Ariz. Admin. Code § R9-7-608 - Mobile Diagnostic Radiographic and Mobile Fluoroscopic Systems, Except Dental Panoramic, Cephalometric, Dental CT, or Dental Intraoral Radiographic Systems
A.
Equipment
1. All requirements of R9-7-607(A)
and (B) apply.
2. For mobile
radiographic units the registrant shall provide a "dead-man" switch, together
with an electrical cord of sufficient length so that the operator can stand out
of the useful beam and at least 1.82 meters (6 feet) from the patient during
all x-ray exposures.
3. A
registrant shall ensure that a cone, spacer frame, or inherent provision is
made so that the equipment is not operated at source-skin distances of less
than 20.3 centimeters (8 inches).
B. Structural shielding. If a mobile unit is
used routinely in one location, it is considered a fixed installation subject
to the shielding requirements in
R9-7-603(C), and
R9-7-607(C).
C. Operating
procedures
1. All provisions of R9-7-607(D)
apply.
2. An individual who
operates a mobile x-ray system shall comply with
R9-7-419(B).