Ariz. Admin. Code § R9-7-910 - Operating Procedures
A. A
registrant shall secure from use a particle accelerator when it is not being
used to prevent unauthorized use.
B. A particle accelerator operator shall use
the switch on the control panel to turn the accelerator beam on and off during
normal operations. The safety interlock system may be used to turn off the
accelerator beam in emergencies.
C.
A registrant shall ensure that all safety and warning systems, including
interlocks, are tested for proper operation at intervals not to exceed three
months, and maintain a record of each test for Department inspection for at
least three years from the date of the test.
D. A registrant shall keep current electrical
circuit diagrams of a particle accelerator and the associated interlock
systems, and maintain the diagrams for inspection by the Department.
E. A registrant shall not bypass an interlock
unless the by-pass is:
1. Authorized in
writing by the Radiation Safety Committee or Radiation Safety Office,
2. Recorded in a permanent log with a notice
of the by-pass posted at any affected interlock and at the control panel,
and
3. Terminated as soon as
possible.
F. A
registrant shall maintain a copy of the current operating and emergency
procedures at the particle accelerator control panel.