40 CFR Appendix Table B-4 to Subpart B of Part 53 - Line Voltage and Room Temperature Test Conditions
Table B-4 to Subpart B of Part 53—Line Voltage and Room Temperature Test Conditions
| Test day | Line
voltage, 1 rms |
Room
temperature, 2 °C |
Comments |
|---|---|---|---|
| 0 | 115 | 25 | Initial set-up and adjustments. |
| 1 | 125 | 20 | |
| 2 | 105 | 20 | |
| 3 | 125 | 30 | Adjustments and/or periodic maintenance permitted at end of tests. |
| 4 | 105 | 30 | |
| 5 | 125 | 20 | |
| 6 | 105 | 20 | Adjustments and/or periodic maintenance permitted at end of tests. |
| 7 | 125 | 30 | Examine test results to ascertain if further testing is required. |
| 8 | 105 | 30 | |
| 9 | 125 | 20 | Adjustments and/or periodic maintenance permitted at end of tests. |
| 10 | 105 | 20 | |
| 11 | 125 | 30 | |
| 12 | 105 | 30 | Adjustments and/or periodic maintenance permitted at end of tests. |
| 13 | 125 | 20 | |
| 14 | 105 | 20 | |
| 15 | 125 | 30 |
1 Voltage specified shall be controlled to ±1 volt.
2 Temperatures shall be controlled to ±1 °C.