Ariz. Admin. Code § R9-7-804 - Open-beam X-ray Systems
A. A
registrant shall label open beam x-ray systems with a readily discernible sign
or signs bearing the radiation symbol and the words:
1. "CAUTION -- HIGH INTENSITY X-RAY BEAM," or
a similar warning, on the x-ray source housing; and
2. "CAUTION RADIATION -- THIS EQUIPMENT
PRODUCES RADIATION WHEN ENERGIZED" or a similar warning, near any switch that
energizes an x-ray tube if the radiation source is an x-ray tube.
B. A registrant shall ensure that
an open beam x-ray system has all of the following warning devices:
1. X-ray tube status (On-Off) indicator in
systems where the primary beam is controlled in this fashion;
2. Shutter status (Open-Closed) indicators
near each port on the radiation housing for systems which control the primary
beam; and
3. A clearly visible
warning light labeled with the words "X-RAY ON," or a similar warning located
near any switch that energizes an x-ray tube, illuminated only when the tube is
energized; and
4. The warning
devices in subsections (B)(1) through (3) shall be labeled so that their
purpose is easily identified.
C. A registrant shall ensure that any
apparatus utilized in beam alignment procedures is designed in such a way that
excessive radiation will not strike the operator. Particular attention shall be
given to viewing devices, in order to ascertain that lenses and other
transparent components attenuate the beam to an acceptable level.
D. A registrant shall provide an interlock
device which prevents entry of any portion of an individual's body into the
primary beam or causes the primary beam to be shut off upon entry into its path
on all open-beam x-ray systems. A registrant may apply to the Department for an
exemption from the requirements of a safety device. An application for
exemption shall include:
1. A description of
the various safety devices that have been evaluated;
2. The reason each device cannot be used;
and
3. A description of the
alternative methods that will be used to minimize accidental exposure,
including procedures to assure that operators and others in the area will be
informed of the absence of safety devices.
E. A registrant shall use only systems
constructed so that:
1. Each x-ray tube
housing is equipped with an interlock that automatically shuts off the tube if
the tube is removed from the radiation source housing or the housing is
disassembled; and
2. With all
shutters closed, radiation measured at a distance of 5 centimeters from the
surface of the system is not capable of producing a dose that exceeds 25 Sv
(2.5 mRem) in one hour for the specified tube rating of the x-ray
tube.
F. A registrant
shall supply each x-ray generating system with a protective cabinet that limits
leakage radiation measured at a distance of 5 cm (2 in) from the cabinet
surface, so that the system is not capable of producing a dose equivalent that
exceeds 25 µSv (2.5 mrem) in one hour.
G. A registrant shall ensure that the local
components of an analytical x-ray system are located and arranged and have
sufficient shielding or access control for the specified tube rating to prevent
the radiation level in any area adjacent to the local component group from
exceeding the dose limits in
R9-7-416.
H. A registrant shall perform a radiation
survey of the local component group of each analytical x-ray system to
demonstrate compliance with subsection (G) upon:
1. Installation,
2. Change in configuration, or
3. Maintenance that affects the radiation
level in any area adjacent to the local component group.
I. A registrant shall maintain a record of
each survey for three years or until the analytical x-ray system is no longer
used, whichever period is shorter.
Notes
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.