Ariz. Admin. Code § R9-7-907 - Shielding and Safety Design
A. An
authorized medical physicist experienced in the principles of radiation
protection and installation design shall be consulted in the design of a
particle accelerator installation and called upon to perform a radiation survey
when the accelerator is first capable of producing radiation. The registrant
shall provide a copy of the installation radiation survey to the Department
before a Department inspection conducted according to R9-7-914.
B. The registrant shall shield each particle
accelerator installation with the primary and secondary protective barriers
necessary to comply with
R9-7-408 and
R9-7-416.
C. At the time of application for
registration and before treatment of the first patient, the applicant shall
provide to the Department a copy of an installation report, signed by the
contractor who installed required shielding material recommended by the
authorized medical physicist who performed the shielding calculations for the
particle accelerator facility.
D.
As part of the annual radiation protection program review required in
R9-7-407(C), the registrant shall document installed facility shielding and
other radiation exposure controls, review patient workload, and note associated
changes, if any, in public exposure that are the result of installed facility
shielding, increased workload, and other radiation exposure controls in use at
the time of the review.