Mich. Admin. Code R. 333.5036 - Shielding plan review
Rule 36.
(1) An
applicant, before registration, shall submit the floor plans, shielding
specifications, and equipment arrangement of all new installations, or
modifications of existing installations, using a radiation machine, to the
department for review and approval. Application for a radiation shielding plan
review shall be completed on an application form supplied by the department
according to the instructions contained in that form. Radiation shielding plans
are not required to be submitted for any of the following:
(a) Dental intraoral or panoramic machines
used in dental clinical facilities.
(b) Cabinet x-ray systems, analytical
systems, process or control gauges, or cold-cathode gas discharge
tubes.
(c) Bone
densitometers.
(d) Mobile or
portable radiographic machines unless routinely used in 1 location.
(e) C-arm fluoroscopic systems having a
maximum source-image receptor distance of less than 45 centimeters that are
used for extremity use only.
(2) The department may require the applicant
to use the services of a health physicist or medical physicist to determine the
shielding requirements before the departments plan review and
approval.
(3) The departments
approval of a plan shall not preclude the requirement of additional
modifications if a subsequent analysis shows an individual could receive a dose
exceeding the limits in R 333.5057 to R 333.5060.
(4) Shielding design goals of 0.1
millisievert (10 mrem) per week for controlled areas and 0.02 millisievert (2
mrem) per week and 0.02 millisievert (2 mrem) in any 1 hour for uncontrolled
areas shall be applied to new facilities and to new construction in existing
facilities.
(a) For the purpose of this
subrule, "controlled area" means a limited access area where the occupational
exposure of personnel to radiation is under the supervision of an individual in
charge of radiation protection. In a controlled area, the access, occupancy,
and working conditions are controlled for the purpose of radiation
protection.
(b) For the purpose of
this subrule, "uncontrolled area" means all areas of the facility and the
surrounding environs that are not controlled for the purpose of radiation
protection.
(5) For
machines installed after the effective date of this part, the registrant shall
maintain for inspection by the department a scale drawing of the room where a
stationary radiation machine system is located. The drawing or accompanying
attachments shall indicate the use of areas adjacent to the room and include an
estimate of the occupancy in each area. In addition, the drawing or attachment
shall include at least 1 of the following:
(a)
The type and thickness of materials, or lead equivalency, of each protective
barrier.
(b) The results of a
survey for radiation levels at the operator's position and at pertinent points
outside the room under specified test conditions.
(6) The department may withhold initial
registration of a radiation machine pending receipt of either of the following:
(a) Plans and specifications for room design
and shielding and approval of those plans and specifications.
(b) Documentation from the applicant that a
certificate of need (CON) has been issued if the use of that machine is a
covered clinical service as defined in the act.