intensity modulation and rotational therapy.
7.12.b.32. Nominal Treatment Distance:
7.12.b.32.A. For electron irradiation, the
distance from the scattering foil, virtual source, or exit window of the
electron beam to the entrance surface of the irradiated object along the
central axis of the useful beam.
7.12.b.32.B. For x-ray irradiation, the
virtual source or target to isocenter distance along the central axis of the
useful beam. For non-isocentric equipment, this distance shall be that
specified by the manufacturer.
7.12.b.33. Patient - an individual subjected
to machine produced external beam radiation for the purposes of medical
therapy.
7.12.b.34. Peak Tube
Potential - the maximum value of the potential difference across the x-ray tube
during an exposure.
7.12.b.35.
Periodic Quality Assurance Check - a procedure which is performed to ensure
that a previous calibration continues to be valid.
7.12.b.36. Phantom - an object behaving in
essentially the same manner as tissue, with respect to absorption or scattering
of the ionizing radiation in question.
7.12.b.37. Practical Range of Electrons -
corresponds to classical electron range where the only remaining contribution
to dose is from Bremsstrahlung X-rays. A further explanation may be found in
"Clinical Electron Beam Dosimetry: Report of AAPM Radiation Therapy Committee
Task Group 25" [Medical Physics 18(1): 73-109, Jan/Feb. 1991] and ICRU Report
35, "Radiation Dosimetry: Electron Beams with Energies Between 1 and 50 MeV",
International Commission on Radiation Units and Measurements, September 15,
1984.
7.12.b.38. Primary Dose
Monitoring System - a system which will monitor the useful beam during
irradiation and which will terminate irradiation when a pre-selected number of
dose monitor units have been delivered.
7.12.b.39. Primary Protective Barrier - the
material, excluding filters, placed in the useful beam.
7.12.b.40. Protective Barrier - a barrier of
radiation absorbing materials used to reduce radiation exposure.
7.12.b.41. Radiation Detector - a device
which, in the presence of radiation provides, by either direct or indirect
means, a signal or other indication suitable for use in measuring one or more
quantities of incident radiation.
7.12.b.42. Radiation Field - the material
which attenuates stray radiation.
7.12.b.43. Radiation Head - the structure
from which the useful beam emerges.
7.12.b.44. Radiation Therapy Physicist - an
individual qualified in accordance with Paragraph 7.12.c.4.
7.12.b.45. Redundant Beam Monitoring System -
a combination of two dose monitoring systems in which each system is designed
to terminate irradiation in accordance with a pre-selected number of dose
monitor units.
7.12.b.46. Scattered
Radiation - ionizing radiation emitted by interaction of ionizing radiation
with matter, the interaction being accompanied by a change in direction of the
radiation. Scattered primary radiation means that scattered radiation which has
been deviated in direction only by materials irradiated by the useful
beam.
7.12.b.47. Secondary Dose
Monitoring System - a system which will terminate irradiation in the event of
failure of the primary dose monitoring system.
7.12.b.48. Secondary Protective Barrier - the
material which attenuates stray radiation.
7.12.b.49. Shadow Tray - a device attached to
the radiation head to support auxiliary beam blocking material.
7.12.b.50. Shutter - a device attached to the
tube housing assembly which can totally intercept the useful beam and which has
a lead equivalency not less than that of the tube housing assembly.
7.12.b.51. Sievert (Sv) - the SI unit of dose
equivalent. The unit of dose equivalent is the joule per kilogram. The previous
unit of dose equivalent (Rem) is being replaced by the Sievert. (One [1] Sv=one
hundred [100] Rem).
7.12.b.52.
Simulator (Radiation Therapy Simulation System) - any x-ray system intended for
localizing the volume to be exposed during radiation therapy and reproducing
the position and size of the therapeutic irradiation field.
7.12.b.53. Source - the region or material
from which the radiation emanates.
7.12.b.54. Source-skin Distance (SSD) - the
distance measured along the beam axis from the center of the front surface of
the x-ray target or electron virtual source to the surface of the irradiated
object or patient.
7.12.b.55.
Stationary beam radiation therapy - radiation therapy without displacement of
one or more mechanical axes relative to the patient during
irradiation.
7.12.b.56. Stray
Radiation - the sum of leakage and scattered radiation.
7.12.b.57. Target - that part of an x-ray
tube or accelerator onto which a beam of accelerated particles is directed to
produce ionizing radiation or other particles.
7.12.b.58. Target-skin Distance (TSD) - the
distance measured along the beam axis from the center of the front surface of
the x-ray target or electron virtual source to the surface of the irradiated
object or patient.
7.12.b.59.
Tenth-value Layer (TVL) - s the thickness of a specified material which
attenuates x-radiation or gamma radiation to an extent such that the air kerma
rate, exposure rate, or absorbed dose rate is reduced to one-tenth (0.1) of the
value measured without the material at the same point.
7.12.b.60. Termination of Irradiation - the
stopping of irradiation in a fashion which will not permit continuance of
irradiation without the resetting of operating conditions at the control
panel.
7.12.b.61. Therapeutic
Radiation Machine - x-ray or electron-producing equipment designed and used for
external beam radiation therapy.
7.12.b.62. Tube - an x-ray tube, unless
otherwise specified.
7.12.b.63.
Tube Housing Assembly - the tube housing with tube installed. It includes
high-voltage or filament transformers and other appropriate elements when such
are contained within the tube housing.
7.12.b.64. Useful Beam - the radiation
emanating from the tube housing port or the radiation head and passing through
the aperture of the beam limiting device when the exposure controls are in a
mode to cause the therapeutic radiation machine to produce radiation.
7.12.b.65. Virtual Source - a point from
which radiation appears to originate.
7.12.b.66. Wedge Filter - a filter which
effects continuous change in transmission over all or a part of the useful
beam.
7.12.b.67. X-ray Tube - any
electron tube which is designed to be used primarily for the production of
x-rays.
7.12.c. General
Administrative Requirements for Facilities Using Therapeutic Radiation
Machines.
7.12.c.1. The registrant shall be
responsible for directing the operation of the therapeutic radiation machines
that have been registered with the agency. The registrant or the registrant's
agent shall ensure that the requirements of Subsection 7.12. are met in the
operation of the therapeutic radiation machines.
7.12.c.2. A therapeutic radiation machine
that does not meet the provisions of this rule shall not be used for
irradiation of patients.
7.12.c.3.
The registrant for any therapeutic radiation machine subject to Subdivisions
7.12.f. or 7.12.g. shall require the radiation therapy physicist to:
7.12.c.3.A. Be registered with the agency,
under the provisions of Section 2. of this rule, as a provider of radiation
services in the area of calibration and compliance surveys of external beam
radiation therapy units; and
7.12.c.3.B. Be certified by the American
Board of Radiology in:
7.12.c.3.B.1.
Therapeutic Radiological Physics; or
7.12.c.3.B.2. Roentgen-ray and Gamma-ray
Physics; or
7.12.c.3.B.3. X-ray and
Radium Physics; or
7.12.c.3.B.4.
Radiological Physics; or
7.12.c.3.C. Be certified by the American
Board of Medical Physics in Radiation Oncology Physics; or
7.12.c.3.D. Hold a master's or doctor's
degree in physics, biophysics, radiological physics, or health physics, and
have completed 1 year of full time training in therapeutic radiological physics
and also 1 year of full time work experience under the supervision of a
radiation therapy physicist at a medical institution. To meet this requirement,
the individual shall have performed the tasks listed in 7.12.d.1., 7.12.f.16.
or 7.12.g.20., and 7.12.f.17. or 7.12.g.21. under the supervision of a
radiation therapy physicist during the year of work experience.
7.12.c.3.E. Notwithstanding the provisions of
7.12.c.3.C., certification pursuant 7.12.c.3.B. and 7.12.c.3.C. shall be
required on or before December 31, 1999, for all persons currently qualifying
as a radiation therapy physicist pursuant to 7.12.c.3.D.
