Ill. Admin. Code tit. 32, § 360.130 - Electronic Brachytheraphy
Current through Register Vol. 46, No. 15, April 8, 2022
a)
Applicability. Electronic brachytherapy devices shall be subject to the
requirements of this Section and shall be exempt from the requirements of
Section
360.110,
unless otherwise noted in this Section.
1) An
electronic brachytherapy device that does not meet the requirements of this
Section shall not be used for irradiation of patients; and
2) An electronic brachytherapy device shall
only be utilized for human use applications specifically approved by the U.S.
Food and Drug Administration (FDA) unless participating in a research study
approved by the registrant's Institutional Review Board.
b) Possession of Survey Instruments. Each
registrant using an electronic brachytherapy device in accordance with this
Section shall possess appropriately calibrated portable monitoring equipment.
At a minimum, the monitoring equipment shall include a portable radiation
measurement survey instrument capable of measuring dose rates over the range 10
µSv (1 mrem) per hour to 10 mSv (1000 mrem) per hour. The survey
instruments shall be operable and calibrated within the prior 12 months for the
applicable electronic brachytherapy source energy.
c) Facility Design Requirements for
Electronic Brachytherapy Devices. Each electronic brachytherapy installation
shall be provided with such primary and secondary barriers as are necessary to
assure compliance with 32 Ill. Adm. Code 340.
1) If applicable, provision shall be made to
prevent simultaneous operation of more than one therapeutic radiation machine
in a treatment room.
2) Access to
the treatment room shall be controlled by a door at each entrance.
3) Each treatment room shall have provisions
to permit continuous aural communication and visual observation of the patient
from the treatment control panel during irradiation. The electronic
brachytherapy device shall not be used for patient irradiation unless the
patient can be observed.
d) Control Panel Functions. The control
panel, in addition to the displays required by other provisions in this
Section, shall:
1) Provide an indication of
whether electrical power is available at the control panel and if activation of
the electronic brachytherapy source is possible;
2) Provide an indication of whether x-rays
are being produced;
3) Provide a
means for indicating electronic brachytherapy source potential and
current;
4) Provide a means for
terminating an exposure at any time; and
5) Include an access control (locking) device
that will prevent unauthorized use of the electronic brachytherapy
device.
e) Timer. A
suitable irradiation control device (timer) shall be provided to terminate the
irradiation after a pre-set time interval or integrated charge on a
dosimeter-based monitor. The timer shall operate according to the
manufacturer's design specifications.
f) Therapeutic Radiological Physicist
Support. The services of a therapeutic radiological physicist shall be required
in facilities having electronic brachytherapy devices. The therapeutic
radiological physicist shall be responsible for:
1) Evaluation of the output from the
electronic brachytherapy source;
2)
Generation of the necessary dosimetric information;
3) Supervision and review of treatment
calculations prior to initial treatment of any treatment site;
4) Establishing the periodic and day-of-use
quality assurance checks and reviewing the data from those checks as required
in subsection (j);
5) Consultation
with the physician in treatment planning, as needed;
6) Performing calculations/assessments
regarding patient treatments that may constitute a misadministration.
7) Determination of the need for shielding or
safe distances for individuals in the room during electronic brachytherapy
treatments, in accordance with the radiation dose limits of 32 Ill. Adm. Code
Part 340;
8) Implementation of the
use of shielding or safe distances as determined in subsection
(f)(7).
g) Operating
Procedures
1) Only individuals approved by
the physician or therapeutic radiological physicist shall be present in the
treatment room during treatment.
2)
Electronic brachytherapy devices shall not be made available for medical use
unless the requirements of this Section have been met.
3) The electronic brachytherapy device shall
be inoperable, either by hardware or password, when unattended by qualified
staff or service personnel.
4)
During operation, the therapeutic radiologic physicist shall ensure that all
persons in the treatment room, and all persons entering the treatment room, are
prevented from exceeding the radiation dose limits of 32 Ill. Adm. Code Part
340.
5) If a patient must be held
in position during treatment, mechanical supporting or restraining devices
shall be used.
6) Written
procedures shall be developed, implemented, and maintained for responding to an
abnormal situation. These procedures shall include:
A) Instructions for responding to equipment
failures and the names of the individuals responsible for implementing
corrective actions; and
B) The
names and telephone numbers of the physician and the therapeutic radiological
physicist to be contacted if the device or console operates
abnormally.
