Ariz. Admin. Code § R9-7-1114 - Inspection and Maintenance of Radiation Machines, Survey Instruments, and Associated Equipment
A. A
registrant shall perform visual and operability checks on survey instruments
and radiation machines before use on each day the equipment is to be used to
ensure that the equipment is in good working condition and required labeling is
present. Survey instrument operability checks shall be performed using check
sources or other authorized means. If equipment problems are found, the
registrant shall remove the equipment from service until it is
repaired.
B. A registrant shall
have written inspection and maintenance procedures for radiation machines and
survey instruments that require inspection and maintenance, at intervals that
do not exceed three months or before first use of the equipment and to ensure
the proper functioning of components important to safety. Replacement
components shall meet design specifications. If equipment problems are
discovered, the registrant shall remove the equipment from service until the
equipment is repaired.
C. A
registrant shall maintain records of equipment problems found in daily checks
and quarterly inspections and retain each record for three years after it is
made. The record shall include the date of the check or inspection, name of the
inspector, equipment involved, any problems found, and any repair or needed
maintenance performed.