7.12.c.4. Written safety procedures and rules
shall be developed by a radiation therapy physicist and shall be available in
the control area of a therapeutic radiation machine, including any restrictions
required for the safe operation of the particular therapeutic radiation
machine. The operator shall be able to demonstrate familiarity with this
rule.
7.12.c.5. Individuals shall
not be exposed to the useful beam except for medical therapy purposes and
unless such exposure has been ordered in writing by a licensed practitioner of
the healing arts who is specifically identified on the registration. This
provision specifically prohibits deliberate exposure of an individual for
training, demonstration or other non-healing-arts purposes.
7.12.c.6. Visiting Authorized User.
7.12.c.6.A. Notwithstanding the provisions of
7.12.c.5., a registrant may permit any physician to act as a visiting
authorized user under the term of the registrant's registration for up to sixty
(60) days per calendar year under the following conditions:
7.12.c.6.A.1. The visiting authorized user
has the prior written permission of the registrant's management and, if the use
occurs on behalf of an institution, the institution's radiation safety
committee; and
7.12.c.6.A.2. The
visiting authorized user meets the requirements established for authorized
users in 7.12.c.3.B.; and
7.12.c.6.a.3. The registrant maintains copies
of all records specified by 7.12.c.6. for five (5) years from the date of the
last visit.
7.12.c.6.B.
All individuals associated with the operation of a therapeutic radiation
machine shall be instructed in and shall comply with the provisions of the
registrant's quality management program. In addition to the requirements of
Subsection 7.12, these individuals are also subject to the requirements of
Subsection 6.5., Subsection 6.9. and Subsection 6.17. of this rule.
7.12.c.7. Information and
Maintenance Record and Associated Information. The registrant shall maintain
the following information in a separate file or package for each therapeutic
radiation machine, for inspection by the agency:
7.12.c.7.A. Report of acceptance
testing;
7.12.c.7.B. Records of all
surveys, calibrations, and periodic quality assurance checks of the therapeutic
radiation machine required by Subsection 7.12., as well as the names of persons
who performed such activities;
7.12.c.7.C. Records of maintenance and
modifications performed on the therapeutic radiation machine after July 1,
2001, as well as the names of persons who performed such services;
7.12.c.7.D. Signature of person authorizing
the return of therapeutic radiation machine to clinical use after service,
repair, or upgrade.
7.12.c.8. Records Retention. All records
required by Subsection 7.12. shall be retained until disposal is authorized by
the agency unless another retention period is specifically authorized in
Subsection 7.12. All required records shall be retained in an active file from
at least the time of generation until the next Agency inspection. Any required
record generated prior to the last agency inspection may be microfilmed or
otherwise archived as long as a complete copy of said record can be retrieved
until such time as the agency authorizes final disposal.
7.12.d. General Technical Requirements for
Facilities Using Therapeutic Radiation Machines.
7.12.d.1. Protection Surveys.
7.12.d.1.A. The registrant shall ensure that
radiation protection surveys of all new facilities, and existing facilities not
previously surveyed are performed with an operable radiation measurement survey
instrument calibrated in accordance with 7.12.h. The radiation protection
survey shall be performed by, or under the direction of, a radiation therapy
physicist or a qualified expert and shall verify that, with the therapeutic
radiation machine in a "BEAM-ON" condition, with the largest clinically
available treatment field and with a scattering phantom in the useful beam of
radiation:
7.12.d.1.A.1. Radiation levels in
restricted areas are not likely to cause personnel exposures in excess of the
limits specified in Subdivision 6.5.a. of this rule.; and
7.12.d.1.A.2. Radiation levels in
unrestricted areas do not exceed the limits specified in Subdivisions 6.13.a.
and 6.13.b. of this rule.
7.12.d.1.B. In addition to the requirements
of Subparagraph 7.12.d.1.A., a radiation protection survey shall also be
performed prior to any subsequent medical use and:
7.12.d.1.B.1. After making any change in the
treatment room shielding;
7.12.d.1.B.2. After making any change in the
location of the therapeutic radiation machine within the treatment
room;
7.12.d.1.B.3. After
relocating the therapeutic radiation machine; or
7.12.d.1.B.4. Before using the therapeutic
radiation machine in a manner that could result in increased radiation levels
in areas outside the external beam radiation therapy treatment room.
7.12.d.1.C. The survey record
shall indicate all instances where the facility, in the opinion of the
radiation therapy physicist or a qualified expert, is in violation of
applicable rules. The survey record shall also include: the date of the
measurements; the reason the survey is required; the manufacturer's name; model
number and serial number of the therapeutic radiation machine; the instruments
used to measure radiation levels; a plan of the areas surrounding the treatment
room that were surveyed; the measured dose rate at several points in each area
expressed in microsieverts or millirems per hour; the calculated maximum level
of radiation over a period of one (1) week for each restricted and unrestricted
area; and the signature of the individual responsible for conducting the
survey;
7.12.d.1.D. If the results
of the surveys required by 7.12.d.1.A. or 7.12.d.1.B. indicate any radiation
levels in excess of the respective limit specified in Subparagraph 7.12.d.1.A.,
the registrant shall lock the control in the "OFF" position and not use the
unit:
7.12.d.1.D.1. Except as may be
necessary to repair, replace, or test the therapeutic radiation machine, the
therapeutic radiation machine shielding, or the treatment room shielding;
or
7.12.d.1.D.2. Until the
registrant has received a specific exemption from the agency.
7.12.d.2. Modification
of radiation therapy unit or room before beginning a treatment program. If the
survey required by Paragraph 7.12.d.1. indicates that an individual in an
unrestricted area may be exposed to levels of radiation greater than those
permitted by Subdivisions 6.13.a. and 6.13.b. of this rule, before beginning
the treatment program the registrant shall:
7.12.d.2.A. Either equip the unit with beam
direction interlocks or add additional radiation shielding to ensure compliance
with Subdivisions 6.13.a. and 6.13.b. of this rule;
7.12.d.2.B. Perform the survey required by
Paragraph 7.12.d.1. again; and
7.12.d.2.C. Include in the report required by
Paragraph 7.12.d.4. the results of the initial survey, a description of the
modification made to comply with Subparagraph 7.12.d.2.A., and the results of
the second survey; or
7.12.d.2.D.
Request and receive a registration amendment under Subdivision 6.13.c. of this
rule that authorizes radiation levels in unrestricted areas greater than those
permitted by Subdivision 6.13.a and 6.13.b. of this rule.
7.12.d.3. Dosimetry Equipment.
7.12.d.3.A. The registrant shall have a
calibrated dosimetry system available for use. The system shall have been
calibrated by the National Institute for Standards and Technology (NIST) or by
an American Association of Physicists in Medicine (AAPM) accredited dosimetry
calibration laboratory (ADCL). The calibration shall have been performed within
the previous twenty four (24) months and after any servicing that may have
affected system calibration. An independent survey shall be conducted by a
qualified expert or radiation therapy physicist other that the person
performing the original survey prior to the equipment being used except as
described in Subparagraph 7.12.d.1.D.
7.12.d.3.A.1. For beams with energies greater
than one (1) MV (one [1] MeV), the dosimetry system shall have been calibrated
for Cobalt-60;
7.12.d.3.A.2. For
beams with energies equal to or less than one (1) MV (one [1] MeV), the
dosimetry system shall have been calibrated at an energy (energy range)
appropriate for the radiation being measured;
7.12.d.3.B. The registrant shall have
available for use a dosimetry system for quality assurance check measurements.
