Ill. Admin. Code tit. 77, § 675.100 - Instrumentation
a) Pure-tone
audiometers utilized for identification audiometry must comply with minimum
specifications established by the American National Standards Institute as
published in the American National Standard Specifications for Audiometers.
(ANSI 3.6 1996)
b) Pure-tone
audiometers utilized for identification audiometry must undergo an
electro-acoustic coupler calibration check a minimum of once per calendar year.
The electro-acoustic calibration check shall include the following
measurements:
1) frequency count;
2) attenuator linearity; and
3) earphone sound pressure level
output.
c) Annual
calibration check services can be acquired from either the Department (for a
$10 fee) or other qualified entity (a dealer or manufacturer who has technical
knowledge and experience in repair and maintenance of audiometric equipment).
Any audiometer that does not meet ANSI 3.6(1996) and/or does not pass the
listening or visual check is required to be calibrated and/or repaired by a
dealer or manufacturer who has technical knowledge and experience in repair and
maintenance of audiometric equipment.
Notes
Amended at 24 Ill. Reg. 4956, effective March 20, 2000