Mich. Admin. Code R. 333.5694 - Mammography equipment evaluations
Rule 694.
(1)
Additional evaluations of stereotactic breast biopsy units or image processors
shall be conducted when a new unit or processor is installed, a unit or
processor is disassembled and reassembled at the same or a new location, or
major components of a stereotactic breast biopsy unit or processor equipment
are changed or repaired. These evaluations shall be used to determine whether
the new or changed equipment meets the requirements of R 333.5676 to R 333.5678
and R 333.5687 to R 333.5691, as applicable. Problems revealed by the
evaluation shall be corrected before the new or changed equipment is put into
service for procedures or film processing.
(2) The equipment evaluations specified in
subrule (1) of this rule shall be performed by a medical physicist or by an
individual under the direct supervision of a medical physicist.