To meet this requirement, the system may be compared with a system that has
been calibrated in accordance with Subparagraph 7.12.d.3.A. This comparison
shall have been performed within the previous twelve (12) months and after each
servicing that may have affected system calibration. The quality assurance
check system may be the same system used to meet the requirement in
7.12.d.3.1.;
7.12.d.3.C. The
registrant shall maintain a record of each dosimetry system calibration,
intercomparison, and comparison for the duration of the registration. For each
calibration, intercomparison, or comparison, the record shall include: the
date; the model numbers and serial numbers of the instruments that were
calibrated, inter-compared, or compared as required by 7.12.d.3.A. and
7.12.d.3.B.; the correction factors that were determined; the names of the
individuals who performed the calibration, intercomparison, or comparison; and
evidence that the intercomparison was performed by, or under the direct
supervision and in the physical presence of, a radiation therapy
physicist.
7.12.d.4.
Reports of external beam radiation therapy surveys and measurements. The
registrant for any therapeutic radiation machine subject to Subdivisions
7.12.f. or 7.12.g. shall furnish a copy of the records required in Paragraphs
7.12.d.1. and 7.12.d.2. to the agency within thirty (30) days following
completion of the action that initiated the record requirement.
7.12.e. Quality Management
Program. The facility shall implement a quality management program. The
facility may use the quality management programs found in either Table 64-23 V
or Table 64-23 W.
7.12.f.
Therapeutic Radiation Machines of Less Than Five Hundred (500) kV.
7.12.f.1. Leakage Radiation. When the x-ray
tube is operated at its maximum rated tube current for the maximum kV, the
leakage air kerma rate shall not exceed the value specified at the distance
specified for that classification of therapeutic radiation machine:
7.12.f.1.A. Five to Fifty (5 to 50) kV
Systems. The leakage air kerma rate measured at any position 5 centimeters from
the tube housing assembly shall not exceed one (1) mGy (one hundred [100] mRad)
in any one hour.
7.12.f.1.B.
Greater than fifty (50) and less than five hundred (500) kV Systems. The
leakage air kerma rate measured at a distance of one (1) meter from the target
in any direction shall not exceed one (1) cGy (one [1] Rad) in any one (1)
hour. This air kerma rate measurement may be averaged over areas no larger than
one hundred (100) square centimeters. In addition, the air kerma rate at a
distance of five (5) centimeters from the surface of the tube housing assembly
shall not exceed thirty (30) cGy (thirty [30] rad) per hour.
7.12.f.1.C. For each therapeutic radiation
machine, the registrant shall determine, or obtain from the manufacturer, the
leakage radiation existing at the positions specified in Subparagraph
7.12.f.1.A. and 7.12.f.1.B. for the specified operating conditions. Records on
leakage radiation measurements shall be maintained at the installation for
inspection by the agency.
7.12.f.2. Permanent beam limiting devices.
Permanent diaphragms or cones used for limiting the useful beam shall provide
at least the same degree of attenuation as required for the tube housing
assembly.
7.12.f.3. Adjustable or
Removable Beam Limiting Devices.
7.12.f.3.A.
All adjustable or removable beam limiting devices, diaphragms, cones or blocks
shall not transmit more than five (5) percent of the useful beam for the most
penetrating beam used;
7.12.f.3.B.
When adjustable beam limiting devices are used, the position and shape of the
radiation field shall be indicated by a light beam.
7.12.f.4. Filter System. The filter system
shall be so designed that:
7.12.f.4.A.
Filters can not be accidentally displaced at any possible tube
orientation;
7.12.f.4.B. For
equipment installed after July 1, 2001, an interlock system prevents
irradiation if the proper filter is not in place;
7.12.f.4.C. The air kerma rate escaping from
the filter slot shall not exceed one (1) cGy (1 Rad) per hour at 1 meter under
any operating conditions; and
7.12.f.4.D. Each filter shall be marked as to
its material of construction and its thickness.
7.12.f.5. Tube Immobilization.
7.12.f.5.A. The x-ray tube shall be so
mounted that it can not accidentally turn or slide with respect to the housing
aperture; and
7.12.f.5.B. The tube
housing assembly shall be capable of being immobilized for stationary portal
treatments.
7.12.f.6.
Source Marking. The tube housing assembly shall be so marked that it is
possible to determine the location of the source to within five (5)
millimeters, and such marking shall be readily accessible for use during
calibration procedures.
7.12.f.7.
Beam Block. Contact therapy tube housing assemblies shall have a removable
shield of material, equivalent in attenuation to five-tenths (0.5) millimeters
of lead at one hundred (100) kV, which can be positioned over the entire useful
beam exit port during periods when the beam is not in use.
7.12.f.8. Timer. A suitable irradiation
control device shall be provided to terminate the irradiation after a pre-set
time interval.
7.12.f.8.A. A timer with a
display shall be provided at the treatment control panel. The timer shall have
a pre-set time selector and an elapsed time or time remaining
indicator;
7.12.f.8.B. The timer
shall be a cumulative timer that activates with an indication of "BEAM-ON" and
retains its reading after irradiation is interrupted or terminated. After
irradiation is terminated and before irradiation can be reinitiated, it shall
be necessary to reset the elapsed time indicator;
7.12.f.8.C. The timer shall terminate
irradiation when a pre-selected time has elapsed, if any dose monitoring system
present has not previously terminated irradiation;
7.12.f.8.D. The timer shall permit accurate
pre-setting and determination of exposure times as short as one (1)
second;
7.12.f.8.E. The timer shall
not permit an exposure if set at zero;
7.12.f.8.F. The timer shall not activate
until the shutter is opened when irradiation is controlled by a shutter
mechanism unless calibration includes a timer error correction to compensate
for mechanical lag; and
7.12.f.8.G.
Timer shall be accurate to within one (1) percent of the selected value or one
(1) second, whichever is greater.
7.12.f.9. Control Panel Functions. The
control panel, in addition to the displays required by other provisions in
Subdivision 7.12.f., shall have:
7.12.f.9.A.
An indication of whether electrical power is available at the control panel and
if activation of the x-ray tube is possible;
7.12.f.9.B. An indication of whether x-rays
are being produced;
7.12.f.9.C. A
means for indicating x-ray tube potential and current;
7.12.f.9.D. The means for terminating an
exposure at any time;
7.12.f.9.E. A
locking device which will prevent unauthorized use of the therapeutic radiation
machine; and
7.12.f.9.F. For
therapeutic radiation machines manufactured after July 1, 2001, a positive
display of specific filter or filters in the beam.
7.12.f.10. Multiple Tubes. When a control
panel may energize more than one (1) x-ray tube:
7.12.f.10.A. It shall be possible to activate
only one x-ray tube at any time;
7.12.f.10.B. There shall be an indication at
the control panel identifying which x-ray tube is activated; and
7.12.f.10.C. There shall be an indication at
the tube housing assembly when that tube is energized.
7.12.f.11. Target-to-Skin Distance (TSD).
There shall be a means of determining the central axis TSD to within one (1)
centimeter and of reproducing this measurement to within two (2) millimeters
thereafter.
7.12.f.12. Shutters.
Unless it is possible to bring the x-ray output to the prescribed exposure
parameters within five (5) seconds after the x-ray "ON" switch is energized,
the beam shall be attenuated by a shutter having a lead equivalency not less
than that of the tube housing assembly. In addition, after the unit is at
operating parameters, the shutter shall be controlled by the operator from the
control panel. An indication of shutter position shall appear at the control
panel.
7.12.f.13. Low Filtration
X-ray Tubes. Each therapeutic radiation machine equipped with a beryllium or
other low-filtration window shall be clearly labeled as such upon the tube
housing assembly and shall be provided with a permanent warning device on the
control panel that is activated when no additional filtration is present, to
indicate that the dose rate is very high.
7.12.f.14. Facility design requirements for
therapeutic radiation machines capable of operating in the range fifty (50) kV
to five hundred (500) kV. In addition to shielding adequate to meet
requirements of Subdivision 7.12.i., the treatment room shall meet the
following design requirements:
7.12.f.14.A.
