Ill. Admin. Code tit. 32, § 320.70 - Additional Requirements for Operators of Class D Radiation Installations
a) Each operator of
a Class D radiation installation shall utilize the services of an individual,
registered with the Agency pursuant to 32 Ill. Adm. Code 410, to implement and
maintain a comprehensive radiation protection program. Activities related to
diagnostic radiation producing machines shall be performed by a registered
diagnostic imaging specialist (DIS). Activities related to therapeutic
radiation machines shall be performed by a registered therapeutic radiological
physicist (TRP). Each operator shall ensure that registered individuals:
1) Conduct an annual performance evaluation
of all radiation machines and verify their inclusion in the radiation
installation registration.
2)
Determine and document in a report to the radiation installation that the
radiation machines evaluated are being maintained and operated in accordance
with standards established by the Agency to protect the public health as set
forth in 32 Ill. Adm. Code: Chapter II, Subchapters b and d. Noncompliance
items shall be readily identified in the report.
3) Establish and oversee the
equipment-related quality assurance practices. Specifically, these quality
assurance practices shall include as a minimum:
A) For therapeutic radiation machines,
compliance with the quality assurance requirements specified in 32 Ill. Adm.
Code 360.110(d) or
360.120(e).
B) For computed tomography machines,
compliance with the quality assurance requirements specified in 32 Ill. Adm.
Code 360.75.
C) For mammography machines, compliance with
the quality assurance requirements specified in 32 Ill. Adm. Code
370.100.
4) Users of digital imaging acquisition
systems shall follow a quality assurance/quality control protocol for image
processing established by the manufacturer for each radiation machine. The
program shall include the entire imaging chain from detector to the monitor
used for interpretation and:
A) The operator
of the radiation installation shall include the protocol in its operating and
safety procedures. The procedures shall be in writing or electronically
available for Agency review.
B) The
operator of the radiation installation shall document the frequency at which
the quality assurance/quality control protocol is performed. Documentation
shall include the date and initials of the individual completing the document
and shall be maintained at the site where performed for inspection by the
Agency.
C) The protocol shall
include, but not be limited to, the following:
i) Cleaning and erasure of all imaging plates
used in computed radiography (CR);
ii) Quality control phantom analysis
including, but not limited to, low contrast detectability, spatial resolution,
and artifacts;
iii) Evaluation of
repeat/reject x-ray examinations;
iv) Review of exposure index, deviation
index, s-number, or dose index values from clinically acquired radiographic
images.
b) Each operator of a Class D radiation
installation shall maintain and have available for review by the Agency:
1) Accurate and thorough radiation machine
evaluation reports.
2) Records of
quality assurance testing performed.
3) Records of calibrations, maintenance or
repair.
4) Records of corrective
action taken for items of non-compliance.
5) Records of digital imaging quality
control.
c) The records
and reports required by this Section shall be maintained for a period of at
least one inspection cycle.