Ill. Admin. Code tit. 32, § 351.330 - Radiation Detection Instruments
a) The licensee or registrant shall keep
sufficient calibrated and operable radiation survey instruments at each field
station and temporary jobsite to make physical radiation surveys as required by
this Part and by 32 Ill. Adm. Code
340.510(a).
Instrumentation shall be capable of measuring 0.001 mSv (0.1 mrem) per hour
through at least 0.5 mSv (50 mrem) per hour.
b) The licensee shall have available
additional calibrated and operable radiation detection instruments sensitive
enough to detect the low radiation and contamination levels that could be
encountered if a sealed source ruptured. The licensee may own the instruments
or may have a procedure to obtain them quickly from a second party.
c) The licensee must have each radiation
survey instrument required under subsections (a) and (b) calibrated:
1) At intervals not to exceed 6 months and
after each instrument servicing (e.g., electronic repair);
2) For linear scale instruments, at two
points located approximately 1/3 and 2/3 of full-scale on each scale; for
logarithmic scale instruments, at midrange of each decade, and at two points of
at least one decade; and for digital instruments, at appropriate points;
and
3) So that accuracy within plus
or minus 20 percent of the calibration standard can be demonstrated on each
scale.
d) The licensee
shall retain calibration records for a period of 3 years after the date of
calibration for inspection by the Agency.