Fla. Admin. Code Ann. R. 59A-13.022 - Emergency Procedures
(1) All PPEC centers
must conform to state standards prepared by the State Fire Marshal: Chapter
69A-36, F.A.C., Uniform Fire Safety Standards for Child Care Facilities, and
must be inspected annually. A copy of the current annual fire inspection
report, conducted by the local authority having jurisdiction over fire safety
or the State Fire Marshal, must be on file at the PPEC center. Documentation of
a satisfactory fire safety inspection shall be provided at the time of the
licensee's annual survey.
(2) There
must be a working telephone, which is neither locked nor a pay station, in the
PPEC center.
(3) Emergency telephone
numbers must be posted on or in the immediate vicinity of all
telephones.
(4) An emergency
generator must exist, with sufficient generating power to continue function of
medical equipment in the event of a power failure. The emergency generator must
be tested every 30 days and satisfactory mechanical operation must be
documented on a log designed for that purpose and signed by the person
conducting the test.
(5) Emergency
transportation must be performed by a licensed E.M.S. provider, with a PPEC
center staff member accompanying each child.
(6) The PPEC center must have an emergency
kit available to provide basic first aid and cardiopulmonary
resuscitation.
Notes
Rulemaking Authority 400.914(1) FS. Law Implemented 400.914(1)(f) FS.
New 3-8-89, Formerly 10D-102.022, Amended 6-22-06.