Ariz. Admin. Code § R9-7-202 - Application for Registration of Ionizing Radiation Producing Machines
A. A person shall
not use a radiation machine except as authorized in this Article.
B. A person possessing a nonexempt radiation
machine shall apply for registration of the machine with the Department within
30 days after its installation. The person applying for registration of a
radiation-producing machine shall use the application forms provided by the
Department. The applicant shall provide the information identified in Appendix
A of this Article.
C. In addition
to the application form or forms, the applicant shall remit the appropriate
registration or licensing fee in R9-7-1306 and provide other information
required by R9-7-208.
D. Each
applicant that applies for registration of a stationary x-ray system, with the
exception of applicants from bone densitometry, cabinet radiography, podiatry,
dental, bone mineral analyzer and mammography facilities, shall provide a scale
drawing of the room in which the x-ray system is located, or provide
measurements from the radiation source to the surrounding barrier surfaces. The
drawing shall denote the type of materials and the thickness (or lead
equivalence) of each barrier of the room (walls, ceilings, floors, doors,
windows). The drawing shall also denote the type and frequency of occupancy in
adjacent areas, including those above and below the x-ray room of concern
(e.g., hallways, offices, parking lots, and lavatories). Estimates of workload
shall also be provided with the drawing.
E. An applicant proposing to use a particle
accelerator for medical purposes shall not use the particle accelerator until
the Department inspection required in R9-7-914 has been completed.