Aural Communication. Provision shall be made for continuous two-way aural
communication between the patient and the operator at the control
panel;
7.12.f.14.B. Viewing
Systems. Provision shall be made to permit continuous observation of the
patient during irradiation and the viewing system shall be so located that the
operator can observe the patient from the control panel. The therapeutic
radiation machine shall not be used for patient irradiation unless at least one
viewing system is operational.
7.12.f.15. Additional Requirements. Treatment
rooms that contain a therapeutic radiation machine capable of operating above
one hundred fifty (150) kV shall meet the following additional requirements:
7.12.f.15.A. All protective barriers shall be
fixed except for entrance doors or beam interceptors;
7.12.f.15.B. The control panel shall be
located outside the treatment room or in a totally enclosed booth, which has a
ceiling, inside the room;
7.12.f.15.C. Interlocks shall be provided
such that all entrance doors, including doors to any interior booths, shall be
closed before treatment can be initiated or continued. If the radiation beam is
interrupted by any door opening, it shall not be possible to restore the
machine to operation without closing the door and reinitiating irradiation by
manual action at the control panel; and
7.12.f.15.D. When any door referred to in
Subparagraph 7.12.f.15.C. is opened while the x-ray tube is activated, the air
kerma rate at a distance of one (1) meter from the source shall be reduced to
less than one (1) mGy (one hundred [100] mRad) per hour.
7.12.f.16. Full Calibration Measurements.
7.12.f.16.A. Full calibration of a
therapeutic radiation machine subject to Subdivision 7.12.f. shall be performed
by, or under the direct supervision of, a radiation therapy physicist:
7.12.f.16.A.1. Before the first medical use
following installation or reinstallation of the therapeutic radiation
machine;
7.12.f.16.A.2. At
intervals not exceeding one (1) year; and
7.12.f.16.A.3. Before medical use under the
following conditions:
7.12.f.16.A.3.(a). When- ever quality
assurance check measurements indicate that the radiation output differs by more
than five (5) percent from the value obtained at the last full calibration and
the difference cannot be reconciled; and
7.12.f.16.A.3.(b). Following any component
replacement, major repair, or modification of components that could
significantly affect the characteristics of the radiation beam.
7.12.f.16.A.4. Notwithstanding the
requirements of Part 7.12.f.16.A.3.:
7.12.f.16.A.4.(a). Full calibration of
therapeutic radiation machines with multi-energy capabilities is required only
for those modes or energies that are not within their acceptable range;
and
7.12.f.16.A.4.(b). If the
repair, replacement or modification does not affect all energies, full
calibration shall be performed on the affected energy that is in most frequent
clinical use at the facility. The remaining energies may be validated with
quality assurance check procedures against the criteria in subpart
7.12.f.16.A.3.(a).
7.12.f.16.B. To satisfy the requirement of
Subparagraph 7.12.f.16.A., full calibration shall include all measurements
recommended for annual calibration by NCRP Report 69, "Dosimetry of X-ray and
Gamma Ray Beams for Radiation Therapy in the Energy Range 10 keV to 50 MeV"
(1981).
7.12.f.16.C. The registrant
shall maintain a record of each calibration for the duration of the
registration. The record shall include: the date of the calibration; the
manufacturer's name, model number, and serial number for both the therapeutic
radiation machine and the x-ray tube; the model numbers and serial numbers of
the instruments used to calibrate the therapeutic radiation machine; and the
signature of the radiation therapy physicist responsible for performing the
calibration.
7.12.f.17.
Periodic Quality Assurance Checks.
7.12.f.17.A. Periodic quality assurance
checks shall be performed on therapeutic radiation machines subject to
Subdivision 7.12.f., which are capable of operation at greater than or equal to
fifty (50) kV.
7.12.f.17.B. To
satisfy the requirement of Subparagraph 7.12.f.17.A., quality assurance checks
shall meet the following requirements:
7.12.f.17.B.1. The registrant shall perform
quality assurance checks in accordance with written procedures established by
the radiation therapy physicist.; and
7.12.f.17.B.2. The quality assurance check
procedures shall specify the frequency at which tests or measurements are to be
performed. The quality assurance check procedures shall specify that the
quality assurance check shall be performed during the calibration specified in
Subparagraph 7.12.f.16.A. The acceptable tolerance for each parameter measured
in the quality assurance check, when compared to the value for that parameter
determined in the calibration specified in Subparagraph 7.12.f.16.A., shall be
stated.
7.12.f.17.C. The
cause for a parameter exceeding a tolerance set by the radiation therapy
physicist shall be investigated and corrected before the system is used for
patient irradiation;
7.12.f.17.D.
Whenever a quality assurance check indicates a significant change in the
operating characteristics of a system, as specified in the radiation therapy
physicist's quality assurance check procedures, the system shall be
recalibrated as required in Subparagraph 7.12.f.16.A.;
7.12.f.17.E. The registrant shall use the
dosimetry system described in Subparagraph 7.12.d.3.B. to make the quality
assurance check required in Subparagraph 7.12.f.17.A.;
7.12.f.17.F. The registrant shall have the
radiation therapy physicist review and sign the results of each radiation
output quality assurance check within one (1) month of the date that the check
was performed;
7.12.f.17.G. The
registrant shall ensure that safety quality assurance checks of therapeutic
radiation machines subject to Subdivision 7.12.f. are performed at intervals
not to exceed one (1) month;
7.12.f.17.H. Notwithstanding the requirements
of Subparagraphs 7.12.f.17.F. and Subparagraph 7.12.f.17.G., the registrant
shall ensure that no therapeutic radiation machine is used to administer
radiation to humans unless the quality assurance checks required by
Subparagraphs 7.12.f.17.F. and Subparagraph 7.12.f.17.G. have been performed
within the thirty (30) day period immediately prior to said
administration;
7.12.f.17.I. To
satisfy the requirement of Subparagraph 7.12.f.17.G., safety quality assurance
checks shall ensure proper operation of:
7.12.f.17.I.1. Electrical interlocks at each
external beam radiation therapy room entrance;
7.12.f.17.I.2. The "BEAM-ON" and termination
switches;
7.12.f.17.I.3. Beam
condition indicator lights on the access doors, control console, and in the
radiation therapy room;
7.12.f.17.I.4. Viewing systems;
7.12.f.17.I.5. If applicable, electrically
operated treatment room doors from inside and outside the treatment
room.
7.12.f.17.J. The
registrant shall maintain a record of each quality assurance check required by
Subparagraph 7.12.f.17.A. and Subparagraph 7.12.f.17.g. for three (3) years.
The record shall include: the date of the quality assurance check; the
manufacturer's name, model number, and serial number of the therapeutic
radiation machine; the manufacturer's name; model number and serial number for
the instruments used to measure the radiation output of the therapeutic
radiation machine; and the signature of the individual who performed the
periodic quality assurance check.
7.12.f.18. Operating Procedures.
7.12.f.18.A. The therapeutic radiation
machine shall not be used for irradiation of patients unless the requirements
of Subparagraph 7.12.f.16. And Subparagraph 7.12.f.17. have been met;
7.12.f.18.B. Therapeutic radiation machines
shall not be left unattended unless secured pursuant to Subparagraph
7.12.f.1.E.;
7.12.f.18.C. When a
patient must be held in position for radiation therapy, mechanical supporting
or restraining devices shall be used;
7.12.f.18.D. The tube housing assembly shall
not be held by an individual during operation unless the assembly is designed
to require such holding and the peak tube potential of the system does not
exceed fifty (50) kV. In such cases, the holder shall wear protective gloves
and apron of not less than five-tenths (0.5) millimeters lead equivalency at
one hundred (100) kV;
7.12.f.18.E.
A copy of the current operating and emergency procedures shall be maintained at
the therapeutic radiation machine control console; and
7.12.f.18.F. No individual other than the
patient shall be in the treatment room during exposures from therapeutic
radiation machines operating above one hundred fifty (150) kV. At energies less
than or equal to one hundred fifty (150) kV, any individual, other than the
patient, in the treatment room shall be protected by a barrier sufficient to
meet the requirements of Subsection 6.5 of this rule.