7) A copy of
the current operating and emergency procedures shall be physically located at
the electronic brachytherapy device control console.
8) Instructions shall be posted at the
electronic brachytherapy device control console to inform the electronic
brachytherapy device operator of the names and telephone numbers of the
physician and the therapeutic radiological physicist to be contacted if the
device or console operates abnormally.
h) Safety Precautions for Electronic
Brachytherapy Devices
1) A therapeutic
radiological physicist shall determine which persons in the treatment room
require radiation monitoring when the beam is energized.
2) A physician and a therapeutic radiological
physicist shall be physically present during the initiation of all patient
treatments involving the electronic brachytherapy device.
3) A therapeutic radiological physicist and
either a physician or an electronic brachytherapy device operator, under the
supervision of a physician, who has been trained in the operation of, and
emergency response for, the electronic brachytherapy device, shall be
physically present during continuation of all patient treatments involving the
electronic brachytherapy device.
4)
A therapeutic radiological physicist shall designate shield locations or safe
distances sufficient to meet the requirements of 32 Ill. Adm. Code 340 for any
individual, other than the patient, in the treatment room;
5) All personnel in the treatment room are
required to remain behind shielding or at a safe distance specified by the
therapeutic radiological physicist during treatment. A therapeutic radiological
physicist shall approve any deviation from this requirement and shall designate
alternative radiation safety protocols, compatible with patient safety, to
provide an equivalent degree of protection.
i) Electronic Brachytherapy Source
Calibration Measurements
1) Calibration of the
electronic brachytherapy source output for an electronic brachytherapy device
shall be performed by, or under the direct supervision of, a therapeutic
radiological physicist.
2)
Calibration of the electronic brachytherapy source output shall be made for
each electronic brachytherapy source, or after any repair affecting the x-ray
beam generation, or when indicated by the electronic brachytherapy source
quality assurance checks.
3)
Calibration of the electronic brachytherapy source output shall utilize a
dosimetry system that meets the requirements of subsection
360.110(d)(4).
4) Calibration of the electronic
brachytherapy source output shall include, as applicable, determination of:
A) The output within 2% of the expected
value, if applicable, or determination of the output if there is no expected
value;
B) Timer accuracy and
linearity over the typical range of use;
C) Proper operation of back-up exposure
control devices;
D) Evaluation that
the relative dose distribution about the source is within 5% of that expected;
and
E) Source positioning accuracy
to within one millimeter within the applicator.
5) Calibration of the x-ray source output
shall be in accordance with the manufacturer's calibration protocol.
6) The registrant shall maintain a record of
each calibration in an auditable form for 5 years. The record shall include:
the date of the calibration; the manufacturer's name, model number and serial
number for the electronic brachytherapy device and a unique identifier for its
electronic brachytherapy source; the model numbers and serial numbers of the
instruments used to calibrate the electronic brachytherapy device; and the name
and signature of the therapeutic radiological physicist responsible for
performing the calibration.
j) Periodic and Day-of-Use Quality Assurance
Checks for Electronic Brachytherapy Devices
1)
Quality assurance checks shall be performed on each electronic brachytherapy
device:
A) At the beginning of each day of
use;
B) Each time the device is
moved to a new room or site; and
C)
After each x-ray tube installation.
2) The registrant shall perform periodic
quality assurance checks required by subsection (j)(1) in accordance with
procedures established by the therapeutic radiological physicist;
3) Quality assurance checks shall include, at
a minimum:
A) Verification that output of the
electronic brachytherapy source falls within 3% of expected values, as
appropriate for the device, as determined by:
i) Output as a function of time; or
ii) Output as a function of setting on a
monitor chamber.
B)
Verification of the consistency of the dose distribution to within 3% of that
found during calibration required by subsection (i); and
C) Validation of the operation of positioning
methods to ensure that the treatment dose exposes the intended location within
1 mm.
4) The registrant
shall use a dosimetry system that has been intercompared within the previous 12
months with the dosimetry system described in subsection (i)(3) to make the
quality assurance checks required in this subsection (j).
