Ariz. Admin. Code § R9-7-908 - Particle Accelerator Controls and Interlock Systems
A registrant shall ensure that:
1. Instrumentation, readouts and controls on
the particle accelerator control panel are clearly identified and easily
discernible;
2. All entrances into
the area that contains the particle accelerator room, target room, or other
high radiation area, are provided with interlocks that shut down the machine if
an entrance door is opened;
3. If
an interlock system connected to an entrance door that provides access to the
therapy suite has been tripped, it is not possible to resume operation of the
particle accelerator by resetting the interlock switch at the entrance where it
had been tripped;
4. Each safety
interlock is on a circuit that allows it to operate independently of all other
safety interlocks;
5. If possible,
the interlock system is fail-safe in design, so that any defect or component
failure in the interlock system prevents operation of the particle accelerator;
and
6. A scram button or other
emergency power cutoff switch is located and easily identifiable in the area
that contains the particle accelerator. The registrant shall ensure that the
scram button prevents persons from restarting the particle accelerator at the
accelerator control panel without resetting the button or switch.