7.12.f.19. Possession of Survey Instruments.
Each facility location authorized to use a therapeutic radiation machine in
accordance with Subdivision 7.12.f. shall possess appropriately calibrated
portable monitoring equipment. As a minimum, such equipment shall include a
portable radiation measurement survey instrument capable of measuring dose
rates over the range ten (10) :Sv (one [1] mRem) per hour to ten (10) mSv (one
thousand [1000] mRem) per hour. The survey instruments shall be operable and
calibrated in accordance with Subdivision 7.12.h.
7.12.g. Therapeutic Radiation Machines -
photon therapy systems (five hundred [500] kV and above) and electron therapy
systems (five hundred [500] keV and above).
7.12.g.1. Possession of Survey Instruments.
Each facility location authorized to use a therapeutic radiation machine in
accordance with Subdivision 7.12.g. shall possess appropriately calibrated
portable monitoring equipment. As a minimum, such equipment shall include a
portable radiation measurement survey instrument capable of measuring dose
rates over the range ten (10) :Sv (one [1] mRem) per hour to ten (10) mSv (one
thousand [1000] mRem) per hour. The survey instruments shall be operable and
calibrated in accordance with Subsection 7.12.h.
7.12.g.2. Leakage Radiation Outside the
Maximum Useful Beam in Photon and Electron Modes.
7.12.g.2.A. The absorbed dose due to leakage
radiation (excluding neutrons) at any point outside the maximum sized useful
beam, but within a circular plane of radius 2 meters which is perpendicular to
and centered on the central axis of the useful beam at the nominal treatment
distance (i.e. patient plane), shall not exceed a maximum of two-tenths (0.2)
percent and an average of one-tenth (0.1) percent of the absorbed dose on the
central axis of the beam at the nominal treatment distance. Measurements shall
be averaged over an area not exceeding one hundred (100) square centimeters at
a minimum of sixteen (16) points uniformly distributed in the plane;
7.12.g.2.B. Except for the area defined in
Subparagraph 7.12.g.2.A., the absorbed dose due to leakage radiation (excluding
neutrons) at one (1) meter from the electron path between the electron source
and the target or electron window shall not exceed five-tenths (0.5) percent of
the absorbed dose on the central axis of the beam at the nominal treatment
distance. Measurements shall be averaged over an area not exceeding one hundred
(100) square centimeters;
7.12.g.2.C. For equipment manufactured after
July 1, 2001, the neutron absorbed dose outside the useful beam shall be in
compliance with International Electrotechnical Commission (IEC) Document
601-2-1; and
7.12.g.2.D. For each
therapeutic radiation machine, the registrant shall determine, or obtain from
the manufacturer, the leakage radiation existing at the positions specified in
Subparagraphs 7.12.g.2.A. through 7.12.g.2.C. for the specified operating
conditions. Records on leakage radiation measurements shall be maintained at
the installation for inspection by the agency.
7.12.g.3. Leakage Radiation Through Beam
Limiting Devices.
7.12.g.3.A. Photon
Radiation. All adjustable or interchangeable beam limiting devices shall
attenuate the useful beam such that at the nominal treatment distance, the
maximum absorbed dose anywhere in the area shielded by the beam limiting
devices shall not exceed two (2) percent of the maximum absorbed dose on the
central axis of the useful beam measured in a ten (10) centimeter by ten (10)
centimeter radiation field;
7.12.g.3.B. Electron Radiation. All
adjustable or interchangeable electron applicators shall attenuate the
radiation, including but not limited to photon radiation generated by electrons
incident on the beam limiting device and electron applicator and other parts of
the radiation head, such that the absorbed dose in a plane perpendicular to the
central axis of the useful beam at the nominal treatment distance shall not
exceed:
7.12.g.3.B.1. A maximum of two (2)
percent and average of five-tenths (0.5) percent of the absorbed dose on the
central axis of the useful beam at the nominal treatment distance. This limit
shall apply beyond a line seven (7) centimeters outside the periphery of the
useful beam; and
7.12.g.3.B.2. A
maximum of ten (10) percent of the absorbed dose on the central axis of the
useful beam at the nominal treatment distance. This limit shall apply beyond a
line two (2) centimeters outside the periphery of the useful beam..
7.12.g.3.C. Measurement of Leakage
Radiation.
7.12.g.3.C.1. Photon Radiation.
Measurements of leakage radiation through the beam limiting devices shall be
made with the beam limiting devices closed and any residual aperture blocked by
at least two (2) tenth value layers of suitable absorbing material. In the case
of overlapping beam limiting devices, the leakage radiation through each set
shall be measured independently at the depth of maximum dose. Measurements
shall be made using a radiation detector of area not exceeding ten (10) square
centimeters;
7.12.g.3.C.2. Electron
Radiation. Measurements of leakage radiation through the electron applicators
shall be made with the electron beam directed into the air and using a
radiation detector of area up to but not exceeding one (1) square centimeter
suitably protected against radiation which has been scattered from material
beyond the radiation detector. Measurements shall be made using one (1)
centimeter of water equivalent build up material.
7.12.g.4. Filters or Wedges.
7.12.g.4.A. Each wedge filter that is
removable from the system shall be clearly marked with an identification
number. For removable wedge filters, the nominal wedge angle shall appear on
the wedge or wedge tray (if permanently mounted to the tray). If the wedge or
wedge tray is significantly damaged, the wedge transmission factor shall be
redetermined;
7.12.g.4.B. If the
absorbed dose rate information required by Paragraph 7.12.g.1. relates
exclusively to operation with a field flattening filter or beam scattering foil
in place, such foil or filter shall be removable only by the use of
tools;
7.12.g.4.C. For equipment
manufactured after July 1, 2001 which utilizes wedge filters, interchangeable
field flattening filters, or interchangeable beam scattering foils:
7.12.g.4.C.1. Irradiation shall not be
possible until a selection of a filter or a positive selection to use "no
filter" has been made at the treatment control panel, either manually or
automatically;
7.12.g.4.C.2. An
interlock system shall be provided to prevent irradiation if the filter
selected is not in the correct position;
7.12.g.4.C.3. A display shall be provided at
the treatment control panel showing the wedge filters, interchangeable field
flattening filters, or interchangeable beam scattering foils in use;
and
7.12.g.4.C.4. An interlock
shall be provided to prevent irradiation if any filter or beam scattering foil
selection operation carried out in the treatment room does not agree with the
filter or beam scattering foil selection operation carried out at the treatment
control panel.
7.12.g.5. Stray radiation in the useful beam.
For equipment manufactured after July 1, 2001, the registrant shall determine
during acceptance testing, or obtain from the manufacturer, data sufficient to
ensure that x-ray stray radiation in the useful electron beam, absorbed dose at
the surface during x-ray irradiation and stray neutron radiation in the useful
x-ray beam are in compliance with International Electrotechnical Commission
(IEC) Document 601-2-1 (most current revision).
7.12.g.6. Beam Monitors. All therapeutic
radiation machines subject to Subdivision 7.12.g. shall be provided with
redundant beam monitoring systems. The sensors for these systems shall be fixed
in the useful beam during treatment to indicate the dose monitor unit rate.
7.12.g.6.A. Equipment manufactured after July
1, 2001 shall be provided with at least 2 independently powered integrating
dose meters. Alternatively, common elements may be used if the production of
radiation is terminated upon failure of any common element.
7.12.g.6.B. Equipment manufactured on or
before July 1, 2001 shall be provided with at least one (1) radiation detector.
