Ill. Admin. Code tit. 32, § 333.40 - Calibration Services
a) The
Department provides the following calibration services under accreditation by
the Conference of Radiation Control Program Directors, Inc.
1) X-ray Calibrations. 5 beam codes at 32
mR/sec. This service is available for the Radcal MDH 1015 x-ray monitor only.
The available beam codes are:
|
BEAM CODE |
HALF-VALUE LAYER |
HOMOGENEITY COEFFICIENT |
|
|
L |
80 |
1.81 mm Aluminum |
0.59 |
|
L |
100 |
2.82 mm Aluminum |
0.59 |
|
M |
30 |
0.36 mm Aluminum |
0.65 |
|
M |
50 |
1.03 mm Aluminum |
0.64 |
|
M |
100 |
4.96 mm Aluminum |
0.73 |
2)
Gamma Ray Calibrations. 1 beam code. This service is available only for ion
chamber and energy compensated G-M tube instruments.
|
BEAM CODE |
INTENSITIES |
|
Cs-137 |
.005 - 40 R/hr |
b) The Department also provides the following
services, which are not performed under accreditation.
1) Microrem Meter Calibrations. 2 points per
scale, gamma intensity range available to 0.015 mR/hr.
2) Pocket Dosimeter Calibrations. 1 point
gamma radiation up to 200 R.
3)
Count-Rate Instrument Calibrations. Pulse rate at 2 points per scale. Alpha
probe response is checked at 2 points. Beta-gamma probes are checked at 1
point. Other checks may be performed, depending on each instrument and probe
combination.
c) The
Department may be able to provide additional calibration services, including
but not limited to calibrations using additional beam codes. Persons desiring
any calibration services are urged to contact the facility by telephone at
(217) 786-7221 to determine the availability of additional services.