7 AAC 18.550 - Survey instruments
(a) For physical surveys required under this
chapter, a registrant shall maintain calibrated and operable ionizing radiation
survey instruments that have a range from two milliroentgens per hour to one
roentgen per hour. An industrial radiographic operation may not be conducted
unless calibrated and operable ionizing radiation survey instrumentation is
available and used at each site where radiographic exposures are
made.
(b) A registrant shall ensure
that each survey instrument is calibrated after any service or repair and at
regular intervals not to exceed six months. A record shall be maintained of the
latest date of calibration.
(c)
Records of surveys and calibrations must be retained for at least five years,
or since the most recent inspection by the department, whichever is
less.