This detector shall be incorporated into a useful beam monitoring
system;
7.12.g.6.C. The detector
and the system into which that detector is incorporated shall meet the
following requirements:
7.12.g.6.C.1. Each
detector shall be removable only with tools and, if movable, shall be
interlocked to prevent incorrect positioning;
7.12.g.6.C.2. Each detector shall form part
of a beam monitoring system from whose readings in dose monitor units the
absorbed dose at a reference point can be calculated;
7.12.g.6.C.3. Each beam monitoring system
shall be capable of independently monitoring, interrupting, and terminating
irradiation; and
7.12.g.6.C.4. For
equipment manufactured after July 1, 2001, the design of the beam monitoring
systems shall ensure that the:
7.12.g.6.C.4.(a). Malfunctioning of one
system shall not affect the correct functioning of the other systems;
and
7.12.g.6.C.4.(b). Failure of
either system shall terminate irradiation or prevent the initiation of
radiation.
7.12.g.6.C.5. Each beam
monitoring system shall have a legible display at the treatment control panel.
For equipment manufactured after July 1, 2001, each display shall:
7.12.g.6.C.5.(a). Maintain a reading until
intentionally reset;
7.12.g.6.C.5.(b). Have only one scale and no
electrical or mechanical scale multiplying factors;
7.12.g.6.C.5.(c). Utilize a design such that
increasing dose is displayed by increasing numbers; and
7.12.g.6.C.5.(d). In the event of power
failure, the beam monitoring information required in Part 7.12.g.5.E.3.
displayed at the control panel at the time of failure shall be retrievable in
at least one system for a twenty (20) minute period of time.
7.12.g.7. Beam
Symmetry.
7.12.g.7.A. Bent-beam linear
accelerators subject to 7.12.g. shall be provided with auxiliary devices to
monitor beam symmetry;
7.12.g.7.B.
The devices referenced in Subparagraph 7.12.g.7.A. shall be able to detect
field asymmetry greater than ten (10) percent; and
7.12.g.7.C. The devices referenced in
Subparagraph 7.12.g.7.A. shall be configured to terminate irradiation if the
specifications in Subparagraph 7.12.g.7.B. can not be maintained.
7.12.g.8. Selection and Display of
Dose Monitor Units.
7.12.g.8.A. Irradiation
shall not be possible until a new selection of a number of dose monitor units
has been made at the treatment control panel;
7.12.g.8.B. The pre-selected number of dose
monitor units shall be displayed at the treatment control panel until reset
manually for the next irradiation;
7.12.g.8.C. After termination of irradiation,
it shall be necessary to reset the dosimeter display before subsequent
treatment can be initiated; and
7.12.g.8.D. For equipment manufactured after
July 1, 2001, after termination of irradiation, it shall be necessary for the
operator to reset the pre-selected dose monitor units before irradiation can be
initiated.
7.12.g.9. Air
Kerma Rate or Absorbed Dose Rate. For equipment manufactured after July 1,
2001, a system shall be provided from whose readings the air kerma rate or
absorbed dose rate at a reference point can be calculated. In addition:.
7.12.g.9.A. The dose monitor unit rate shall
be displayed at the treatment control panel;
7.12.g.9.B. If the equipment can deliver
under any conditions an air kerma rate or absorbed dose rate at the nominal
treatment distance more than twice the maximum value specified by the
manufacturer, a device shall be provided which terminates irradiation when the
air kerma rate or absorbed dose rate exceeds a value twice the specified
maximum. The dose rate at which the irradiation will be terminated shall be a
record maintained by the registrant;
7.12.g.9.C. If the equipment can deliver
under any fault condition or conditions an air kerma rate or absorbed dose rate
at the nominal treatment distance more than ten (10) times the maximum value
specified by the manufacturer, a device shall be provided to prevent the air
kerma rate or absorbed dose rate anywhere in the radiation field from exceeding
twice the specified maximum value and to terminate irradiation if the excess
absorbed dose at the nominal treatment distance exceeds four (4) Gy (four
hundred [400] Rad); and
7.12.g.9.D.
For each therapeutic radiation machine, the registrant shall determine, or
obtain from the manufacturer, the maximum values specified in 7.12.g.9.B. and
7.12.g.9.C. for the specified operating conditions. Records of these maximum
values shall be maintained at the installation for inspection by the
agency.
7.12.g.10.
Termination of irradiation by the beam monitoring system or systems during
stationary beam radiation therapy.
7.12.g.10.A. Each primary system shall
terminate irradiation when the pre-selected number of dose monitor units has
been detected by the system;
7.12.g.10.B. If the original design of the
equipment included a secondary dose monitoring system, that system shall be
capable of terminating irradiation when not more than fifteen (15) percent or
forty (40) dose monitor units above the pre-selected number of dose monitor
units set at the control panel has been detected by the secondary dose
monitoring system; and
7.12.g.10.C.
For equipment manufactured after July 1, 2001, an indicator on the control
panel shall show which monitoring system has terminated irradiation.
7.12.g.11. Termination of
Irradiation. It shall be possible to terminate irradiation and equipment
movement or go from an interruption condition to termination condition at any
time from the operator's position at the treatment control panel.
7.12.g.12. Interruption of Irradiation. If a
therapeutic radiation machine has an interrupt mode, it shall be possible to
interrupt irradiation and equipment movements at any time from the treatment
control panel. Following an interruption it shall be possible to restart
irradiation by operator action without any reselection of operating conditions.
If any change is made of a pre-selected value during an interruption,
irradiation and equipment movements shall be automatically
terminated.
7.12.g.13. Timer. A
suitable irradiation control device shall be provided to terminate the
irradiation after a pre-set time interval.
7.12.g.13.A. A timer shall be provided which
has a display at the treatment control panel. The timer shall have a pre-set
time selector and an elapsed time indicator;
7.12.g.13.B. The timer shall be a cumulative
timer that activates with an indication of "BEAM-ON" and retains its reading
after irradiation is interrupted or terminated. After irradiation is terminated
and before irradiation can be reinitiated, it shall be necessary to reset the
elapsed time indicator;
7.12.g.13.C. The timer shall terminate
irradiation when a pre-selected time has elapsed, if the dose monitoring
systems have not previously terminated irradiation.
7.12.g.14. Selection of Radiation Type.
Equipment capable of both x-ray therapy and electron therapy shall meet the
following additional requirements:
7.12.g.14.A. Irradiation shall not be
possible until a selection of radiation type (x-rays or electrons) has been
made at the treatment control panel;
7.12.g.14.B. The radiation type selected
shall be displayed at the treatment control panel before and during
irradiation;
7.12.g.14.C. An
interlock system shall be provided to ensure that the equipment can principally
emit only the radiation type that has been selected;
7.12.g.14.D. An interlock system shall be
provided to prevent irradiation with x-rays, except to obtain an image, when
electron applicators are fitted;
7.12.g.14.E. An interlock system shall be
provided to prevent irradiation with electrons when accessories specific for
x-ray therapy are fitted; and
7.12.g.14.F. An interlock system shall be
provided to prevent irradiation if any selected operations carried out in the
treatment room do not agree with the selected operations carried out at the
treatment control panel.
7.12.g.15. Selection of Energy. Equipment
capable of generating radiation beams of different energies shall meet the
following requirements:
7.12.g.15.A.
Irradiation shall not be possible until a selection of energy has been made at
the treatment control panel;
7.12.g.15.B. The nominal energy value
selected shall be displayed at the treatment control panel until reset manually
for the next irradiation. After termination of irradiation, it shall be
necessary to reset the nominal energy value selected before subsequent
treatment can be initiated;
7.12.g.15.C. Irradiation shall not be
possible until the appropriate flattening filter or scattering foil for the
selected energy is in its proper location; and
7.12.g.15.D. For equipment manufactured after
July 1, 2001, the selection of energy shall be in compliance with International
Electrotechnical Commission (IEC) Document 601-2-1.
