Ariz. Admin. Code § R9-7-450 - Sealed Sources
A. A
licensee shall only receive, possess, and use radioactive materials contained
in a sealed source that has been manufactured, labeled, packaged, and
distributed in accordance with a specific license for its manufacture and
distribution. The license to manufacture and distribute a sealed source shall
be issued by the Department, the U.S. Nuclear Regulatory Commission, a
Licensing State, or another Agreement State.
B. A licensee who possesses and uses a sealed
source, or any device or equipment that contains a sealed source, shall follow
the radiation safety and handling instructions approved by the Department or
follow the radiation safety and handling instructions furnished by the
manufacturer on the label attached to the source, on the permanent container of
the source, or in a leaflet or brochure that accompanies the source, and
maintain the instructions in a legible and conveniently available form. If the
handling instructions, leaflet, or brochure is no longer available and a copy
cannot be obtained from the manufacturer, the licensee shall notify the
Department that the source handling information is no longer
available.
C. Inventories:
1. An inventory shall be conducted at
intervals not to exceed six months, unless a shorter interval is specified by
license condition.
2. The records
of the inventory shall be maintained for three years from the date of the
inventory, and shall be available for inspection by the Department.
3. The information recorded shall include:
a. The kind and quantity of radioactive
material,
b. The model and serial
number of the source or the device in which it is mounted,
c. The location of the sealed
source,
d. The date of the
inventory, and
e. The signature of
the person performing the inventory.
D. Any licensee who possesses and uses sealed
sources in the practice of medicine shall conduct a physical inventory
according to the requirements in 9 A.A.C. 7, Article 7.
E. Sealed sources, containing radioactive
material, shall not be opened unless authorized by license condition.
F. Sealed sources and machines, devices, or
equipment containing sealed sources shall be used in accordance with procedures
described in the manufacturer's instructions and the safety precautions
described in the Nuclear Regulatory Commission Sealed Sources and Device
Registry, unless the instructions or precautions conflict with these rules or
license condition.
Notes
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.