Mich. Admin. Code R. 333.5325 - X-ray equipment
Rule 325.
(1)
All x-ray tube housings in fixed radiographic installations shall be of the
diagnostic type.
(2) The aluminum
equivalent of the total filtration in the useful beam shall not be less than
the values shown in table 325-1.
TABLE 325-1
|
Operating Potential |
Minimum Total Filter (Inherent plus added) |
|
Below 50 kVp |
0.5 mm aluminum |
|
50-70 kVp |
1.5 mm aluminum |
|
Above 70 kVp |
2.5 mm aluminum |
(3) If the filter in the machine is not
accessible for examination and the total filtration is not known, subrule (2)
of this rule may be assumed to have been met if the half-value layer is not
less than any of the following:
(a) 0.6 mm
aluminum at 49 kVp.
(b) 1.6 mm
aluminum at 70 kVp.
(c) 2.6 mm
aluminum at 90 kVp.
(4)
Under conditions of subrule (3) of this rule for tube potentials above 90 kVp,
subrule (2) of this rule may be assumed to have been met if the half-value
layer is not less than that specified in table 325-2.
(5) The half-value layer of the useful beam
for a given x-ray tube potential shall not be less than the values shown in
table 325-2.
TABLE 325-2
|
Design Operating Range (kVp) |
Measured Potential (kVp) |
Half-value Layer (mm aluminum) |
|
Below 50 |
30 |
0.3 |
|
40 |
0.4 |
|
|
49 |
0.5 |
|
|
50 to 70 |
50 |
1.2 |
|
60 |
1.3 |
|
|
70 |
1.5 |
|
|
Above 70 |
80 |
2.3 |
|
90 |
2.5 |
|
|
100 |
2.7 |
|
|
110 |
3.0 |
|
|
120 |
3.2 |
|
|
130 |
3.5 |
|
|
140 |
3.8 |
|
|
150 |
4.1 |
(6) To determine the half-value layer at an
x-ray tube potential which is not listed in table 325-2, linear interpolation
or extrapolation may be made. Positive means shall be provided to ensure that
at least the minimum filtration needed to achieve these beam quality
requirements is in the useful beam during each exposure.
(7) Machines equipped with beryllium window
x-ray tubes with removable filters shall contain keyed filter interlock
switches in the tube housing and suitable indication on the control panel of
the added filter in the useful beam. The total filtration permanently in the
useful beam shall not be less than 0.5 millimeter aluminum equivalent and shall
be clearly indicated on the tube housing.
(8) Beryllium window x-ray tubes shall not be
used routinely for general purpose diagnostic examinations. Such a tube may
comprise an x-ray subsystem if needed for special soft tissue technique in
accord with subrule (7) of this rule.
(9) Beam-limiting devices, such as
diaphragms, cones, or adjustable collimators, capable of restricting the useful
beam to the area radiographically recorded shall be provided to define the beam
and shall provide the same degree of attenuation as that required of the tube
housing.
(10) Beam-limiting devices
shall be calibrated in terms of the size of the projected useful beam at
specified source-image receptor distances (SID). This calibration shall be
clearly and permanently recorded on the beam-limiting device. Calibration of
adjustable beam-limiting devices shall permit reproducible settings.
(11) X-ray systems designed for only 1 image
receptor size at a fixed SID shall be provided with means to limit the field at
the plane of the image receptor to dimensions no greater than those of the
image receptor, and to align the center of the x-ray field with the center of
the image receptor to within 2% of the SID.
(12) General purpose radiographic x-ray
systems shall be equipped with adjustable beam-limiting devices containing
light localizers that define the entire field.
(13) The size of the x-ray beam projected by
fixed aperture beam-limiting devices, except those used for stereoradiography,
shall not exceed the dimensions of the image receptor by more than 2% of the
SID when the axis of the x-ray beam is perpendicular to the plane of the image
receptor.
(14) The calibrated field
size indicator on adjustable beam-limiting devices shall be accurate to within
2% of the SID. The light field shall be aligned with the x-ray field with the
same degree of accuracy. The field size projected by automatic adjustable
beam-limiting devices shall provide the same precision.
(15) For radiographic procedures resulting in
multiple views on a single image receptor, the beam-limiting device shall limit
the x-ray field size to the recorded radiographic image size within 2% of the
SID. Covering a portion of the image receptor with radio-opaque material is not
a substitute for proper x-ray field limitation.
(16) Radiographic x-ray machines used for
purposes other than mammography or extremity radiography only shall be capable
of operation at not less than an average current of 100 milliamperes (mA)
during all radiographic techniques used. A machine not capable of sustained
operation at not less than an average of 100 mA for the duration of a given
technique shall not be used for that technique.
(17) A device shall be provided which
terminates the exposure at a preset time interval or exposure limit. The
operator shall be able to terminate the exposure at any time by discontinuing
pressure upon the exposure switch except that during serial radiography means
may be provided to permit completion of a single exposure in
progress.
(18) The exposure switch,
except for those used in conjunction with spot film devices in fluoroscopy,
shall be securely fixed so that the operator is required to be behind a fixed
shield which intercepts the useful beam and any radiation which has been
scattered only once.
(19) When 2 or
more x-ray tube heads are operated from a single exposure switch, there shall
be indication at the control panel showing which tube is connected and ready to
be energized and means to prevent energizing more than 1 tube head
simultaneously. Machines designed for simultaneous multiple tube operation
shall have positive means for selecting single tube or multiple tube
operation.
(20) The control panel
shall provide positive visual identification of the production of x-rays when
the x-ray tube is energized. A milliammeter may comply with this
subrule.
(21) A signal audible to
the operator shall indicate that the exposure has ended.
(22) The technique factors to be used during
an exposure shall be indicated before the exposure begins. When automatic
exposure controls are used, only those technique factors which are set before
the exposure shall be indicated. On equipment having fixed technique factors,
this requirement may be met by permanent markings. Indication of technique
factors shall be visible from the operator's position.
Notes
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.