7.12.g.16. Selection of Stationary Beam
Radiation Therapy or Moving Beam Radiation Therapy. Therapeutic radiation
machines capable of both stationary beam radiation therapy and moving beam
radiation therapy shall meet the following requirements:
7.12.g.16.A. Irradiation shall not be
possible until a selection of stationary beam radiation therapy or moving beam
radiation therapy has been made at the treatment control panel;
7.12.g.16.B. The mode of operation shall be
displayed at the treatment control panel;
7.12.g.16.C. An interlock system shall be
provided to ensure that the equipment can operate only in the mode that has
been selected;
7.12.g.16.D. An
interlock system shall be provided to prevent irradiation if any selected
parameter in the treatment room does not agree with the selected parameter at
the treatment control panel;
7.12.g.16.E. Moving beam radiation therapy
shall be controlled to obtain the selected relationships between incremental
dose monitor units and incremental movement. For equipment manufactured after
July 1, 2001:
7.12.g.16.E.1. An interlock
system shall be provided to terminate irradiation if the number of dose monitor
units delivered in any ten (10) degrees of rotation or one (1) cm of linear
motion differs by more than twenty (20) percent from the selected
value;
7.12.g.16.E.2. Where angle
terminates the irradiation in moving beam radiation therapy, the dose monitor
units delivered shall differ by less than five (5) percent from the dose
monitor unit value selected;
7.12.g.16.E.3. An interlock shall be provided
to prevent motion of more than five (5) degrees or one (1) cm beyond the
selected limits during moving beam radiation therapy;
7.12.g.16.E.4. An interlock shall be provided
to require that a selection of direction be made at the treatment control panel
in all units which are capable of both clockwise and counter-clockwise moving
beam radiation therapy.
7.12.g.16.E.5. Moving beam radiation therapy
shall be controlled with both primary position sensors and secondary position
sensors to obtain the selected relationships between incremental dose monitor
units and incremental movement.
7.12.g.16.E.6. Where the beam monitor system
terminates the irradiation in moving beam radiation therapy, the termination of
irradiation shall be as required by Paragraph 7.12.g.10.; and
7.12.g.16.E.7. For equipment manufactured
after July 1, 2001, an interlock system shall be provided to terminate
irradiation if movement:
7.12.g.16.E.7.(a). Occurs during stationary
beam radiation therapy; or
7.12.g.16.E.7.(b). Does not start or stops
during moving beam radiation therapy unless such stoppage is a pre-planned
function.
7.12.g.17. Facility Design Requirements for
Therapeutic Radiation Machines Operating Above 500 kV. In addition to shielding
adequate to meet requirements of 1.12.i., the following design requirements are
made:
7.12.g.17.A. Protective Barriers. All
protective barriers shall be fixed, except for access doors to the treatment
room or movable beam interceptors;
7.12.g.17.B. Control Panel. In addition to
other requirements specified in Subsection 7.12., the control panel shall also:
7.12.g.17.B.1. Be located outside the
treatment room;
7.12.g.17.B.2.
Provide an indication of whether electrical power is available at the control
panel and if activation of the radiation is possible;
7.12.g.17.B.3. Provide an indication of
whether radiation is being produced; and
7.12.g.17.B.4. Include an access control
(locking) device that will prevent unauthorized use of the therapeutic
radiation machine.
7.12.g.17.C. Viewing Systems. Windows,
mirrors, closed-circuit television or an equivalent viewing system shall be
provided to permit continuous observation of the patient following positioning
and during irradiation and shall be so located that the operator may observe
the patient from the treatment control panel. The therapeutic radiation machine
shall not be used for patient irradiation unless at least one viewing system is
operational;
7.12.g.17.D. Aural
Communications. Provision shall be made for continuous two-way aural
communication between the patient and the operator at the control panel. The
therapeutic radiation machine shall not be used for irradiation of patients
unless continuous two-way aural communication is possible;
7.12.g.17.E. Room Entrances. Treatment room
entrances shall be provided with warning lights in a readily observable
position near the outside of all access doors, which will indicate when the
useful beam is "ON" and when it is "OFF";
7.12.g.17.F. Entrance Interlocks. Interlocks
shall be provided such that all access controls are activated before treatment
can be initiated or continued. If the radiation beam is interrupted by any
access control, it shall not be possible to restore the machine to operation
without resetting the access control and reinitiating irradiation by manual
action at the control panel;
7.12.g.17.G. Beam Interceptor Interlocks. If
the shielding material in any protective barrier requires the presence of a
beam interceptor to ensure compliance with Subdivision 6.13.a. and Subsection
6.13.b. of this rule, interlocks shall be provided to prevent the production of
radiation, unless the beam interceptor is in place, whenever the useful beam is
directed at the designated barriers;
7.12.g.17.H. Emergency Cutoff Switches. At
least 1 emergency power cutoff switch shall be located in the radiation therapy
room and shall terminate all equipment electrical power including radiation and
mechanical motion. This switch is in addition to the termination switch
required by 7.12.g.11. All emergency power cutoff switches shall include a
manual reset so that the therapeutic radiation machine cannot be restarted from
the unit's control console without resetting the emergency cutoff
switch;
7.12.g.17.I. Safety
Interlocks. All safety interlocks shall be designed so that any defect or
component failure in the safety interlock system prevents or terminates
operation of the therapeutic radiation machine; and
7.12.g.17.J. Surveys for Residual Radiation.
Surveys for residual activity shall be conducted on all therapeutic radiation
machines capable of generating photon and electron energies above ten (10) MV
prior to machining, removing, or working on therapeutic radiation machine
components which may have become activated due to photo-neutron
production.
7.12.g.18.
Radiation Therapy Physicist Support.
7.12.g.18.A. The services of a radiation
therapy physicist shall be required in facilities having therapeutic radiation
machines with energies of five hundred (500) kV and above. The radiation
therapy physicist shall be responsible for:
7.12.g.18.A.1. Full calibrations required by
Paragraph 7.12.g.20. and protection surveys required by Paragraph
7.12.d.1.;
7.12.g.18.A.2.
Supervision and review of dosimetry;
7.12.g.18.A.3. Beam data acquisition and
transfer for computerized dosimetry, and supervision of its use;
7.12.g.18.A.4. Quality assurance, including
quality assurance check review required by Subparagraph 7.12.g.21.E.
7.12.g.18.A.5. Consultation with the
authorized user in treatment planning, as needed; and
7.12.g.18.A.6. Perform
calculations/assessments regarding misadministrations.
7.12.g.18.B. If the radiation therapy
physicist is not a full-time employee of the registrant, the operating
procedures required by Subparagraph 7.12.g.24. shall also specifically address
how the radiation therapy physicist is to be contacted for problems or
emergencies, as well as the specific actions, if any, to be taken until the
radiation therapy physicist can be contacted.
7.12.g.19. Operating Procedures.
7.12.g.19.A. No individual, other than the
patient, shall be in the treatment room during treatment or during any
irradiation for testing or calibration purposes;
7.12.g.19.B. Therapeutic radiation machines
shall not be made available for medical use unless the requirements of
7.12.a.1., 7.12.g.20. and 7.12.21. have been met;
7.12.g.19.C. Therapeutic radiation machines,
when not in operation, shall be secured to prevent unauthorized use;
7.12.g.19.D. When adjustable beam limiting
devices are used, the position and shape of the radiation field shall be
indicated by a light field.
7.12.g.19.E. If a patient must be held in
position during treatment, mechanical supporting or restraining devices shall
be used; and
7.12.g.19.F. A copy of
the current operating and emergency procedures shall be maintained at the
therapeutic radiation machine control console.
7.12.g.20. Acceptance testing, commissioning
and full calibration measurements.
7.12.g.20.A. Acceptance testing,
commissioning and full calibration of a therapeutic radiation machine subject
to Subsection 7.12. shall be performed by, or under the direct supervision of,
a radiation therapy physicist.
7.12.g.20.B. Acceptance testing and
commissioning shall be performed in accordance with "AAPM Code of Practice for
Radiotherapy Accelerators: Report of AAPM Radiation Therapy Task Group 45" and
shall be conducted before the first medical use following installation or
reinstallation of the therapeutic radiation machine.
