Mich. Admin. Code R. 333.5409 - Conditions of operation
Rule 409.
(1)
An operator shall properly utilize the beam-limiting devices provided to
restrict the useful beam to the smallest area consistent with clinical
requirements. Particular care shall be taken to align accurately the x-ray beam
with the patient and film.
(2) The
operator shall ensure the presence of adequate filtration before a radiographic
procedure pursuant to R 333.5405(2).
(3) When a patient or film is held in
position for radiography, mechanical supporting or restraining devices shall be
available and shall be used unless contraindicated. Proper use of these devices
shall permit the operator to stand behind the primary control shield during
most radiographic procedures.
(4)
If the patient or film is held by 1 or more individuals, each individual shall
wear protective gloves and body aprons of 0.5 millimeter minimum lead
equivalence. Each individual shall be so positioned that no part of his or her
body is struck by the useful beam and that his or her body is as far as
possible from the edge of the useful beam.
(5) Only individuals whose presence is
necessary shall be permitted in the radiographic room during an exposure. These
individuals shall be protected as specified in subrule (4) of this rule unless
protected by an approved primary barrier.
(6) If an auxiliary foot switch is provided
as specified in R 333.5405(12), it shall be used only by a licensed
veterinarian and only at times when sufficient personnel are not available to
permit use of the primary exposure switch specified in R
333.5405(11).
(7) To protect the
feet of the veterinarian or his or her assistant from the primary beam while
restraining patients, the underside of the radiographic table shall be
protected by at least 1.6 millimeter (1/16 inch) lead or equivalent protection
approved by the department.
(8)
Personnel monitoring shall be performed in controlled areas for each individual
occupationally exposed to ionizing radiation from diagnostic x-ray equipment.
Individual monitoring devices, such as film badge dosimeters or
thermoluminescent dosimeters, shall be permanently assigned to each
occupationally exposed individual. This monitoring shall be continuous during
employment as a radiation worker.
(9) Personnel exposure records shall be kept
on permanent available file at the facility where the exposure
occurs.
(10) Monitoring devices
used to estimate whole body exposure shall normally be worn on the chest or
abdomen. Monitoring of all other body parts shall comply with R
333.5065.
(11) Monitoring devices
worn to estimate personnel occupational exposure shall not be worn by the
individual when he or she is exposed as a patient for a medical or dental
reason.
(12) Medical x-ray screen
type films and intensifying screens shall be employed to reduce patient
exposure except in cases where a noticeable decrease in image definition
resulting from increased sensitivity may reduce the clinical value of the
examination.
(13) Film processing
materials and techniques shall be those recommended by the x-ray film and
processing materials manufacturers unless otherwise tested to ensure maximum
information content of the developed film. Sight developing is not permitted
except under extreme emergency conditions. Correct temperature control and
development time are necessary to minimize radiation dose to the
patient.
(14) A radiographic x-ray
system shall not be left unattended without locking the apparatus, room, or
building in some manner which prevents use of the apparatus by unauthorized
persons.