Ariz. Admin. Code § R13-10-108 - Examination and Quality Assurance Requirements for Analysts
A. The Department
shall require an Analyst permit applicant to successfully demonstrate the
applicant's proficiency in making alcohol concentration determinations from
test specimens in accordance with subsection (B). The applicant shall be
examined only on the methods that relate to the type of determination for which
the applicant desires a permit.
B.
An applicant shall, before receiving an initial Analyst permit or renewal of an
existing Analyst permit, participate in and successfully complete proficiency
testing administered by the Department. An applicant shall successfully analyze
samples by testing at least three suitable reference standards or control
samples with a known alcohol concentration in the range of 0.00 to 0.40 grams
per 100 milliliters of blood and having the results agree with the established
value within the limits of ± 0.01 grams per 100 milliliters of blood or
±10 percent, whichever is greater. Proficiency testing shall be
administered by the Department as follows:
1.
An applicant shall correctly analyze all proficiency samples in the set
provided by the Department.
2. When
returning the results of analyses to the Department, the applicant shall attach
an affidavit attesting that the applicant analyzed the proficiency samples
without help or input from any other person.
3. An applicant failing to correctly analyze
all proficiency samples in the set will be provided an opportunity to
successfully analyze a second set of samples.
4. The Department shall deny the application
of an applicant who declines or fails to correctly analyze the second set of
proficiency samples and shall not issue a permit.
5. An applicant who fails to successfully
analyze the second set of proficiency samples and whose application is denied
may reapply for an analyst's permit beginning 90 days from the date of
denial.
C. An analyst
who conducts alcohol concentration determinations shall implement and maintain
a quality assurance program. This program shall be designed to ensure the
validity of test results by providing for:
1.
Chain of custody,
2. Quality
control,
3. Analytical
procedures,
4. Documentation of
test results, and
5. Participation
in proficiency testing.