7.12.g.20.C. Full calibration shall include
measurement of all parameters required by Table II of "Comprehensive QA for
Radiation Oncology: Report of AAPM Radiation Therapy Committee Task Group 40"
and shall be performed in accordance with "AAPM Code of Practice for
Radiotherapy Accelerators: Report of AAPM Radiation Therapy Task Group 45".
Although it shall not be necessary to complete all elements of a full
calibration at the same time, all parameters (for all energies) shall be
completed at intervals not exceeding twelve (12) calendar months, unless a more
frequent interval is required in Table ii.
7.12.g.20.D. The radiation therapy physicist
shall perform all elements of a full calibration necessary to determine that
all parameters are within acceptable limits:
7.12.g.20.D.1. Whenever quality assurance
check measurements indicate that the radiation output differs by more than five
(5) percent from the value obtained at the last full calibration and the
difference cannot be reconciled. Therapeutic radiation machines with
multi-energy or multi-mode capabilities shall only require measurements for
those modes or energies that are not within their acceptable range;
and
7.12.g.20.D.2. Following any
component replacement, major repair, or modification of components that could
significantly affect the characteristics of the radiation beam. If the repair,
replacement or modification does not affect all modes or energies, measurements
shall be performed on the effected mode or energy that is in most frequent
clinical use at the facility. The remaining energies or modes may be validated
with quality assurance check procedures against the criteria in
7.12.g.20.D.1.
7.12.g.20.E. The registrant shall maintain a
record of each calibration in an auditable form for the duration of the
registration. The record shall include: the date of the calibration; the
manufacturer's name, model number and serial number for the therapeutic
radiation machine; the model numbers and serial numbers of the instruments used
to calibrate the therapeutic radiation machine; and the signature of the
radiation therapy physicist responsible for performing the
calibration.
7.12.g.21.
Periodic Quality Assurance Checks.
7.12.g.21.A. Periodic quality assurance
checks shall be performed on all therapeutic radiation machines subject to
Subdivision 7.12. at intervals not to exceed those specified in "comprehensive
qa for radiation oncology: report of AAPM radiation therapy committee task
group 40";
7.12.g.21.B. To satisfy
the requirement of Subparagraph 7.12.g.21.A., quality assurance checks shall
include determination of central axis radiation output and a representative
sampling of periodic quality assurance checks contained in "comprehensive qa
for radiation oncology: report of AAPM radiation therapy committee task group
40". Representative sampling shall include all referenced periodic quality
assurance checks in an interval not to exceed 12 consecutive calendar
months;
7.12.g.21.C. The registrant
shall use a dosimetry system that has been inter-compared within the previous
12 months with the dosimetry system described in Subparagraph 7.12.d.3.A. to
make the periodic quality assurance checks required in Subparagraphs
7.12.g.21.B.;
7.12.g.21.D. The
registrant shall perform periodic quality assurance checks required by
Subparagraph 7.12.g.21.A. in accordance with procedures established by the
radiation therapy physicist;
7.12.g.21.E. The registrant shall review the
results of each periodic radiation output check according to the following
procedures:
7.12.g.21.E.1. The authorized
user and radiation therapy physicist shall be immediately notified if any
parameter is not within its acceptable tolerance. The therapeutic radiation
machine shall not be made available for subsequent medical use until the
radiation therapy physicist has determined that all parameters are within their
acceptable tolerances;
7.12.g.21.E.2. If all quality assurance check
parameters appear to be within their acceptable range, the quality assurance
check shall be reviewed and signed by either the authorized user or radiation
therapy physicist within three (3) treatment days; and
7.12.g.21.E.3. The radiation therapy
physicist shall review and sign the results of each radiation output quality
assurance check at intervals not to exceed one (1) month.
7.12.g.21.F. Therapeutic radiation machines
subject to Subsection 7.12. shall have safety quality assurance checks listed
in "Comprehensive QA for Radiation Oncology: Report of AAPM Radiation Therapy
Committee Task Group 40" performed at intervals not to exceed one (1)
week;
7.12.g.21.G. To satisfy the
requirement of Subparagraph 7.12.g.21.F., safety quality assurance checks shall
ensure proper operation of:
7.12.g.21.G.1.
Electrical interlocks at each external beam radiation therapy room
entrance;
7.12.g.21.G.2. Proper
operation of the "BEAM-ON", interrupt and termination switches;
7.12.g.21.G.3. Beam condition indicator
lights on the access doors, control console, and in the radiation therapy
room;
7.12.g.21.G.4. Viewing
systems;
7.12.g.21.G.5.
Electrically operated treatment room doors from inside and outside the
treatment room;
7.12.g.21.G.6. At
least one emergency power cutoff switch. If more than one emergency power
cutoff switch is installed and not all switches are tested at once, each switch
shall be tested on a rotating basis. Safety quality assurance checks of the
emergency power cutoff switches may be conducted at the end of the treatment
day in order to minimize possible stability problems with the therapeutic
radiation machine.
7.12.g.21.H. The registrant shall promptly
repair any system identified in Subparagraph 7.12.g.21.G. that is not operating
properly; and
7.12.g.21.I. The
registrant shall maintain a record of each quality assurance check required by
Subparagraphs 7.12.g.21.A. and 7.21.g.21.G. for three (3) years. The record
shall include: the date of the quality assurance check; the manufacturer's
name, model number, and serial number of the therapeutic radiation machine; the
manufacturer's name, model number and serial number for the instruments used to
measure the radiation output of the therapeutic radiation machine; and the
signature of the individual who performed the periodic quality assurance
check.
7.12.h.
Calibration of Survey Instruments.
7.12.h.1.
The registrant shall ensure that the survey instruments used to show compliance
with Subsection 7.12. have been calibrated before first use, at intervals not
to exceed twelve (12) months, and following repair.
7.12.h.2. To satisfy the requirements of
Paragraph 7.12.h.1., the registrant shall:
7.12.h.2.A. Calibrate all required scale
readings up to ten (10) mSv (one thousand [1000] mRem) per hour with an
appropriate radiation source that is traceable to the National Institute of
Standards and Technology (NIST);
7.12.h.2.B. Calibrate at least two (2) points
on each scale to be calibrated. These points should be at approximately
one-third (1/3) and two-thirds (2/3) of full-scale; and
7.12.h.3. To satisfy the requirements of
Paragraph 7.12.h.2., the registrant shall:
7.12.h.3.B. Consider a point as calibrated if
the indicated dose rate differs from the calculated dose rate by not more than
ten (10) percent; and
7.12.h.3.C.
Consider a point as calibrated if the indicated dose rate differs from the
calculated dose rate by not more than twenty (20) percent if a correction
factor or graph is conspicuously attached to the instrument.
7.12.h..3.D. The registrant shall retain a
record of each calibration required in Paragraph 7.12.h.1. for three (3) years.
The record shall include:
7.12.h.3.D.1. A
description of the calibration procedure; and
7.12.h.3.D.2. A description of the source
used and the certified dose rates from the source, and the rates indicated by
the instrument being calibrated, the correction factors deduced from the
calibration data, the signature of the individual who performed the
calibration, and the date of calibration.
7.12.h.3.E. The registrant may obtain the
services of individuals registered by the agency, the US Nuclear Regulatory
Commission, an agreement state, or a licensing state to perform calibrations of
survey instruments. Records of calibrations that contain information required
by Paragraph 7.12.h.4. shall be maintained by the
registrant.
7.12.i. Shielding and Safety Design
Requirements.
7.12.i.1. Each therapeutic
radiation machine subject to Subdivision 7.12.g. or 7.12.h. shall be provided
with such primary or secondary barriers as are necessary to ensure compliance
with Subsections 6.5. and 6.13. of this rule.
7.12.i.2. Facility design information for all
new installations of a therapeutic radiation machine or installations of a
therapeutic radiation machine of higher energy into a room not previously
approved for that energy shall be submitted for agency approval prior to actual
installation of the therapeutic radiation machine. The minimum facility design
information that must be submitted is contained in Table 64-23 U.