5) The registrant shall review the results of
each radiation output quality assurance check according to the following
procedures:
A) A physician and therapeutic
radiological physicist shall be immediately notified if any parameter is not
within its acceptable tolerance. The electronic brachytherapy device shall not
be made available for subsequent medical use until the therapeutic radiological
physicist has determined that all parameters are within their acceptable
tolerances; and
B) The therapeutic
radiological physicist shall review and sign the results of each radiation
output quality assurance check at intervals not to exceed 30 days.
6) Quality assurance checks shall,
at a minimum, assure:
A) Proper operation of
radiation exposure indicator lights on the electronic brachytherapy device and
on the control console;
B) Proper
operation of viewing and intercom systems in each electronic brachytherapy
facility, if applicable;
C) Proper
operation of radiation monitors, if applicable;
D) The integrity of all cables, catheters or
parts of the device that carry high voltages; and
E) Connecting guide tubes, transfer tubes,
transfer-tube-applicator interfaces, and treatment spacers are free from any
defects that interfere with proper operation.
7) If the results of the safety device
quality assurance checks indicate the malfunction of any system, a registrant
shall secure the control console in the OFF position and not use the electronic
brachytherapy device except as may be necessary to repair, replace, or check
the malfunctioning system.
8) The
registrant shall maintain a record of each quality assurance check in an
auditable form for 3 years.
A) The record
shall include the date of the quality assurance check; the manufacturer's name,
model number and serial number for the electronic brachytherapy device; the
name and signature of the individual who performed the periodic quality
assurance check; and the name and signature of the therapeutic radiological
physicist who reviewed the quality assurance check; and
B) The record shall also include the unique
identifier for the electronic brachytherapy source; the manufacturer's name;
and the model number and serial number for the instruments used to measure the
radiation output of the electronic brachytherapy device.
k) Therapy-Related Computer
Systems. The registrant shall perform acceptance testing on the treatment
planning system of electronic brachytherapy-related computer systems in
accordance with the manufacturer's acceptance testing protocol.
1) Acceptance testing shall be performed by,
or under the direct supervision of, a therapeutic radiological physicist. At a
minimum, the acceptance testing shall include, as applicable, verification of:
A) The source-specific input parameters
required by the dose calculation algorithm;
B) The accuracy of dose, dwell time, and
treatment time calculations at representative points;
C) The accuracy of isodose plots and graphic
displays;
D) The accuracy of the
software used to determine radiation source positions from radiographic images;
and
E) If the treatment planning
system is different from the treatment delivery system, the accuracy of
electronic transfer of the treatment delivery parameters to the treatment
delivery unit from the treatment planning system.
2) The position indicators in the applicator
shall be compared to the actual position of the source or planned dwell
positions, as appropriate, at the time of commissioning.
3) Prior to each patient treatment regimen,
the parameters for the treatment shall be evaluated and approved by the
physician and the therapeutic radiological physicist for correctness through
means independent of that used for the determination of the
parameters.
l) Training
1) A registrant shall provide instruction,
initially and at least annually, to all individuals who operate the electronic
brachytherapy device, as appropriate to the individual's assigned duties, in
the operating procedures identified in subsection (g). If the interval between
patients exceeds one year, retraining of the individuals shall be
provided.
2) Physicians,
therapeutic radiological physicists, and electronic brachytherapy device
operators shall receive device specific instruction initially from the
manufacturer and annually from either the manufacturer or other qualified
trainer. The training shall be of a duration recommended by the manufacturer's
training protocol. The training shall include, but not be limited to:
A) Device-specific radiation safety
requirements;
B) Device
operation;
C) Clinical use for the
types of use approved by the FDA;
D) Emergency procedures, including an
emergency drill; and
E) The
registrant's quality assurance program.
3) A registrant shall retain a record of
individuals receiving instruction required by this subsection (l) for 3 years.
The record shall include a list of the topics covered, the date of the
instruction, the names of the attendees, and the names of the individuals who
provided the instruction.
m) Mobile Electronic Brachytherapy Service. A
registrant providing mobile electronic brachytherapy service shall, as a
minimum:
1) Check all radiation survey
instruments before medical use at each address of use or on each day of use,
whichever is more restrictive.
2)
Account for the electronic brachytherapy source in the electronic brachytherapy
device before departure from the client's address.
3) Perform, at each location on each day of
use, all of the required quality assurance checks specified in subsection (j)
to assure proper operation of the device.
Notes
The following state regulations pages link to this page.
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.