Ariz. Admin. Code § R9-7-1130 - Personnel Monitoring
A. An
individual shall not act as a radiographer or a radiographer's assistant
unless, at all times during radiographic operations, the individual wears, on
the trunk of the body, a direct reading dosimeter, an operating alarm rate
meter, and either a film badge, a TLD, or an optically stimulated luminescence
(OSL) dosimeter. At permanent radiography installations where other required
alarm or warning devices are in routine use, an alarm rate meter is not
required.
1. A registrant shall provide
pocket dosimeters that have a range from zero to 2 millisieverts (200
millirems) and ensure that the dosimeters are recharged at the start of each
shift. Electronic personnel dosimeters are permitted in place of ion-chamber
pocket dosimeters.
2. The
registrant shall assign a film badge, TLD, or OSL dosimeter to one individual,
who shall wear the assigned equipment.
3. The registrant shall replace film badges
at least monthly and replace TLDs or OSL dosimeters at least
quarterly.
4. After replacement,
the registrant shall ensure that each film badge or TLD is processed as soon as
possible.
B. A
radiographer or radiographer's assistant shall record exposures noted from
direct reading dosimeters, such as pocket dosimeters or electronic personnel
dosimeters, at the beginning and end of each shift.
C. A registrant shall check each pocket
dosimeter or electronic personnel dosimeter at least yearly for correct
response to radiation, and discontinue use of a dosimeter if it is not accurate
within plus or minus 20% of the true radiation exposure.
D. If an individual's pocket dosimeter has an
off-scale reading, or the electronic personnel dosimeter reads greater than 2
millisieverts (200 millirems), and radiation exposure cannot be ruled out as
the cause, a registrant shall send the individual's film badge, TLD, or OSL
dosimeter for processing within 24 hours. The registrant shall not allow the
individual to work with a radiation machine until the individual's radiation
exposure is determined. Using the information from the badge or dosimeter, the
RSO or the RSO's designee shall calculate the affected individual's cumulative
radiation exposure, as prescribed in Article 4 of this Chapter and include the
results in records maintained in accordance with subsection (G).
E. If an individual's monitoring device is
lost or damaged, the individual shall cease work immediately until the
registrant provides a replacement film badge, TLD, or OSL dosimeter and the RSO
or the RSO's designee calculates the exposure for the time period from issuance
to discovery of a lost or damaged film badge, TLD, or OSL dosimeter. The
registrant shall include the calculated exposure and the time period for which
the film badge, TLD, or OSL dosimeter was lost or damaged in the records
maintained in accordance with subsection (G).
F. For each alarm rate meter a registrant
shall ensure that:
1. At the start of a shift
each individual with an alarm rate meter checks that the alarm functions
(sounds) before using the device;
2. Each device is set to give an alarm signal
at a preset dose rate of 5 mSv/hr (500 mrem/hr) with an accuracy of plus or
minus 20% of the true radiation dose rate;
3. A special means is necessary to change the
preset alarm function on the device; and
4. Each device is calibrated at periods that
do not to exceed 12 months for correct response to radiation
G. Each registrant shall maintain
the following personnel monitoring records:
1.
Each dosimeter reading and the yearly operability check required by subsections
(B) and (C) for three years after each record is made;
2. A record of each alarm rate meter
calibration for three years after the record is made;
3. Any report received from the film badge,
TLD, or OSL processor. The registrant shall maintain these records until the
Department terminates the registration; and
4. Any estimation of an exposure evidenced by
an off-scale personnel direct-reading dosimeter or a lost or damaged film
badge, TLD, or OSL dosimeter. The records shall be maintained until the
Department terminates the registration.
Notes
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.